CLASS II ONGOING Device recall · Reported 28 Feb 2024 · FDA Enforcement Report

The device is an X-ray system that is capable recalled

Canon Medical System USA is recalling The device is an X-ray system that is capable of radiographic and fluoroscopic studies, distributed in California, Louisiana. The recall was initiated on 15 Nov 2023 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI: 04987670100192, Serial Number: B1E22X2005
Quantity recalled1

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1154-2024
FDA event ID
93844
Recalling firm
Canon Medical System USA, Tustin, CA
Classification
Class II
Status
Ongoing
Recall initiated
15 Nov 2023
FDA classified
16 Feb 2024
Reported
28 Feb 2024
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

15 Nov 2023Recall initiated by company
16 Feb 2024Classified by FDA+93 days
28 Feb 2024Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

VAREX changed the manufacturing of the internal board, and the detectors displayed abnormal images. Internal board of the X-ray flat panel detector may display abnormal images, such as horizontal stripes, which may lead to the inability to complete an ongoing exam, or the patient may need to be examined with another x-ray system.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

The device is an X-ray system that is capable of radiographic and fluoroscopic studies and is used in an interventional setting. The system consists of a C-arm which is equipped with an X-ray tube, beam limiter and X-ray receptor, X-ray controller, computers with system and processing software, and a patient radiographic table.

Where it was distributed

US: CA, LA

CaliforniaLouisiana

Questions about this recall

Why was it recalled?

VAREX changed the manufacturing of the internal board, and the detectors displayed abnormal images. Internal board of the X-ray flat panel detector may display abnormal images, such as horizontal stripes, which may lead to the inability to complete an ongoing exam, or the patient may need to be examined with another x-ray system.

Which lots are affected?

UDI-DI: 04987670100192, Serial Number: B1E22X2005

Where was it sold?

US: CA, LA

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

1 other product · FDA event 93844

This product is part of a recall event; the main page for the event is The Infx-8000F is designed to take advantage of the latest recalled.