CLASS II Device recall · Reported 4 Nov 2015 · FDA Enforcement Report
CardioCall ECG Event Recorder recalled by Spacelabs Healthcare
Spacelabs Healthcare is recalling CardioCall ECG Event Recorder, distributed in Nebraska, South Carolina, Washington. The recall was initiated on 3 Sep 2015 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0172-2016
- FDA event ID
- 72211
- Recalling firm
- Spacelabs Healthcare, Snoqualmie, WA
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 3 Sep 2015
- FDA classified
- 27 Oct 2015
- Reported
- 4 Nov 2015
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Surgical Instruments
Timeline
| 3 Sep 2015 | Recall initiated by company | |
| 27 Oct 2015 | Classified by FDA | +54 days |
| 4 Nov 2015 | Published in enforcement report | +8 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
When the battery is inserted into the CardioCall ECG Event Recorder, the device beeps twice and is followed by a continuous ringing sound that only stops when battery is removed. The device has a self-test failure and will not operate normally.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
Spacelabs Healthcare CardioCall ECG Event Recorder, Models VS20, VS20x4, ST80, ST80x4, PCBAs PN # 670-1670-00 Product Usage: The CardioCall event recorder is worn by the patient to record a short period of electrocardiogram which can be sent to his physician.
Where it was distributed
Worldwide Distribution - US Distribution in the states of Nebraska, South Carolina, and Washington and in the following countries of AUSTRALIA, BELGIUM, BOLIVIA, CANADA, CHILE, DENMARK, GERMANY, GUATEMALA, HONG KONG, ITALY, NETHERLANDS, NEW ZEALAND, POLAND, PORTUGAL, QATAR, SINGAPORE, TURKEY, UNITED ARAB EMIRATES, and UNITED KINGDOM.
Questions about this recall
Why was it recalled?
When the battery is inserted into the CardioCall ECG Event Recorder, the device beeps twice and is followed by a continuous ringing sound that only stops when battery is removed. The device has a self-test failure and will not operate normally.
Which lots are affected?
PCBAs PN #: 670-1670-00. SERIAL NUMBERS in the US: CARD-200990, CARD-201287, CARD-623478, CARD-623775, CARD-624198, CARD-624224, CARD-624314, CARD-624387, CARD-624404, CARD-624413, CARD-624891, CARD-625061, CARD-625089, CARD-625124, CARD-625232, CARD-625269, CARD-625296, CARD-625304, CARD-625322, and CARD-625421. SERIAL NUMBERS OUTSIDE THE US: CARD-000019, CARD-000028, CARD-000037, CARD-000046, CARD-000064, CARD-000073, CARD-000073, CARD-000082, CARD-000091, CARD-000109, CARD-000118, CARD-000127, CARD-000136, CARD-000145, CARD-000163, CARD-000172, CARD-000181, CARD-000190, CARD-000208, CARD-000217, CARD-000226, CARD-000235, CARD-000244, CARD-000271, CARD-000280, CARD-000299, CARD-000307, CARD-000325, CARD-000334, CARD-000343, CARD-000352, CARD-000361, CARD-000370, CARD-000398, CARD-000406, CARD-000415, CARD-000424, CARD-000433, CARD-000442, CARD-000451, CARD-000479, CARD-000479, CARD-000488, CARD-000497, CARD-000505, CARD-000514, CARD-000523, CARD-000532, CARD-000550, CARD-000578, CARD-000596, CARD-000604, CARD-000613, CARD-000622, CARD-000631, CARD-000640, CARD-000659, CARD-000668, CARD-000677, CARD-000686, CARD-000703, CARD-000721, CARD-000730, CARD-000749, CARD-000758, CARD-000767, CARD-000776, CARD-000785, CARD-000785, CARD-000794, CARD-000802, CARD-000811, CARD-000820, CARD-000820, CARD-000839, CARD-000848, CARD-000857, CARD-000866, CARD-000875, CARD-000893, CARD-000910, CARD-000910, CARD-000929, CARD-000938, CARD-000947, CARD-000956, CARD-000974, CARD-000983, CARD-000992, CARD-001009, CARD-001018, CARD-001027, CARD-001027, CARD-001036, CARD-001045, CARD-001054, CARD-001063, CARD-001072, CARD-001072, CARD-001081, CARD-001090, CARD-001108, CARD-001117, CARD-001135, CARD-001144, CARD-001153, CARD-001171, CARD-001180, CARD-001199, CARD-001199, CARD-001207, CARD-001216, CARD-001225, CARD-001234, CARD-001243, CARD-001252, CARD-001261, CARD-001270, CARD-001289, CARD-001298, CARD-001306, CARD-001315, CARD-001333, CARD-001379, CARD-001388, CARD-001397, CARD-001405, CARD-001414, CARD-001423, CARD-001441, CARD-001603, CARD-002602, CARD-003304, CARD-003601, CARD-003908, CARD-005203, CARD-006103, CARD-200035, CARD-200116, CARD-200125, CARD-200486, CARD-200963, CARD-201016, CARD-201052, CARD-201124, CARD-201133, CARD-300016, CARD-300025, CARD-300034, CARD-300052, CARD-300070, CARD-300089, CARD-300098, CARD-300160, CARD-300188, CARD-300197, CARD-300205, CARD-300214, CARD-300223, CARD-300232, CARD-300269, CARD-300278, CARD-300313, CARD-300322, CARD-300331, CARD-300368, CARD-300395, CARD-300412, CARD-300421, CARD-300430, CARD-300458, CARD-300467, CARD-300476, CARD-300494, CARD-300511, CARD-300539, CARD-300548, CARD-300566, CARD-300576, CARD-300647, CARD-300656, CARD-300674, CARD-300692, CARD-300728, CARD-300755, CARD-300773, CARD-300782, CARD-300809, CARD-300818, CARD-300836, CARD-300881, CARD-300890, CARD-300917, CARD-300926, CARD-300944, CARD-301006, CARD-301042, CARD-301079, CARD-301114, CARD-301150, CARD-301178, CARD-301439, CARD-301484, CARD-400213, CARD-621894, CARD-623973, CARD-624846, CARD-625043, CARD-625160, CARD-625188, CARD-625250, CARD-625476, CARD-625511, CARD-625665, CARD-625818, CARD-627023, CARD-627546, CARD-627555, CARD-627564, CARD-627636, CARD-627645, CARD-627654, CARD-627663, CARD-627672, CARD-627708, CARD-627726, CARD-627744, CARD-627753, CARD-627771, CARD-627780, CARD-627834, CARD-627861, CARD-627898, CARD-627960, CARD-627997, CARD-628013, CARD-628031, CARD-628059, CARD-628077, CARD-628095, CARD-628112, CARD-628176, CARD-628220, CARD-628239, CARD-630065, CARD-630137, CARD-630146, CARD-630281, CARD-630443, CARD-630461, CARD-630498, CARD-630885, CARD-631235, CARD-631271, and CARD-631307.
Where was it sold?
Worldwide Distribution - US Distribution in the states of Nebraska, South Carolina, and Washington and in the following countries of AUSTRALIA, BELGIUM, BOLIVIA, CANADA, CHILE, DENMARK, GERMANY, GUATEMALA, HONG KONG, ITALY, NETHERLANDS, NEW ZEALAND, POLAND, PORTUGAL, QATAR, SINGAPORE, TURKEY, UNITED ARAB EMIRATES, and UNITED KINGDOM.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Spacelabs Healthcare
All 47| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Xhibit Central Station recalled by Spacelabs Healthcare | Spacelabs Healthcare | Device Malfunction | Nationwide |
| Device | CLASS II | Spacelabs Ultraview SL Command Module SL Command Module recalled | Spacelabs Healthcare | Packaging Defects | Nationwide |
| Device | CLASS II | Multi-parameter Command Module recalled by Spacelabs Healthcare | Spacelabs Healthcare | Labeling Errors | Nationwide |
| Device | CLASS II | Sentinel V11 recalled by Spacelabs Healthcare | Spacelabs Healthcare | Device Malfunction | Nationwide |
| Device | CLASS II | Branding is not applied to this product DVI display cables recalled | Spacelabs Healthcare | Device Malfunction | Nationwide |
Other Surgical Instruments recalls
All 3,809| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline Convenience Kits: IV Start KIT recalled by Medline Industries | Medline Industries | Sterility | Nationwide |
| Device | CLASS II | Medline Convenience Kits containing BD ChloraPrep Clear recalled | Medline Industries | Packaging Defects | Nationwide |
| Device | CLASS II | Medline convenience kits labeled as: Open Heart Pack, Medline Kit recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Medline convenience kits labeled as: GYN Robotic CDS, Medline Kit recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Linvatec Genesys CrossFT Suture Anchor with Disposable Driver recalled | ConMed | Other | Nationwide |