CLASS II ONGOING Device recall · Reported 2 Oct 2024 · FDA Enforcement Report

GE Giraffe Warmer recalled by Wipro GE Healthcare Private

Wipro GE Healthcare Private is recalling GE Giraffe Warmer. The recall was initiated on 9 Aug 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSerial Number GBW24230381SA
Quantity recalled1

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-3080-2024
FDA event ID
95216
Recalling firm
Wipro GE Healthcare Private, Bengaluru
Classification
Class II
Status
Ongoing
Recall initiated
9 Aug 2024
FDA classified
20 Sep 2024
Reported
2 Oct 2024
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Packaging Defects

Timeline

9 Aug 2024Recall initiated by company
20 Sep 2024Classified by FDA+42 days
2 Oct 2024Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

GE HealthCare has become aware that complete electrical safety testing was not conducted during manufacturing of certain Giraffe and Panda iRes Warmers. There are multiple electrical safety protections included in the product design. However, because the electrical safety testing in manufacturing was incomplete for these devices, there is the potential for leakage current to exceed IEC 60601 limits.

Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.

Product as listed by the FDA

GE Giraffe Warmer, Model/Catalog Number M1118179-01090979. Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a controlled open environment.

Where it was distributed

US and Australia, Denmark, France, Germany, Ireland, ITALY, New Zealand, Poland, Portugal, Slovenia, Spain, SWITZERLAND, United Arab Emirates, United Kingdom.

Questions about this recall

Why was it recalled?

GE HealthCare has become aware that complete electrical safety testing was not conducted during manufacturing of certain Giraffe and Panda iRes Warmers. There are multiple electrical safety protections included in the product design. However, because the electrical safety testing in manufacturing was incomplete for these devices, there is the potential for leakage current to exceed IEC 60601 limits.

Which lots are affected?

Serial Number GBW24230381SA

Where was it sold?

US and Australia, Denmark, France, Germany, Ireland, ITALY, New Zealand, Poland, Portugal, Slovenia, Spain, SWITZERLAND, United Arab Emirates, United Kingdom.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

1 other product · FDA event 95216

This product is part of a recall event; the main page for the event is GE Panda iRES Warmer recalled by Wipro GE Healthcare Private.

More recalls from Wipro GE Healthcare Private

All 10

Other Other Medical Devices recalls

All 13,908