CLASS II ONGOING Device recall · Reported 5 Apr 2023 · FDA Enforcement Report

CRP Immunoturbidimetric reagent recalled by Randox Laboratories

Randox Laboratories is recalling CRP Immunoturbidimetric reagent, distributed in Ohio, Pennsylvania, West Virginia. The recall was initiated on 1 Feb 2023 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesGTIN: 05055273201765 Batch: 588434
Quantity recalled2 kits

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1286-2023
FDA event ID
91745
Recalling firm
Randox Laboratories, Kearneysville, WV
Classification
Class II
Status
Ongoing
Recall initiated
1 Feb 2023
FDA classified
24 Mar 2023
Reported
5 Apr 2023
Type
Voluntary: Firm initiated
Customers notified by
E-Mail
Reason category
Other Reasons

Timeline

1 Feb 2023Recall initiated by company
24 Mar 2023Classified by FDA+51 days
5 Apr 2023Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The Antibody (R2 reagent) is showing a positive bias compared to previous batches on patient samples. Quality Controls do not show the same bias and will be within range

Product as listed by the FDA

CRP Immunoturbidimetric reagent. Intended for the quantitative in vitro determination of CRP concentration in serum Catalogue number: CP3826

Where it was distributed

PA, OH, WV

OhioPennsylvaniaWest Virginia

Questions about this recall

Why was it recalled?

The Antibody (R2 reagent) is showing a positive bias compared to previous batches on patient samples. Quality Controls do not show the same bias and will be within range

Which lots are affected?

GTIN: 05055273201765 Batch: 588434

Where was it sold?

PA, OH, WV

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Randox Laboratories

All 98

Other Test Kits & Diagnostics recalls

All 2,663