CLASS II ONGOING Device recall · Reported 19 Mar 2025 · FDA Enforcement Report
GE Healthcare Revolution Apex Select, System, X-ray recalled
GE Medical Systems is recalling GE Healthcare Revolution Apex Select, System, X-ray, Tomography, Computed. The recall was initiated on 18 Feb 2025 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1301-2025
- FDA event ID
- 96370
- Recalling firm
- GE Medical Systems, Waukesha, WI
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 18 Feb 2025
- FDA classified
- 7 Mar 2025
- Reported
- 19 Mar 2025
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Packaging Defects
- Device type
- Imaging Equipment
Timeline
| 18 Feb 2025 | Recall initiated by company | |
| 7 Mar 2025 | Classified by FDA | +17 days |
| 19 Mar 2025 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.
Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.
Product as listed by the FDA
GE Healthcare Revolution Apex Select, System, X-ray, Tomography, Computed
Where it was distributed
Worldwide distribution.
Questions about this recall
Why was it recalled?
GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.
Which lots are affected?
System ID Number 907729APEX1, UDI (01)00195278460271(11)231100(21)REV2X2300082CN, Mfg. Lot/Serial Number REV2X2300082CN
Where was it sold?
Worldwide distribution.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other products in this recall event
6 other products · FDA event 96370This product is part of a recall event; the main page for the event is GE Healthcare Revolution Apex, System, X-ray, Tomography, Computed recalled.
More recalls from GE Medical Systems
All 160| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Optima XR240amx, Mobile Digital Radiographic System recalled | GE Medical Systems | Device Malfunction | Nationwide |
| Device | CLASS II | Optima XR220amx, Mobile Digital-Ready Radiographic System recalled | GE Medical Systems | Device Malfunction | Nationwide |
| Device | CLASS II | Optima XR200amx, Mobile Digital-Ready Radiographic System recalled | GE Medical Systems | Device Malfunction | Nationwide |
| Device | CLASS II | Giraffe OmniBed Field Replaceable Unit (FRU) recalled | GE Medical Systems | Other | Nationwide |
| Device | CLASS II | Allia IGS 7 Pulse angiographic X-ray system recalled | GE Medical Systems | Other | Nationwide |
Other Imaging Equipment recalls
All 2,336| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline Convenience Kits containing BD ChloraPrep Clear recalled | Medline Industries | Packaging Defects | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Ultrasound Breast Pack recalled | Medline Industries | Sterility | Nationwide |
| Device | CLASS II | Alphenix Models Infx-8000C Infx-8000F Infx-8000V This device recalled | Canon Medical System USA | Other | Nationwide |
| Device | CLASS II | Merge PACS and Merge OrthoPACS recalled by Merge Healthcare | Merge Healthcare | Labeling Errors | Nationwide |
| Device | CLASS II | Philips Azurion recalled by Philips Medical Systems | Philips Medical Systems Nederl | Potency & Assay |