CLASS I ONGOING Device recall · Reported 9 Jul 2025 · FDA Enforcement Report

Servo-n Ventilator System recalled by Maquet Critical Care AB

Maquet Critical Care AB is recalling Servo-n Ventilator System, distributed nationwide in the US. The recall was initiated on 30 May 2025 and the FDA has classified it as Class I.

Check your package

Lot, batch or date codesModel Number: 6694800. UDI-DI: 07325710001110. All software versions.
Quantity recalled7,688 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2014-2025
FDA event ID
97032
Recalling firm
Maquet Critical Care AB, Solna
Classification
Class I
Status
Ongoing
Recall initiated
30 May 2025
FDA classified
27 Jun 2025
Reported
9 Jul 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Labeling Errors

Timeline

30 May 2025Recall initiated by company
27 Jun 2025Classified by FDA+28 days
9 Jul 2025Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Potential for inaccurate measurement of the patient circuit compliance during the patient circuit test due to combination of incorrect pre-use leakage test and use of an incorrect active humidifier.

Product as listed by the FDA

Servo-n Ventilator System. Model Number: 6694800.

Where it was distributed

US Nationwide distribution and Puerto Rico. No international distribution.

Nationwide Puerto Rico

Questions about this recall

Why was it recalled?

Potential for inaccurate measurement of the patient circuit compliance during the patient circuit test due to combination of incorrect pre-use leakage test and use of an incorrect active humidifier.

Which lots are affected?

Model Number: 6694800. UDI-DI: 07325710001110. All software versions.

Where was it sold?

US Nationwide distribution and Puerto Rico. No international distribution.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class I mean?

Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

2 other products · FDA event 97032

This product is part of a recall event; the main page for the event is Servo-u Ventilator System recalled by Maquet Critical Care AB.

More recalls from Maquet Critical Care AB

All 5

Other CPAP & Ventilators recalls

All 554