CLASS II ONGOING Device recall · Reported 12 Aug 2026 · FDA Enforcement Report
Flow Anesthesia Machine Flow-i C20 6677200 Flow-i C30 recalled
Maquet Critical Care AB is recalling Flow Anesthesia Machine Flow-i C20 6677200 Flow-i C30 6677300 Flow-i C40 6677400 Flow-i, distributed nationwide in the US. The recall was initiated on 24 Jun 2026 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2860-2026
- FDA event ID
- 99359
- Recalling firm
- Maquet Critical Care AB, Solna
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 24 Jun 2026
- FDA classified
- 3 Aug 2026
- Reported
- 12 Aug 2026
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Other Medical Devices
Timeline
| 24 Jun 2026 | Recall initiated by company | |
| 3 Aug 2026 | Classified by FDA | +40 days |
| 12 Aug 2026 | Published in enforcement report | +9 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Events observed where anesthetic agent was not delivered when the vaporizer was activated. Failure to deliver the anesthetic agent may result in the patient receiving insufficient anesthesia and may lead to intraoperative awareness.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
Flow Anesthesia Machine Flow-i C20 6677200 Flow-i C30 6677300 Flow-i C40 6677400 Flow-i C20 6888520 Flow-i C30 6888530 Flow-i C40 6888540 Flow-c 6887700 Flow-e 6887900
Where it was distributed
Worldwide - US Nationwide distribution to the states of AL, AR, CA, CO, DC, DE, FL, GA, IN, MA, MI, MO, NY, OH, OR, PA, SC, TX, VA, WA, WV.
Nationwide AlabamaArkansasCaliforniaColoradoDistrict of ColumbiaDelawareFloridaGeorgiaIndianaMassachusettsMichiganMissouriNew YorkOhioOregonPennsylvaniaSouth CarolinaTexasVirginiaWashington
Questions about this recall
Why was it recalled?
Events observed where anesthetic agent was not delivered when the vaporizer was activated. Failure to deliver the anesthetic agent may result in the patient receiving insufficient anesthesia and may lead to intraoperative awareness.
Which lots are affected?
Flow-i C20 6677200 / UDI-DI 07325710001349 Flow-i C30 6677300 / UDI-DI 07325710001363 Flow-i C40 6677400 / UDI-DI 07325710001318 Flow-i C20 6888520 / UDI-DI 07325710010617 Flow-i C30 6888530 / UDI-DI 07325710010624 Flow-i C40 6888540 / UDI-DI 07325710010631 Flow-c 6887700 / UDI-DI 07325710009765 Flow-e 6887900 / UDI-DI 07325710010457 Software versions 4.8, 4.10, and 4.11.
Where was it sold?
Worldwide - US Nationwide distribution to the states of AL, AR, CA, CO, DC, DE, FL, GA, IN, MA, MI, MO, NY, OH, OR, PA, SC, TX, VA, WA, WV.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Maquet Critical Care AB
All 5| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Getinge Y Sensor Module, Servo-u, Servo recalled | Maquet Critical Care AB | Device Malfunction | Nationwide |
| Device | CLASS I | Servo-u MR Ventilator System recalled by Maquet Critical Care AB | Maquet Critical Care AB | Labeling Errors | Nationwide |
| Device | CLASS I | Servo-n Ventilator System recalled by Maquet Critical Care AB | Maquet Critical Care AB | Labeling Errors | Nationwide |
| Device | CLASS I | Servo-u Ventilator System recalled by Maquet Critical Care AB | Maquet Critical Care AB | Labeling Errors | Nationwide |
Other Other Medical Devices recalls
All 13,857| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline Convenience Kits: Major Pack recalled by Medline Industries | Medline Industries | Other | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Peds Sinus Pack recalled by Medline Industries | Medline Industries | Other | Nationwide |
| Device | CLASS II | SmartPath to dStream for XR and 3.0T recalled by Philips North America | Philips North America | Device Malfunction | Nationwide |
| Device | CLASS II | Achieva 1.5T, Achieva 1.5T Conversion recalled by Philips North America | Philips North America | Device Malfunction | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Mfc Craniotomy Pack recalled | Medline Industries | Other | Nationwide |