CLASS II Device recall · Reported 4 Jun 2014 · FDA Enforcement Report

The Centricity PACS Workstation is intended recalled by GE Healthcare

GE Healthcare is recalling The Centricity PACS Workstation is intended for use as a primary diagnostic and analysis, distributed nationwide in the US. The recall was initiated on 3 Apr 2014 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesCentricity PACS versions: 3.2; 3.2.0.1; 3.2.0.2; 3.2.1; 3.2.2; 3.2.2.1; 3.2.2.2; 3.2.2.3; 3.2.6; 3.2 SP7; 3.2 SP8; 4.0; 4.0.1; 4.0 SP3
Quantity recalled297 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1665-2014
FDA event ID
68092
Recalling firm
GE Healthcare, Barrington, IL
Classification
Class II
Status
Terminated
Recall initiated
3 Apr 2014
FDA classified
28 May 2014
Reported
4 Jun 2014
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

3 Apr 2014Recall initiated by company
28 May 2014Classified by FDA+55 days
4 Jun 2014Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

It has been identified that in a reject image workflow where multiple studies are simultaneously opened on the Centricity PACS RA1000 workstation and images are rejected from the active displayed study, images for the previously displayed study may also be rejected. The issue can occur with a previously displayed study that remains open in the background. If one or more, but not all, images in

Product as listed by the FDA

The Centricity PACS Workstation is intended for use as a primary diagnostic and analysis tool for diagnostic images by trained healthcare professional, including radiologists, physicians, technologists, clinicians, and nurses. It is also intended for use as a clinical review workstation throughout the healthcare facility. The workstation interface provides the user with a means to display, manipulate, archive, print, and export images when connected wit the Centricity PACS infrastructure.

Where it was distributed

Worldwide Distribution-USA (nationwide) including the states of AK, AL, AZ, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, VA, VT, WA, WI, and WV, and the countries of Australia, Canada, Chile, France, Germany, India, Ireland, Israel, Italy, Japan, Korea (South), Kuwait, Lebanon, Netherlands, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, Taiwan, Turkey, and United Kingdom.

Nationwide AlaskaAlabamaArizonaCaliforniaColoradoFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusettsMarylandMaineMichiganMinnesota

Questions about this recall

Why was it recalled?

It has been identified that in a reject image workflow where multiple studies are simultaneously opened on the Centricity PACS RA1000 workstation and images are rejected from the active displayed study, images for the previously displayed study may also be rejected. The issue can occur with a previously displayed study that remains open in the background. If one or more, but not all, images in

Which lots are affected?

Centricity PACS versions: 3.2; 3.2.0.1; 3.2.0.2; 3.2.1; 3.2.2; 3.2.2.1; 3.2.2.2; 3.2.2.3; 3.2.6; 3.2 SP7; 3.2 SP8; 4.0; 4.0.1; 4.0 SP3

Where was it sold?

Worldwide Distribution-USA (nationwide) including the states of AK, AL, AZ, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, VA, VT, WA, WI, and WV, and the countries of Australia, Canada, Chile, France, Germany, India, Ireland, Israel, Italy, Japan, Korea (South), Kuwait, Lebanon, Netherlands, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, Taiwan, Turkey, and United Kingdom.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from GE Healthcare

All 512

Other Test Kits & Diagnostics recalls

All 2,656