CLASS II Device recall · Reported 6 May 2015 · FDA Enforcement Report

Test Request Distributor (TRD 3 and TRD 4) recalled

Dako North America is recalling Test Request Distributor (TRD 3 and TRD 4), a, distributed nationwide in the US. The recall was initiated on 1 Apr 2015 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesN/A Serial Number: 1311004352, 1308004862, 1307000537, 1405002266, 1410008844, 1307006995, 1311001075, 1402002186, 1401010308, 1402005995, 1306001697, 1311007317, 1410004066, 1404006200, 1410004067, 1405002263, 1404007225, 1410002144, 1311007321, 1404004226, 1401005960, 1305006010, 1401000919, 1405000371, 1401005963, 1403005465, 1304008205, 1401003711, 1401001048, 1310010425, 1401003715, 1311004354, 1311004355
Quantity recalled33 (7 in U.S. and 26 Internationally)

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1548-2015
FDA event ID
70898
Recalling firm
Dako North America, Carpinteria, CA
Classification
Class II
Status
Terminated
Recall initiated
1 Apr 2015
FDA classified
29 Apr 2015
Reported
6 May 2015
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

1 Apr 2015Recall initiated by company
29 Apr 2015Classified by FDA+28 days
6 May 2015Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

If a user requests slides from the LIS or TPID, then updates a request by changing the test, the TRD software will reject the update. The LIS and printed slide label will display the updated test with the requested change, but the Omnis system will execute the initial test, not reflecting the change. The system will not warn the user that the test request was rejected.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Test Request Distributor (TRD 1.3 and TRD 1.4), a software module used on the Dako Omnis System, an automated slide stainer for in vitro diagnostic use. The TRD is intended to distribute patient case, appropriate parts information, track changes, and transform test requests from LIS or manual entries to connected Dako systems.

Where it was distributed

Worldwide Distribution - U.S Nationwide in the states of IN and PA and the countries of: Spain, Belgium, France, Netherlands, Denmark, Finland, Sweden, Germany, UK, and Korea.

Nationwide Pennsylvania

Questions about this recall

Is Test Request Distributor (TRD 3 and TRD 4), a recalled?

Yes. Dako North America recalled Test Request Distributor (TRD 3 and TRD 4), a on 1 Apr 2015. The recall is Class II and its status is Terminated. Recall number Z-1548-2015.

Why was it recalled?

If a user requests slides from the LIS or TPID, then updates a request by changing the test, the TRD software will reject the update. The LIS and printed slide label will display the updated test with the requested change, but the Omnis system will execute the initial test, not reflecting the change. The system will not warn the user that the test request was rejected.

Which lots are affected?

N/A Serial Number: 1311004352, 1308004862, 1307000537, 1405002266, 1410008844, 1307006995, 1311001075, 1402002186, 1401010308, 1402005995, 1306001697, 1311007317, 1410004066, 1404006200, 1410004067, 1405002263, 1404007225, 1410002144, 1311007321, 1404004226, 1401005960, 1305006010, 1401000919, 1405000371, 1401005963, 1403005465, 1304008205, 1401003711, 1401001048, 1310010425, 1401003715, 1311004354, 1311004355

Where was it sold?

Worldwide Distribution - U.S Nationwide in the states of IN and PA and the countries of: Spain, Belgium, France, Netherlands, Denmark, Finland, Sweden, Germany, UK, and Korea.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.