CLASS II ONGOING Device recall · Reported 24 Jan 2024 · FDA Enforcement Report
Prima TT Glenoid Monoblock/Modular reverse TT baseplates recalled
Limacorporate is recalling Prima TT Glenoid Monoblock/Modular reverse TT baseplates: Description, distributed in California, Illinois, Indiana, Michigan, Missouri, Oklahoma. The recall was initiated on 29 Aug 2023 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0726-2024
- FDA event ID
- 93668
- Recalling firm
- Limacorporate, San Daniele Del Friuli, N/A
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 29 Aug 2023
- FDA classified
- 12 Jan 2024
- Reported
- 24 Jan 2024
- Type
- Voluntary: Firm initiated
- Customers notified by
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Reason category
- Potency & Assay
- Device type
- Other Medical Devices
Timeline
| 29 Aug 2023 | Recall initiated by company | |
| 12 Jan 2024 | Classified by FDA | +136 days |
| 24 Jan 2024 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Potential for baseplate peripheral holes to be out of specification due to a manufacturing issue, which could lead to loosening of the implant, or the screw moving from its original position.
Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.
Product as listed by the FDA
PRIMA TT Glenoid Monoblock/Modular reverse TT baseplates: Description/Product Code: BASEPLATE D.25MM REGULAR/1975.14.500, BASEPLATE D.28MM REGULAR/1975.14.800, BASEPLATE D.25MM FULL WEDGED 10-degrees/1975.14.510, BASEPLATE D.28MM FULL WEDGED 10-degrees/1975.14.810, BASEPLATE D.25MM FULL WEDGED 15-degrees/1975.14.615, BASEPLATE D.28MM FULL WEDGED 15-degrees/1975.14.815, BASEPLATE D.28MM FULL WEDGED 15-degrees X/1975.14.865, BASEPLATE D.25 MM FULL WEDGED 20-degrees/1975.14.620, BASEPLATE D.28 MM FULL WEDGED 20-degrees/1975.14.820, BASEPLATE D.28 MM FULL WEDGED 20-degrees X/1975.14.870
Where it was distributed
US: IL, MO, CA, IN, MI, OK
Questions about this recall
Why was it recalled?
Potential for baseplate peripheral holes to be out of specification due to a manufacturing issue, which could lead to loosening of the implant, or the screw moving from its original position.
Which lots are affected?
Product Code/UDI-DI: 1975.14.500/08033390244824, 1975.14.800/08033390244893, 1975.14.510/08033390231299, 1975.14.810/08033390244909, 1975.14.615/08033390244848, 1975.14.815/08033390244916, 1975.14.865/08033390244930, 1975.14.620/08033390231312, 1975.14.820/08033390244923, 1975.14.870/08033390244947
Where was it sold?
US: IL, MO, CA, IN, MI, OK
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Limacorporate
All 16| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Prima Eccentrical Adaptor With Screw recalled by Limacorporate S.p.A | Limacorporate | Other | Nationwide |
| Device | CLASS II | Prima Eccentrical Adaptor With Screw recalled by Limacorporate S.p.A | Limacorporate | Other | Nationwide |
| Device | CLASS II | TEMA Elbow - Axle Small recalled by Limacorporate S.p.A | Limacorporate | Other | FL, TX |
| Device | CLASS II | Physica HPS Tibial Liner #6 H10 recalled by Limacorporate S.p.A | Limacorporate | Packaging Defects | 5 states |
| Device | CLASS II | REF 1975.14.500, Prima TT Genoid Monoblock Reverse TT Baseplate recalled | Limacorporate | Potency & Assay | Nationwide |
Other Other Medical Devices recalls
All 13,986| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | ANNOUNCED | Luna Bilevel Positive Airway Pressure (BPAP) System recalledLuna | React Health | Other | VT |
| Device | ANNOUNCED | Beds, Rails, Mattresses, and accessories recalled by MedlineMedline | Medline | Other | |
| Device | CLASS II | Hemochron Jr., Whole Blood Microcoagulation System recalled | Accriva Diagnostics | Other | Nationwide |
| Device | CLASS II | Senographe Pristina 3D recalled by GE Medical Systems | GE Medical Systems | Other | WI |
| Device | CLASS II | Mobile column TruSystem recalled by Baxter Healthcare | Baxter Healthcare | Device Malfunction | Nationwide |