CLASS II Device recall · Reported 26 Dec 2018 · FDA Enforcement Report

Carousel MAS1025-01, part of the Brain Surgical Accessories recalled

Mazor Robotics is recalling Carousel MAS1025-01, part of the Brain Surgical Accessories Intra-operative Kit (KIT0295). The recall was initiated on 26 Nov 2014 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesAffected products: all Carousels under Level 01 were affected and therefore are applicable to this issue. The related lots include: WP065611, WP066389, WP066411, WP066412, WP066420. Explanation of the coding system: a part number (e.g. MAS1025-01) is composed of a generic part number related to the part (e.g. MAS1025) and the level of the part (e.g. 01).
Quantity recalled5 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0609-2019
FDA event ID
81521
Recalling firm
Mazor Robotics, North Industrial Park
Classification
Class II
Status
Terminated
Recall initiated
26 Nov 2014
FDA classified
14 Dec 2018
Reported
26 Dec 2018
Type
Voluntary: Firm initiated
Customers notified by
E-Mail
Reason category
Other Reasons
Device type
Spinal Implants

Timeline

26 Nov 2014Recall initiated by company
14 Dec 2018Classified by FDA+1479 days
26 Dec 2018Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

In some situations the carousel position pin (locker) may cause the carousel upper plate to pop out and the carousel may then inadvertently move.

Product as listed by the FDA

Carousel MAS1025-01, part of the Brain Surgical Accessories Intra-operative Kit (KIT0295), accessory used with the Renaissance System (TPL0038). Indicated for precise positioning of surgical instruments or spinal implants during general spinal and brain surgery. It is intended to be used whenever the clinician and/or patient benefits from generated 3D imaging of, high contrast objects.

Where it was distributed

Affected devices were distributed in the US to four different medical facilities in four states

Questions about this recall

Why was it recalled?

In some situations the carousel position pin (locker) may cause the carousel upper plate to pop out and the carousel may then inadvertently move.

Which lots are affected?

Affected products: all Carousels under Level 01 were affected and therefore are applicable to this issue. The related lots include: WP065611, WP066389, WP066411, WP066412, WP066420. Explanation of the coding system: a part number (e.g. MAS1025-01) is composed of a generic part number related to the part (e.g. MAS1025) and the level of the part (e.g. 01).

Where was it sold?

Affected devices were distributed in the US to four different medical facilities in four states

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.