CLASS II ONGOING Device recall · Reported 11 Nov 2026 · FDA Enforcement Report

CryoPatch SynerGraft (SG) Pulmonary Human Cardiac Patch recalled

Artivion is recalling CryoPatch SynerGraft (SG) Pulmonary Human Cardiac Patch, distributed nationwide in the US. The recall was initiated on 18 Sep 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSerial/Donor Number: 12617554/177871, UDI-DI: 00877234000478, Shipped: 22May2024, Implanted: 05June2024
Quantity recalled1

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0325-2027
FDA event ID
100012
Recalling firm
Artivion, Kennesaw, GA
Classification
Class II
Status
Ongoing
Recall initiated
18 Sep 2026
FDA classified
30 Oct 2026
Reported
11 Nov 2026
Type
Voluntary: Firm initiated
Reason category
Other Reasons

Timeline

18 Sep 2026Recall initiated by company
30 Oct 2026Classified by FDA+42 days
11 Nov 2026Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

A cryopreserved pulmonary human cardiac patch, taken from a 64-year-old human donor, was outside the cardiac tissue donor age specification of newborn to 60 years. The patch was implanted, and due to limitations of long-term in vivo data, the possibility of premature graft dysfunction cannot be excluded.

Product as listed by the FDA

CryoPatch SynerGraft (SG) Pulmonary Human Cardiac Patch, Model: SGP010

Where it was distributed

US Nationwide distribution in the state of IL.

Nationwide Illinois

Questions about this recall

Why was it recalled?

A cryopreserved pulmonary human cardiac patch, taken from a 64-year-old human donor, was outside the cardiac tissue donor age specification of newborn to 60 years. The patch was implanted, and due to limitations of long-term in vivo data, the possibility of premature graft dysfunction cannot be excluded.

Which lots are affected?

Serial/Donor Number: 12617554/177871, UDI-DI: 00877234000478, Shipped: 22May2024, Implanted: 05June2024

Where was it sold?

US Nationwide distribution in the state of IL.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 17 Nov 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.