CLASS II ONGOING Device recall · Reported 30 Sep 2026 · FDA Enforcement Report
Vitek 2 Ast-N461 Test Kit, Gram Negative Susceptibility recalled
Biomerieux is recalling Vitek 2 Ast-N461 Test Kit, Gram Negative Susceptibility Testing - Automated, Reference. The recall was initiated on 24 Jul 2026 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-3231-2026
- FDA event ID
- 99749
- Recalling firm
- Biomerieux, Hazelwood, MO
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 24 Jul 2026
- FDA classified
- 18 Sep 2026
- Reported
- 30 Sep 2026
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Test Kits & Diagnostics
Timeline
| 24 Jul 2026 | Recall initiated by company | |
| 18 Sep 2026 | Classified by FDA | +56 days |
| 30 Sep 2026 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
The issue is related to card pouch damage associated with a portion of the lot which was produced on a specific card pouching device, in a specific timeframe, causing a potential risk of false resistant results. False resistant results may lead to the use of less effective antibiotic treatments or the association of multiple antibiotics having negative side effects, which could potentially impact patient s health.
Product as listed by the FDA
VITEK 2 AST-N461 TEST KIT, GRAM NEGATIVE SUSCEPTIBILITY TESTING - AUTOMATED, Reference Number 424811
Where it was distributed
International distribution to the countries of Austria, Germany, Luxembourg.
Questions about this recall
Why was it recalled?
The issue is related to card pouch damage associated with a portion of the lot which was produced on a specific card pouching device, in a specific timeframe, causing a potential risk of false resistant results. False resistant results may lead to the use of less effective antibiotic treatments or the association of multiple antibiotics having negative side effects, which could potentially impact patient s health.
Which lots are affected?
UDI-DI 03573026637712; Lot Number 1333393504
Where was it sold?
International distribution to the countries of Austria, Germany, Luxembourg.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 6 Oct 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Biomerieux
All 477| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Vitek 2 Ast-N447 Test Kit Product reference recalled by Biomerieux | Biomerieux | Other | MO |
| Device | CLASS II | Vitek-negative Susceptibility Cards Containing o Polymyxin recalled | Biomerieux | Sterility | Nationwide |
| Device | CLASS II | Vitek¿-negative AST Test Kits (Cards) recalled by bioMerieux | bioMerieux | Sterility | NC |
| Device | CLASS I | Vitek 2 - AST Test kits (in vitro diagnostic test cards recalled | Biomerieux | Other | MO |
| Device | CLASS II | Myla comprises Ast Filters in conjunction with Vitek MS recalled | Biomerieux | Other | Nationwide |
Other Test Kits & Diagnostics recalls
All 2,658| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | DASH Rapid PCR Instrument recalled by Nuclein | Nuclein | Device Malfunction | Nationwide |
| Device | CLASS II | STA Staclot Protein recalled by Diagnostica Stago | Diagnostica Stago | Potency & Assay | Nationwide |
| Device | CLASS II | DxC 500i Clinical Analyzer Modules using recalled by Beckman Coulter | Beckman Coulter | Packaging Defects | Nationwide |
| Device | CLASS II | DxC 500 AU Clinical Analyzer Modules using recalled by Beckman Coulter | Beckman Coulter | Packaging Defects | Nationwide |
| Device | CLASS II | Syva EMIT 2000 Quinidine Assay recalled | Siemens Healthcare Diagnostics | Potency & Assay | Nationwide |