CLASS II Device recall · Reported 5 Feb 2014 · FDA Enforcement Report

DeRoyal (R) Aquarius Hot/Cold Therapy Combo Unit recalled

DeRoyal Industries is recalling DeRoyal (R) Aquarius Hot/Cold Therapy Combo Unit, w/ Knee/Shoulder Blnkt NS, distributed nationwide in the US. The recall was initiated on 17 Sep 2013 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesLot Numbers: 32901145, 33063999, 3132641, 3217556, 33276468, 33369411, 33412431, 33415413
Quantity recalled409 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0836-2014
FDA event ID
66742
Recalling firm
DeRoyal Industries, New Tazewell, TN
Classification
Class II
Status
Terminated
Recall initiated
17 Sep 2013
FDA classified
27 Jan 2014
Reported
5 Feb 2014
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Temperature Abuse

Timeline

17 Sep 2013Recall initiated by company
27 Jan 2014Classified by FDA+132 days
5 Feb 2014Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The unit displays an error code of hose kinking when no hose kink is present; the unit fails to monitor the temperature of the water in the blanket as intended and controlling it through the automatic temperature control. This failure could result in temperature injuries to the treatment site; and there have been two reports of electric shock.

Foods that need refrigeration and were stored or shipped too warm can grow harmful bacteria without any change in look or smell.

Product as listed by the FDA

DeRoyal (R) Aquarius Hot/Cold Therapy Combo Unit, w/ Knee/Shoulder Blnkt NS, REF T653NS, Rx only, Manufacturer DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849. Physical Therapy.

Where it was distributed

Nationwide Distribution.

Nationwide

Questions about this recall

Is DeRoyal (R) Aquarius Hot/Cold Therapy Combo Unit, w/ Knee/Shoulder Blnkt NS recalled?

Yes. DeRoyal Industries recalled DeRoyal (R) Aquarius Hot/Cold Therapy Combo Unit, w/ Knee/Shoulder Blnkt NS on 17 Sep 2013. The recall is Class II and its status is Terminated. Recall number Z-0836-2014.

Why was it recalled?

The unit displays an error code of hose kinking when no hose kink is present; the unit fails to monitor the temperature of the water in the blanket as intended and controlling it through the automatic temperature control. This failure could result in temperature injuries to the treatment site; and there have been two reports of electric shock.

Which lots are affected?

Lot Numbers: 32901145, 33063999, 3132641, 3217556, 33276468, 33369411, 33412431, 33415413

Where was it sold?

Nationwide Distribution.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from DeRoyal Industries

All 226

Other Hip & Knee Implants recalls

All 3,097