CLASS II Device recall · Reported 15 Nov 2017 · FDA Enforcement Report
Delta Cup Beater Positioner - Aligner recalled by Limacorporate S.p.A
Limacorporate is recalling Delta Cup Beater Positioner - Aligner. The recall was initiated on 8 Sep 2017 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0068-2018
- FDA event ID
- 78109
- Recalling firm
- Limacorporate, San Daniele del Friuli, N/A
- Classification
- Class II
- Status
- Completed
- Recall initiated
- 8 Sep 2017
- FDA classified
- 6 Nov 2017
- Reported
- 15 Nov 2017
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Hip & Knee Implants
Timeline
| 8 Sep 2017 | Recall initiated by company | |
| 6 Nov 2017 | Classified by FDA | +59 days |
| 15 Nov 2017 | Published in enforcement report | +9 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
An internal analysis on complaints received indicated that the welding process may have been performed sub optimally in the 12 lot # involved , leading to reduced mechanical strength of the device and possible breakage during impaction when the instrument is subjected to repeated multi-axial stresses.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
Delta Cup Beater Positioner - Aligner, model # 9057.20.555 Used during implantation of the Delta Acetabular System
Where it was distributed
Worldwide Distribution - US Distribution and to the countries of : Australia, Austria, Belgium, Brazil, Czech Republic, Germany, Italy, New Zealand, Poland, Russia, Slovakia, Slovenia, Spain, South Africa, Switzerland, Turkey, UK and Germany
Questions about this recall
Why was it recalled?
An internal analysis on complaints received indicated that the welding process may have been performed sub optimally in the 12 lot # involved , leading to reduced mechanical strength of the device and possible breakage during impaction when the instrument is subjected to repeated multi-axial stresses.
Which lots are affected?
Lot # 15AG0HJ, 15AG0HQ, 15AG0UY, 15AG00P, 15AG040, 15AG08Y, 14AG243, 15AG00Q, 15AG0UZ, 15AG0V0, 15AG0HN, 16AG02J
Where was it sold?
Worldwide Distribution - US Distribution and to the countries of : Australia, Austria, Belgium, Brazil, Czech Republic, Germany, Italy, New Zealand, Poland, Russia, Slovakia, Slovenia, Spain, South Africa, Switzerland, Turkey, UK and Germany
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Limacorporate
All 16| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Prima Eccentrical Adaptor With Screw recalled by Limacorporate S.p.A | Limacorporate | Other | Nationwide |
| Device | CLASS II | Prima Eccentrical Adaptor With Screw recalled by Limacorporate S.p.A | Limacorporate | Other | Nationwide |
| Device | CLASS II | TEMA Elbow - Axle Small recalled by Limacorporate S.p.A | Limacorporate | Other | FL, TX |
| Device | CLASS II | Prima TT Glenoid Monoblock/Modular reverse TT baseplates recalled | Limacorporate | Potency & Assay | 6 states |
| Device | CLASS II | Physica HPS Tibial Liner #6 H10 recalled by Limacorporate S.p.A | Limacorporate | Packaging Defects | 5 states |
Other Hip & Knee Implants recalls
All 3,123| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Unity Knee Asc Modular Handle included recalled by Corin Medical | Corin Medical | Device Malfunction | Nationwide |
| Device | CLASS II | Halyard Anterior Hip Pack recalled by Avid Medical | Avid Medical | Sterility | Nationwide |
| Device | CLASS II | Halyard Anterior Hip Pack recalled by Avid Medical | Avid Medical | Sterility | Nationwide |
| Device | CLASS II | Medline medical convenience kits labeled as: Hip Scope Pack recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | TS Femoral Prep Tray, Triathlon recalled by Howmedica Osteonics | Howmedica Osteonics | Sterility | NJ |