CLASS II Device recall · Reported 11 Mar 2026 · FDA Enforcement Report
Endocavity ultrasonic probe recalled by Esaote S.p.a
Esaote is recalling Endocavity ultrasonic probe, distributed nationwide in the US. The recall was initiated on 14 Mar 2025 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1490-2026
- FDA event ID
- 98339
- Recalling firm
- Esaote, Sesto Fiorentino, N/A
- Classification
- Class II
- Status
- Completed
- Recall initiated
- 14 Mar 2025
- FDA classified
- 3 Mar 2026
- Reported
- 11 Mar 2026
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Packaging Defects
- Device type
- Other Medical Devices
Timeline
| 14 Mar 2025 | Recall initiated by company | |
| 3 Mar 2026 | Classified by FDA | +354 days |
| 11 Mar 2026 | Published in enforcement report | +8 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
A potential weakness has been identified in the probe body, so that, in a specific production batch, liquid may leak from the terminal part of the casing near the cable.
Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.
Product as listed by the FDA
Esaote Endocavity ultrasonic probe, Model E 3-12, REF 120000038
Where it was distributed
US: Nationwide distribution in the states of NC and TX.
Questions about this recall
Why was it recalled?
A potential weakness has been identified in the probe body, so that, in a specific production batch, liquid may leak from the terminal part of the casing near the cable.
Which lots are affected?
UDI 0805604453891, Serial Numbers: 241700533C, 241700529C, 241700607C, 241700536C, 241700613C.
Where was it sold?
US: Nationwide distribution in the states of NC and TX.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Esaote
All 9| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | AC/DC Power Supply recalled by Esaote S.p.a | Esaote | Other | Nationwide |
| Device | CLASS II | AC/DC Power Supply recalled by Esaote S.p.a | Esaote | Other | Nationwide |
| Device | CLASS II | MyLab Seven recalled by Esaote S.p.a | Esaote | Device Malfunction | Nationwide |
| Device | CLASS II | MyLab Alpha recalled by Esaote S.p.a | Esaote | Device Malfunction | Nationwide |
| Device | CLASS II | MyLab Gamma recalled by Esaote S.p.a | Esaote | Device Malfunction | Nationwide |
Other Other Medical Devices recalls
All 13,986| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | ANNOUNCED | Luna Bilevel Positive Airway Pressure (BPAP) System recalledLuna | React Health | Other | VT |
| Device | ANNOUNCED | Beds, Rails, Mattresses, and accessories recalled by MedlineMedline | Medline | Other | |
| Device | CLASS II | Hemochron Jr., Whole Blood Microcoagulation System recalled | Accriva Diagnostics | Other | Nationwide |
| Device | CLASS II | Senographe Pristina 3D recalled by GE Medical Systems | GE Medical Systems | Other | WI |
| Device | CLASS II | Mobile column TruSystem recalled by Baxter Healthcare | Baxter Healthcare | Device Malfunction | Nationwide |