CLASS II Device recall · Reported 11 Nov 2015 · FDA Enforcement Report
EnVision Flex/Hrp visualization reagent found recalled
Dako North America is recalling EnVision Flex/Hrp visualization reagent found in the following kits: EnVision Flex, High, distributed nationwide in the US. The recall was initiated on 4 Sep 2015 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0220-2016
- FDA event ID
- 72097
- Recalling firm
- Dako North America, Carpinteria, CA
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 4 Sep 2015
- FDA classified
- 5 Nov 2015
- Reported
- 11 Nov 2015
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Packaging Defects
- Device type
- Test Kits & Diagnostics
Timeline
| 4 Sep 2015 | Recall initiated by company | |
| 5 Nov 2015 | Classified by FDA | +62 days |
| 11 Nov 2015 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
There is a defect in one lot of a buffer, non-specific background staining that has, in some instances, interfered with interpretation of the staining results.
Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.
Product as listed by the FDA
EnVision FLEX/HRP visualization reagent found in the following kits: EnVision FLEX, High pH (Link) EnVision FLEX+, Mouse, High pH (Link) EnVision FLEX Mini Kit, High pH (Link) EnVision FLEX, High pH (Dako Omnis) Multi-component, sealed package. For in vitro diagnostic use.
Where it was distributed
Worldwide Distribution - US (nationwide) Distribution including Puerto Rico and to the states of : TX, SD, CA, FL, MA, MD, WV, NC, OH, DE, MI, NJ, CO, HI, AR, NY, LA, GA, NV, RI, IL, OK, CT, MO, VA, and DC., and to the countries of : Canada, Brazil and Costa Rica.
Nationwide ArkansasCaliforniaColoradoConnecticutDistrict of ColumbiaDelawareFloridaGeorgiaHawaiiIllinoisLouisianaMassachusettsMarylandMichiganMissouriNorth CarolinaNew JerseyNevadaNew YorkOhio
Questions about this recall
Is EnVision Flex/Hrp visualization reagent found in the following kits: EnVision Flex, High recalled?
Yes. Dako North America recalled EnVision Flex/Hrp visualization reagent found in the following kits: EnVision Flex, High on 4 Sep 2015. The recall is Class II and its status is Terminated. Recall number Z-0220-2016.
Why was it recalled?
There is a defect in one lot of a buffer, non-specific background staining that has, in some instances, interfered with interpretation of the staining results.
Which lots are affected?
Item No.: K800021-2, K800221-2, K802321-2 (Link) K801021-2, K801221-2, K802421-2 (AS/AS+) GV80011-2, GV82311-2 (Dako Omnis) Catalog/Model No.: K8000 K8002 K8023 GV800 SM802 EnVision FLEX / HRP Lot 20019374 used in: Envision FLEX, High pH (link) Lot No.: 20019097, 20020777, 20021381 Envision FLEX+, Mouse, High pH (link) Lot No.: 20019103, 20019113, 20020772 EnVision FLEX Mini Kit, High pH (Link) Lot No.: 20021362 DM 842 EnVision FLEX / HRP (Dako Omnis) Lot 20019013 used in: EnVision FLEX, High pH (Dako Omnis) Lot No.: 20019102
Where was it sold?
Worldwide Distribution - US (nationwide) Distribution including Puerto Rico and to the states of : TX, SD, CA, FL, MA, MD, WV, NC, OH, DE, MI, NJ, CO, HI, AR, NY, LA, GA, NV, RI, IL, OK, CT, MO, VA, and DC., and to the countries of : Canada, Brazil and Costa Rica.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Dako North America
All 8| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | PD-L1 IHC 22C3 pharmDx is a companion diagnostic used to recalled | Dako North America | Other | 29 states |
| Device | CLASS II | Dako HER2 CISH pharmDx Kit, product recalled by Dako North America | Dako North America | Other | Nationwide |
| Device | CLASS II | Test Request Distributor (TRD 3 and TRD 4) recalled | Dako North America | Device Malfunction | Nationwide |
| Device | CLASS II | Autostainer Plus Link Instrument (AS100) recalled by Dako North America | Dako North America | Packaging Defects | Nationwide |
| Device | CLASS II | Dako Autostainer Link recalled by Dako North America | Dako North America | Device Malfunction | Nationwide |
Other Test Kits & Diagnostics recalls
All 2,656| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | STA Staclot Protein S recalled by Diagnostica Stago | Diagnostica Stago | Potency & Assay | Nationwide |
| Device | CLASS II | DxC 500i Clinical Analyzer Modules using recalled by Beckman Coulter | Beckman Coulter | Packaging Defects | Nationwide |
| Device | CLASS II | DxC 500 AU Clinical Analyzer Modules using recalled by Beckman Coulter | Beckman Coulter | Packaging Defects | Nationwide |
| Device | CLASS II | Syva EMIT 2000 Quinidine Assay recalled | Siemens Healthcare Diagnostics | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Beckman Coulter, Emit 2000 Quinidine Assay recalled | Siemens Healthcare Diagnostics | Lead & Heavy Metals | Nationwide |