CLASS II ONGOING Device recall · Reported 8 Oct 2025 · FDA Enforcement Report
DxC 500 AU Module w/ISE, DxC 500i, With ISE recalled by Beckman Coulter
Beckman Coulter is recalling DxC 500 AU Module w/ISE, DxC 500i, With ISE. The recall was initiated on 4 Aug 2025 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0032-2026
- FDA event ID
- 97390
- Recalling firm
- Beckman Coulter, Brea, CA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 4 Aug 2025
- FDA classified
- 2 Oct 2025
- Reported
- 8 Oct 2025
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Potency & Assay
- Device type
- Test Kits & Diagnostics
Timeline
| 4 Aug 2025 | Recall initiated by company | |
| 2 Oct 2025 | Classified by FDA | +59 days |
| 8 Oct 2025 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
On a clinical analyzer, when ordering a Clinical Chemistry (CC) combination test simultaneously with an Immunoassay (IA) test, if the IA test is processed between the CC tests, CC sample status will remain In Progress with no errors, may cause processing delays.
Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.
Product as listed by the FDA
DxC 500 AU Module w/ISE, DxC 500i, With ISE, REF: C63522 and DxC 500 AU Module, DxC 500i, Without ISE, REF: C6352; part of the DxC 500i Clinical Analyzer
Where it was distributed
Internationally distribution to the countries of Poland, India, Dominican Republic, Lebanon, Italy, Brazil, El Salvador, Saudi Arabia, Germany, France, Nepal, Canada, Chile, Spain, Malaysia, Switzerland, Morocco, Philippines, Curacao, Qatar, United Arab Emirates, Hungary, Czechia, Oman, New Zealand, Thailand, Indonesia, Algeria, United Kingdom of Great Britain and Northern Ireland, Namibia, Turkey, Republic of Korea
Questions about this recall
Why was it recalled?
On a clinical analyzer, when ordering a Clinical Chemistry (CC) combination test simultaneously with an Immunoassay (IA) test, if the IA test is processed between the CC tests, CC sample status will remain In Progress with no errors, may cause processing delays.
Which lots are affected?
REF/UDI-DI: C63522/14987666545089, C63521/14987666545072. Software Versions: SW 1.3.3 and 1.3.4
Where was it sold?
Internationally distribution to the countries of Poland, India, Dominican Republic, Lebanon, Italy, Brazil, El Salvador, Saudi Arabia, Germany, France, Nepal, Canada, Chile, Spain, Malaysia, Switzerland, Morocco, Philippines, Curacao, Qatar, United Arab Emirates, Hungary, Czechia, Oman, New Zealand, Thailand, Indonesia, Algeria, United Kingdom of Great Britain and Northern Ireland, Namibia, Turkey, Republic of Korea
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Beckman Coulter
All 340| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Rotor Assy-SW41, Reference Number recalled by Beckman Coulter | Beckman Coulter | Device Malfunction | Nationwide |
| Device | CLASS II | Rotor Assy-SW40, Reference Number recalled by Beckman Coulter | Beckman Coulter | Device Malfunction | Nationwide |
| Device | CLASS II | IQ200 Sprint recalled by Beckman Coulter | Beckman Coulter | Packaging Defects | Nationwide |
| Device | CLASS II | DxU 840m Iris recalled by Beckman Coulter | Beckman Coulter | Packaging Defects | Nationwide |
| Device | CLASS II | DxC 500 AU Clinical Chemistry Analyzers: AU w/out ISE recalled | Beckman Coulter | Device Malfunction | Nationwide |
Other Test Kits & Diagnostics recalls
All 2,675| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Nextseq 550Dx Instrument recalled by Illumina | Illumina | Other | Nationwide |
| Device | CLASS II | NovaSeq 6000Dx Instrument recalled by Illumina | Illumina | Other | Nationwide |
| Device | CLASS II | Liaison Plex Gram-Negative Blood Culture Assay recalled by Luminex | Luminex | Potency & Assay | Nationwide |
| Device | CLASS II | MiSeq Dx Instrument recalled by Illumina | Illumina | Other | Nationwide |
| Device | CLASS II | Techno-Path Multichem AE Unassayed Level recalled | Techno-path Manufacturing | Potency & Assay | Nationwide |