CLASS II ONGOING Device recall · Reported 5 Feb 2025 · FDA Enforcement Report
Calibrate CCX Interbody System for spinal fusion procedures recalled
Alphatec Spine is recalling Calibrate CCX Interbody System for spinal fusion procedures: Implants: Calibrate Ccx-o 5, distributed nationwide in the US. The recall was initiated on 15 Dec 2024 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1065-2025
- FDA event ID
- 96009
- Recalling firm
- Alphatec Spine, Carlsbad, CA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 15 Dec 2024
- FDA classified
- 30 Jan 2025
- Reported
- 5 Feb 2025
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Spinal Implants
Timeline
| 15 Dec 2024 | Recall initiated by company | |
| 30 Jan 2025 | Classified by FDA | +46 days |
| 5 Feb 2025 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Complaints have been received regarding post-operative implant collapse.
Product as listed by the FDA
Calibrate CCX Interbody System for spinal fusion procedures: Implants: (1) Calibrate CCX-O, 7 x 10 x 25mm 5 degrees, Part #325-07102505, nonsterile; (2) Calibrate CCX-O, 7 x 10 x 25mm 10 degrees, Part #325-07102510, nonsterile; (3) Calibrate CCX-O, 7 x 10 x 25mm 15 degrees, Part #325-07102515, nonsterile; (4) Calibrate CCX-O, 7 X 10 X 30mm 5 degrees, Part #325-07103005, nonsterile; (5) Calibrate CCX-O, 7 x 10 x 30mm 10 degrees, Part #325-07103010, nonsterile; (6) Calibrate CCX-O, 7 x 10 x 30mm 15 degrees, Part #325-07103015, nonsterile; (7) Calibrate CCX-O, 7 x 10 x 35mm 10 degrees, Part #325-07103510, nonsterile; (8) Calibrate CCX-O, 7 x 10 x 35mm 15 degrees, Part #325-07103515, nonsterile; (9) Calibrate CCX-O, 9 x 10 x 25mm 5 degrees, Part #325-09102505, nonsterile; (10) Calibrate CCX-O, 9 x 10 x 25mm 10 degrees, Part #325-09102510, nonsterile; (11) Calibrate CCX-O, 9 x 10 x 30mm 5 degrees, Part #325-09103005, nonsterile; (12) Calibrate CCX-O, 9 x 10 x 30mm 10 degrees, Part #325-09103010, nonsterile; (13) Calibrate CCX-O, 11 x 10 x 30mm 10 degrees, Part #325-11103010, nonsterile; (14) Calibrate CCX-S, 7 x 10 x 22mm 5 degrees, Part #326-07102205, nonsterile; (15) Calibrate CCX-S, 7 x 10 x 22mm 10 degrees, Part #326-07102210, nonsterile; (16) Calibrate CCX-S, 7 x 10 x 22mm 15 degrees, Part #326-07102215, nonsterile; (17) Calibrate CCX-S, 7 x 10 x 26mm 5 degrees, Part #326-07102605, nonsterile; (18) Calibrate CCX-S, 7 x 10 x 26mm 10 degrees, Part #326-07102610, nonsterile; (19) Calibrate CCX-S, 7 x 10 x 26mm 15 degrees, Part #326-07102615, nonsterile; (20) Calibrate CCX-S, 7 x 10 x 30mm 10 degrees, Part #326-07103010, nonsterile; (21) Calibrate CCX-S, 9 x 10 x 22mm 5 degrees, Part #326-09102205, nonsterile; (22) Calibrate CCX-S, 9 x 10 x 22mm 10 degrees, Part #326-09102210, nonsterile; (23) Calibrate CCX-S, 9 x 10 x 26mm 5 degrees, Part #326-09102605, nonsterile; (24) Calibrate CCX-S, 9 x 10 x 26mm 10 degrees, Part #326-09102610, nonsterile; and (25) Calibrate CCX-S, 11 x 10 x 26mm 10 degrees, Part #326-11102610, nonsterile. Convenience kits containing affected implants: (1) Calibrate CCX Oblique Implants, REF CLCCXOIMPA, containing 24 implants/kit, nonsterile; and (2) Calibrate CCX Straight Implants, REF CLCCXSIMPA, containing 36 implants/kit, nonsterile.
Where it was distributed
US Nationwide distribution in the states of AZ, CA, CO, CT, FL, GA, IA, IL, IN, MA, MD, MI, MN, NC, NJ, NY, OH, OK, OR, PA, SC, SD, TN, and UT.
Nationwide ArizonaCaliforniaColoradoConnecticutFloridaGeorgiaIowaIllinoisIndianaMassachusettsMarylandMichiganMinnesotaNorth CarolinaNew JerseyNew YorkOhioOklahomaOregonPennsylvania
Questions about this recall
Why was it recalled?
Complaints have been received regarding post-operative implant collapse.
Which lots are affected?
Implants: (1) Lot numbers AP02103, AP02374, AP02426, AP02442, AP02766, AP02967, AP03207, UDI-DI 00190376533308; (2) Lot numbers AP02104, AP02375, AP02427, AP02770, AP02780, AP02833, AP02834, AP02845, AP02846, AP02894, UDI-DI 00190376532967; (3) Lot numbers AP02105, AP02376, AP02782, AP02893, AP02957, AP03109, AP03110, UDI-DI 00190376532974; (4) Lot numbers AP02114, AP02377, AP02428, AP02429, AP02767, AP02919, UDI-DI 00190376532998; (5) Lot numbers AP02106, AP02263, AP02378, AP02430, AP02771, AP02783, AP02836, AP02837, AP02838, AP02839, UDI-DI 00190376533001; (6) Lot numbers AP02107, AP02379, AP02431, AP02781, AP02835, AP02847, AP02848, AP02892. UDI-DI 00190376533018; (7) Lot numbers AP02108, AP02380, AP02432, AP02433, UDI-DI 00190376533049; (8) Lot number AP02144, UDI-DI 00190376533056; (9) Lot numbers AP02109, AP02257, AP02381, AP02434, AP02435, AP02436, AP02768, AP03180, UDI-DI 00190376533322; (10) Lot numbers AP02110, AP02382, AP02772, AP02962, UDI-DI 00190376533339; (11) Lot numbers AP02111, AP02259, AP02383, AP02437, AP02438, UDI-DI 00190376533360; (12) Lot numbers AP02112, AP02384, AP02626, AP02711, AP02773, UDI-DI 00190376533377; (13) Lot numbers AP02113, AP02385, AP02439, AP02440, AP02441, AP02774, AP02921, UDI-DI 00190376532653; (14) Lot numbers AP02115, AP02258, AP02283, AP02301, AP02710, AP02925, AP02971, AP03100, UDI-DI 00190376530277; (15) Lot numbers AP02116, AP02302, AP02871, AP02872, AP02896, UDI-DI 00190376530284; (16) Lot numbers AP02117, AP02303, AP02623, UDI-DI 00190376530291; (17) Lot numbers AP02118, AP02304, AP02775, AP02869, AP02875, UDI-DI 00190376530314; (18) Lot numbers AP02119, AP02305, AP02422, AP02777, AP02785, AP02843, UDI-DI 00190376530321; (19) Lot numbers AP02120, AP02306, AP02624, AP02744, AP02895, AP02922, UDI-DI 00190376530338; (20) Lot numbers AP02121, AP02307, AP02423, UDI-DI 00190376530369; (21) Lot numbers AP02122, AP02312, AP02424, UDI-DI 00190376530758; (22) Lot numbers AP02123, AP02306, AP02308, AP02625, UDI-DI 00190376530765; (23) Lot numbers AP02124, AP02309, AP02425, AP02776, AP02786, AP02844, UDI-DI 00190376530796; (24) Lot numbers AP02125, AP02310, AP02778, AP02784, AP02840, AP02841, AP02842, AP02870, AP03095, UDI-DI 00190376530802; and (25) Lot numbers AP02126, AP02311, AP02779, UDI-DI 00190376531281. Convenience kits containing affected implants: (1) REF CLCCXOIMPA - Serial numbers 2000494, 2000495, 2000496, 2000497, 2000498, 2000499, 2000500, 2000501, 2000502, and 2000503; UDI-DI 00190376553535; and (2) REF CLCCXSIMPA - Serial numbers 2000727, 2000728, 2000730, and 2000731; UDI-DI 00190376553542.
Where was it sold?
US Nationwide distribution in the states of AZ, CA, CO, CT, FL, GA, IA, IL, IN, MA, MD, MI, MN, NC, NJ, NY, OH, OK, OR, PA, SC, SD, TN, and UT.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Alphatec Spine
All 46| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | SafeOp Patient Interface recalled by Alphatec Spine | Alphatec Spine | Device Malfunction | Nationwide |
| Device | CLASS II | Atec Lateral Navigation Disc Prep Instruments Lif Nav, Osteotome recalled | Alphatec Spine | Packaging Defects | Nationwide |
| Device | CLASS II | Atec Lateral Navigation Disc Prep Instruments Lif Nav, Osteotome recalled | Alphatec Spine | Packaging Defects | Nationwide |
| Device | CLASS II | IdentiTi ALIF Standalone Interbody System recalled by Alphatec Spine | Alphatec Spine | Device Malfunction | Nationwide |
| Device | CLASS II | LIF AMP, Adjustable Awl recalled by Alphatec Spine | Alphatec Spine | Other | Nationwide |
Other Spinal Implants recalls
All 1,161| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline surgical convenience kits labeled as: Spinal CDS recalled | Medline Industries | Packaging Defects | Nationwide |
| Device | CLASS II | Medline surgical convenience kits labeled as recalled | Medline Industries | Packaging Defects | Nationwide |
| Device | CLASS II | Extended Polyaxial Head, Attached Gimbal recalled | Seaspine Orthopedics | Other | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Anterior Spine Pack recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Neuro; Neuro; Crani; Ortho Spine Pack; Bmhm Neuro Pack recalled | Avid Medical | Sterility | Nationwide |