CLASS II ONGOING Device recall · Reported 26 Mar 2025 · FDA Enforcement Report

Capnocheck BCI Airway Adapter recalled by Smiths Medical ASD

Smiths Medical ASD is recalling Capnocheck BCI Airway Adapter, distributed nationwide in the US. The recall was initiated on 14 Feb 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI 30610586038444 Lots WW8214 022980-1-22 022404-1-17 021860-1-10 021110-1-7 020609-1-2 016288-1-1 022980-1-24 022404-1-18 021860-1-14 021110-1-8 020609-1-3 017774-1-1 022980-1-23 022404-1-19 021860-1-13 021110-1-9 020609-1-4 018139-1-1 022980-1-26 022404-1-20 021860-1-15 021110-1-10 020609-1-5 018184-1-1 022980-1-27 022404-1-21 021860-1-17 021110-1-11 020609-1-6 018252-1-1 022980-1-28 022404-1-22 021860-1-16 021110-1-12 020609-1-7 018458-1-1 022980-1-25 022404-1-23 021860-1-18 021110-1-13 020609-1-8 018457-1-1 022980-1-32 022404-1-24 021860-1-20 021110-1-14 020609-1-9 018614-1-1 022980-1-30 022404-1-25 021860-1-21 021110-1-16 020609-1-10 018641-1-1 022980-1-29 022404-1-26 021860-1-19 021110-1-15 020609-1-11 018789-1-1 022980-1-31 022404-2-2 021860-2-1 021110-1-17 020609-1-12 018789-2-1 022980-1-33 022404-1-27 021860-2-2 021110-1-18 020609-1-13 018789-3-1 023592-1-3 022404-2-1 021860-1-22 021110-1-19 020609-1-14 018789-4-1 023592-1-2 022404-2-3 021860-2-5 021110-1-20 020609-1-15 018936-1-1 023592-1-1 022404-2-4 021860-2-3 021110-1-21 020609-1-16 018936-2-1 023592-1-5 022404-2-5 021860-2-4 021110-1-22 020609-1-19 018936-3-1 023592-1-6 022404-2-6 021860-2-6 021110-1-23 020609-1-20 018936-4-1 023592-1-4 022404-2-7 021860-2-8 021110-1-24 020609-1-17 019179-1-1 023592-1-7 022404-2-8 021860-2-7 021110-1-25 020609-1-21 019254-1-1 023592-1-9 022404-2-9 021860-2-10 021110-1-26 020609-1-18 019323-1-1 023592-1-8 022404-2-10 021860-2-11 021110-1-27 020609-1-23 019383-1-1 023592-1-12 022404-2-11 021860-2-9 021110-1-28 020609-1-22 019409-1-1 023592-1-11 022404-2-12 021860-2-12 021110-1-29 020609-1-24 019410-1-1 023592-1-10 022404-2-13 021860-2-13 021110-1-30 020609-1-26 019549-1-1 023592-1-15 022980-1-3 021860-2-14 021110-1-33 020609-1-27 019613-1-1 023592-1-16 022980-1-1 021860-2-15 021110-1-31 020609-1-25 019671-1-1 023592-1-17 022980-1-2 021860-2-16 021110-1-34 020609-1-28 019747-1-1 023592-1-18 022980-1-4 021860-2-17 021110-1-35 020609-1-30 019782-1-1 023592-1-19 022980-1-5 021860-2-18 021110-1-32 020609-1-31 019810-1-1 023592-1-20 022980-1-6 022404-1-2 021110-1-36 020609-1-29 019920-1-1 9755313 022980-1-7 022404-1-1 021110-1-37 020609-1-34 019974-1-1 9755312 022980-1-8 022404-1-3 021110-1-38 020609-1-32 020000-1-1 9757103 022980-1-10 022404-1-4 021110-1-39 020609-1-33 020022-1-1 9757119 022980-1-11 022404-1-5 021110-1-40 020609-1-35 020022-2-1 9755311 022980-1-9 022404-1-6 021860-1-3 020609-1-36 020074-1-1 9759068 022980-1-12 022404-1-7 021860-1-4 020609-1-37 020126-1-1 9757614 022980-1-13 022404-1-8 021860-1-2 020609-1-38 020127-1-1 9761695 022980-1-15 022404-1-9 021860-1-1 020609-1-39 020159-1-1 9760404 022980-1-14 022404-1-10 021860-1-5 020609-1-40 020160-1-1 9762385 022980-1-16 022404-1-11 021860-1-7 021110-1-1 020227-1-1 9759947 022980-1-17 022404-1-12 021860-1-6 021110-1-2 020323-1-1 9759328 022980-1-18 022404-1-13 021860-1-11 021110-1-3 020346-1-1 9762393 022980-1-19 022404-1-14 021860-1-8 021110-1-4 020386-1-1 9763712 022980-1-20 022404-1-15 021860-1-9 021110-1-5 020435-1-1 9763703 022980-1-21 022404-1-16 021860-1-12 021110-1-6 020466-1-1 020609-1-1 020561-1-1 020551-1-1 020505-1-1 /
Quantity recalled175,994 US; 45,223 OUS

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1405-2025
FDA event ID
96238
Recalling firm
Smiths Medical ASD, Minneapolis, MN
Classification
Class II
Status
Ongoing
Recall initiated
14 Feb 2025
FDA classified
20 Mar 2025
Reported
26 Mar 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Labeling Errors

Timeline

14 Feb 2025Recall initiated by company
20 Mar 2025Classified by FDA+34 days
26 Mar 2025Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Airway adapter does not meet specification and may break off when attaching a valve or tube. Device failure may result in inaccurate readings, a break in airway circuit, and loss of ventilation. A break in the airway circuit could result in inability to ventilate, hypoxia, cardiopulmonary collapse, or death.

Product as listed by the FDA

CAPNOCHECK BCI Airway Adapter; Model Number BSR; Product Code WW8214

Where it was distributed

US Nationwide distribution. International distribution pending.

Nationwide

Questions about this recall

Why was it recalled?

Airway adapter does not meet specification and may break off when attaching a valve or tube. Device failure may result in inaccurate readings, a break in airway circuit, and loss of ventilation. A break in the airway circuit could result in inability to ventilate, hypoxia, cardiopulmonary collapse, or death.

Which lots are affected?

UDI-DI 30610586038444 Lots WW8214 022980-1-22 022404-1-17 021860-1-10 021110-1-7 020609-1-2 016288-1-1 022980-1-24 022404-1-18 021860-1-14 021110-1-8 020609-1-3 017774-1-1 022980-1-23 022404-1-19 021860-1-13 021110-1-9 020609-1-4 018139-1-1 022980-1-26 022404-1-20 021860-1-15 021110-1-10 020609-1-5 018184-1-1 022980-1-27 022404-1-21 021860-1-17 021110-1-11 020609-1-6 018252-1-1 022980-1-28 022404-1-22 021860-1-16 021110-1-12 020609-1-7 018458-1-1 022980-1-25 022404-1-23 021860-1-18 021110-1-13 020609-1-8 018457-1-1 022980-1-32 022404-1-24 021860-1-20 021110-1-14 020609-1-9 018614-1-1 022980-1-30 022404-1-25 021860-1-21 021110-1-16 020609-1-10 018641-1-1 022980-1-29 022404-1-26 021860-1-19 021110-1-15 020609-1-11 018789-1-1 022980-1-31 022404-2-2 021860-2-1 021110-1-17 020609-1-12 018789-2-1 022980-1-33 022404-1-27 021860-2-2 021110-1-18 020609-1-13 018789-3-1 023592-1-3 022404-2-1 021860-1-22 021110-1-19 020609-1-14 018789-4-1 023592-1-2 022404-2-3 021860-2-5 021110-1-20 020609-1-15 018936-1-1 023592-1-1 022404-2-4 021860-2-3 021110-1-21 020609-1-16 018936-2-1 023592-1-5 022404-2-5 021860-2-4 021110-1-22 020609-1-19 018936-3-1 023592-1-6 022404-2-6 021860-2-6 021110-1-23 020609-1-20 018936-4-1 023592-1-4 022404-2-7 021860-2-8 021110-1-24 020609-1-17 019179-1-1 023592-1-7 022404-2-8 021860-2-7 021110-1-25 020609-1-21 019254-1-1 023592-1-9 022404-2-9 021860-2-10 021110-1-26 020609-1-18 019323-1-1 023592-1-8 022404-2-10 021860-2-11 021110-1-27 020609-1-23 019383-1-1 023592-1-12 022404-2-11 021860-2-9 021110-1-28 020609-1-22 019409-1-1 023592-1-11 022404-2-12 021860-2-12 021110-1-29 020609-1-24 019410-1-1 023592-1-10 022404-2-13 021860-2-13 021110-1-30 020609-1-26 019549-1-1 023592-1-15 022980-1-3 021860-2-14 021110-1-33 020609-1-27 019613-1-1 023592-1-16 022980-1-1 021860-2-15 021110-1-31 020609-1-25 019671-1-1 023592-1-17 022980-1-2 021860-2-16 021110-1-34 020609-1-28 019747-1-1 023592-1-18 022980-1-4 021860-2-17 021110-1-35 020609-1-30 019782-1-1 023592-1-19 022980-1-5 021860-2-18 021110-1-32 020609-1-31 019810-1-1 023592-1-20 022980-1-6 022404-1-2 021110-1-36 020609-1-29 019920-1-1 9755313 022980-1-7 022404-1-1 021110-1-37 020609-1-34 019974-1-1 9755312 022980-1-8 022404-1-3 021110-1-38 020609-1-32 020000-1-1 9757103 022980-1-10 022404-1-4 021110-1-39 020609-1-33 020022-1-1 9757119 022980-1-11 022404-1-5 021110-1-40 020609-1-35 020022-2-1 9755311 022980-1-9 022404-1-6 021860-1-3 020609-1-36 020074-1-1 9759068 022980-1-12 022404-1-7 021860-1-4 020609-1-37 020126-1-1 9757614 022980-1-13 022404-1-8 021860-1-2 020609-1-38 020127-1-1 9761695 022980-1-15 022404-1-9 021860-1-1 020609-1-39 020159-1-1 9760404 022980-1-14 022404-1-10 021860-1-5 020609-1-40 020160-1-1 9762385 022980-1-16 022404-1-11 021860-1-7 021110-1-1 020227-1-1 9759947 022980-1-17 022404-1-12 021860-1-6 021110-1-2 020323-1-1 9759328 022980-1-18 022404-1-13 021860-1-11 021110-1-3 020346-1-1 9762393 022980-1-19 022404-1-14 021860-1-8 021110-1-4 020386-1-1 9763712 022980-1-20 022404-1-15 021860-1-9 021110-1-5 020435-1-1 9763703 022980-1-21 022404-1-16 021860-1-12 021110-1-6 020466-1-1 020609-1-1 020561-1-1 020551-1-1 020505-1-1 /

Where was it sold?

US Nationwide distribution. International distribution pending.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

1 other product · FDA event 96238

This product is part of a recall event; the main page for the event is Capnocheck BCI Airway Adapter recalled by Smiths Medical ASD.

ReportedClassRecallCompanyReasonWhere
CLASS IICapnocheck BCI Airway Adapter recalled by Smiths Medical ASDSmiths Medical ASDLabeling ErrorsNationwide

More recalls from Smiths Medical ASD

All 284

Other Other Medical Devices recalls

All 13,778