CLASS II ONGOING Device recall · Reported 15 Jan 2025 · FDA Enforcement Report

GE Healthcare Optima XR220amx, Mobile Digital Radiographic System recalled

GE Medical Systems is recalling GE Healthcare Optima XR220amx, Mobile Digital Radiographic System. The recall was initiated on 25 Nov 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codes1) 5555000-12: UDI/DI none, Serial Numbers: 082407100532; UDI/DI to be provided, Serial Numbers: 082407160512; 2) 5555000-5: UDI/DI 840682115261, Serial Numbers: 1557XR0205, 352867XRA, 478765CDP1, 509962KV220, 601358OPTX220, 620421P3, 813872OPT3, SB2201700099WK; UDI/DI none, Serial Numbers: 479314220, 00028MOB12, 203694SB3, 205820DP1, 208666KDM, 209334OPT1, 210703M3, 240566PX2, 240964OPT1, 248465DP3, 256HHPORT2, 256HHPORT3, 256MHPORT1, 281MSL220, 281MSL2202, 305KEN220A, 306766DMOB2, 310680PORTDR2, 314251NAV7, 325574OPTITBD2, 330489DIGPORT, 333OPTI2, 333OPTI3, 333OPTI4, 352596OPT1, 352LEEDP1, 352VILDP1, 386FHODP3, 401OPTI1, 405636P2, 405636P240, 405752OPT1, 405752OPT2, 405752OPT4, 405DELCITY220, 407303OPTIMAX220, 408259AP6, 417347DPORT1, 417556P1, 417556P2, 417556P3, 419251DP1, 419251DP2, 419383DP1, 434654OPTIMA, 435CVXR220, 480MAYO220A, 501364OPTI1, 501364OPTI2, 509474DR1, 510437OPT1, 510727OPT3, 512SAMCSB, 513870DP1, 520TMCOPT2, 541426XR220, 559447OPT1, 559784OPT2, 561694OPT1, 561694OPT2, 561965DPORT5, 561965DPORT7, 573882DP1, 573882DP2, 573882DP3, 580421OPT1, 601288SB1, 601288SB2, 601288SB3, 601288SB6, 601928D220, 608364OPMX220, 613938XR3, 626457GEPORT1, 703369MDIG2, 703664M7, 706596SFP2, 706651DP, 715838DP2, 718206OPT240C, 718267DP5, 718670PTDR1, 720UCHSB7, 727848SB1, 731541SB1, 731541SB1A, 731541SB2, 731541SB3, 732324PA1, 732828DP11, 757388DP4, 770385220D1, 772468OPT1, 785776AMX220P1, 804281SMDM1, 806296OPTI220, 813302OPT3, 814837OPT, 815300M220C, 817927JPSXR220G, 817MCAXR220B, 828213DXP1, 828262P2202, 830RHHM1, 830RHHM2, 843847220B, 850325DP2, 863699OP220, 903886XR220, 905527XR220, 905845MOB4, 908685DP2, 909464220P1, 910615OPTIMA1, 910615OPTIMA2, 915521OPTIMA1, 915595OPT1, 940898XR220A, 940898XR240D, 985867M1, AXXXXX38, ORMCPORTA, ORMCPORTB, ORMCPORTC, ORMCPORTD, TAMXDRP03, WD0005, XR1036454WK5; 3) 5555000-6: UDI/DI 840682115261, Serial Numbers: 06051MOB06, 228809XR220, 228818XR220, 270798XR220B, 281KINGOPT3, 303450SB3, 303695SB3, 416323MOB3, 530888OPT1, 540981ODP3R, 540981ODP9, 615769OPTX220, 713529MDR220, 717782OPT4, 7982XR0192, 802775SB1, 813681OPT2, 918307OPT2, MHSWOPT2201, SA2513MM03, SB2201600071WK, SB2201700085WK, SJRO220P1, SJRO220P3, UTAHJAZZ, XR1864CO; UDI/DI none, Serial Numbers: 083007001584712, 0850070865, 0850070866, 0850070870, 0850070913, 0856070254, 0856070267, 0856070268, 091026DEMO2245, 10181MOB16, 140024RX14, 206598USB7, 209943OTP1, 214PH220G, 228575D220, 254HHSMOB1, 256429PORT2, 270688SB5, 281338BAPOR1, 281338CL221, 281338MDR220, 281348KWDM2, 281348KWDM3, 281KINGOPT1, 281MWB2202, 281MWB2203, 304255OPT220, 304623OPT220, 308865GSDP1, 309779RDIG2, 314525NAV15, 314525NAV16, 318675PW1, 318675PW2, 318769RDPT1, 318769RDPT2, 330480DP1, 330480DP2, 330480DP3, 330480DP4, 370OPTI1, 409938DR1220, 412232DMX2, 412372DMX1, 412647DMX2, 419291DP4, 419893DP1, 423493DP2, 432685OPT2, 434799GE220, 469764XR220H, 504454XR221, 504785MD1, 504873TMD8, 513732DP1, 563742NDIG1, 574267XR220, 586263DP1, 586493DP1, 601798D220, 614257DP4, 615284STMDR2, 615284STMDR3, 615316DP1, 615342DP1, 615391SUR, 702SIEXR221, 702SIEXR226, 713359BSDM3, 713440DP1, 713441OPCDX1, 714966GEPORT2, 720UCHSB9, 740592OPTIMA2, 77522MOB14, 801479SB2, 810342DP2, 817255NHXR220A, 817HUG220C, 817PLAZA220A, 817SAGERXR220A, 865541DP2, 865545DP1, 865545DP3, 903CSM222, 904OPMCDP3, 914737PORTWK3, 915577XR220, 916UCDAMXDR23, 916UCDAMXDR27, 918579OPT2, 919774DIGPRT, 928757XR222, 936539DM1, 936539DM2, 949517DR1, 949764DR2, 949764DR3, 949764DR4, 972420XR220B, 972420XR220C, 972566XR220F, 979MGXR220, LLPM220P1, MHKOPT220, SA1056MM06, SA2161MM04, XR1031657WK8, ZA1067RX30; 4) A0659PB: UDI/DI 840682115261, Serial Numbers: 702880EROP1; UDI/DI none, Serial Numbers: 206598USB6; 5) to be provided: UDI/DI none, Serial Numbers: 240826OPT1, 772SRDP1, 817HUG220B 860545SB2A; UDI/DI to be provided, Serial Numbers: 907852OPT22019, SA1075MM04.
Quantity recalled281 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0812-2025
FDA event ID
95958
Recalling firm
GE Medical Systems, Waukesha, WI
Classification
Class II
Status
Ongoing
Recall initiated
25 Nov 2024
FDA classified
7 Jan 2025
Reported
15 Jan 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

25 Nov 2024Recall initiated by company
7 Jan 2025Classified by FDA+43 days
15 Jan 2025Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement performed, some bolts used in the Column Safety mechanism can be missing. If these bolts are missing and a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work properly causing the arm with X-ray tube and collimator to drop downward on the column track.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

GE Healthcare Optima XR220amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-12; 2) 5555000-5; 3) 5555000-6; 4) A0659PB; 5) to be provided

Where it was distributed

Worldwide

Questions about this recall

Why was it recalled?

GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement performed, some bolts used in the Column Safety mechanism can be missing. If these bolts are missing and a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work properly causing the arm with X-ray tube and collimator to drop downward on the column track.

Which lots are affected?

1) 5555000-12: UDI/DI none, Serial Numbers: 082407100532; UDI/DI to be provided, Serial Numbers: 082407160512; 2) 5555000-5: UDI/DI 840682115261, Serial Numbers: 1557XR0205, 352867XRA, 478765CDP1, 509962KV220, 601358OPTX220, 620421P3, 813872OPT3, SB2201700099WK; UDI/DI none, Serial Numbers: 479314220, 00028MOB12, 203694SB3, 205820DP1, 208666KDM, 209334OPT1, 210703M3, 240566PX2, 240964OPT1, 248465DP3, 256HHPORT2, 256HHPORT3, 256MHPORT1, 281MSL220, 281MSL2202, 305KEN220A, 306766DMOB2, 310680PORTDR2, 314251NAV7, 325574OPTITBD2, 330489DIGPORT, 333OPTI2, 333OPTI3, 333OPTI4, 352596OPT1, 352LEEDP1, 352VILDP1, 386FHODP3, 401OPTI1, 405636P2, 405636P240, 405752OPT1, 405752OPT2, 405752OPT4, 405DELCITY220, 407303OPTIMAX220, 408259AP6, 417347DPORT1, 417556P1, 417556P2, 417556P3, 419251DP1, 419251DP2, 419383DP1, 434654OPTIMA, 435CVXR220, 480MAYO220A, 501364OPTI1, 501364OPTI2, 509474DR1, 510437OPT1, 510727OPT3, 512SAMCSB, 513870DP1, 520TMCOPT2, 541426XR220, 559447OPT1, 559784OPT2, 561694OPT1, 561694OPT2, 561965DPORT5, 561965DPORT7, 573882DP1, 573882DP2, 573882DP3, 580421OPT1, 601288SB1, 601288SB2, 601288SB3, 601288SB6, 601928D220, 608364OPMX220, 613938XR3, 626457GEPORT1, 703369MDIG2, 703664M7, 706596SFP2, 706651DP, 715838DP2, 718206OPT240C, 718267DP5, 718670PTDR1, 720UCHSB7, 727848SB1, 731541SB1, 731541SB1A, 731541SB2, 731541SB3, 732324PA1, 732828DP11, 757388DP4, 770385220D1, 772468OPT1, 785776AMX220P1, 804281SMDM1, 806296OPTI220, 813302OPT3, 814837OPT, 815300M220C, 817927JPSXR220G, 817MCAXR220B, 828213DXP1, 828262P2202, 830RHHM1, 830RHHM2, 843847220B, 850325DP2, 863699OP220, 903886XR220, 905527XR220, 905845MOB4, 908685DP2, 909464220P1, 910615OPTIMA1, 910615OPTIMA2, 915521OPTIMA1, 915595OPT1, 940898XR220A, 940898XR240D, 985867M1, AXXXXX38, ORMCPORTA, ORMCPORTB, ORMCPORTC, ORMCPORTD, TAMXDRP03, WD0005, XR1036454WK5; 3) 5555000-6: UDI/DI 840682115261, Serial Numbers: 06051MOB06, 228809XR220, 228818XR220, 270798XR220B, 281KINGOPT3, 303450SB3, 303695SB3, 416323MOB3, 530888OPT1, 540981ODP3R, 540981ODP9, 615769OPTX220, 713529MDR220, 717782OPT4, 7982XR0192, 802775SB1, 813681OPT2, 918307OPT2, MHSWOPT2201, SA2513MM03, SB2201600071WK, SB2201700085WK, SJRO220P1, SJRO220P3, UTAHJAZZ, XR1864CO; UDI/DI none, Serial Numbers: 083007001584712, 0850070865, 0850070866, 0850070870, 0850070913, 0856070254, 0856070267, 0856070268, 091026DEMO2245, 10181MOB16, 140024RX14, 206598USB7, 209943OTP1, 214PH220G, 228575D220, 254HHSMOB1, 256429PORT2, 270688SB5, 281338BAPOR1, 281338CL221, 281338MDR220, 281348KWDM2, 281348KWDM3, 281KINGOPT1, 281MWB2202, 281MWB2203, 304255OPT220, 304623OPT220, 308865GSDP1, 309779RDIG2, 314525NAV15, 314525NAV16, 318675PW1, 318675PW2, 318769RDPT1, 318769RDPT2, 330480DP1, 330480DP2, 330480DP3, 330480DP4, 370OPTI1, 409938DR1220, 412232DMX2, 412372DMX1, 412647DMX2, 419291DP4, 419893DP1, 423493DP2, 432685OPT2, 434799GE220, 469764XR220H, 504454XR221, 504785MD1, 504873TMD8, 513732DP1, 563742NDIG1, 574267XR220, 586263DP1, 586493DP1, 601798D220, 614257DP4, 615284STMDR2, 615284STMDR3, 615316DP1, 615342DP1, 615391SUR, 702SIEXR221, 702SIEXR226, 713359BSDM3, 713440DP1, 713441OPCDX1, 714966GEPORT2, 720UCHSB9, 740592OPTIMA2, 77522MOB14, 801479SB2, 810342DP2, 817255NHXR220A, 817HUG220C, 817PLAZA220A, 817SAGERXR220A, 865541DP2, 865545DP1, 865545DP3, 903CSM222, 904OPMCDP3, 914737PORTWK3, 915577XR220, 916UCDAMXDR23, 916UCDAMXDR27, 918579OPT2, 919774DIGPRT, 928757XR222, 936539DM1, 936539DM2, 949517DR1, 949764DR2, 949764DR3, 949764DR4, 972420XR220B, 972420XR220C, 972566XR220F, 979MGXR220, LLPM220P1, MHKOPT220, SA1056MM06, SA2161MM04, XR1031657WK8, ZA1067RX30; 4) A0659PB: UDI/DI 840682115261, Serial Numbers: 702880EROP1; UDI/DI none, Serial Numbers: 206598USB6; 5) to be provided: UDI/DI none, Serial Numbers: 240826OPT1, 772SRDP1, 817HUG220B 860545SB2A; UDI/DI to be provided, Serial Numbers: 907852OPT22019, SA1075MM04.

Where was it sold?

Worldwide

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

2 other products · FDA event 95958

More recalls from GE Medical Systems

All 160

Other Imaging Equipment recalls

All 2,336