CLASS II Device recall · Reported 6 May 2015 · FDA Enforcement Report
Genesis Malleable Penile Prosthesis recalled
Coloplast Manufacturing US is recalling Genesis Malleable Penile Prosthesis, distributed nationwide in the US. The recall was initiated on 9 Apr 2015 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1538-2015
- FDA event ID
- 70960
- Recalling firm
- Coloplast Manufacturing US, Minneapolis, MN
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 9 Apr 2015
- FDA classified
- 27 Apr 2015
- Reported
- 6 May 2015
- Type
- Voluntary: Firm initiated
- Customers notified by
- Visit
- Reason category
- Labeling Errors
- Device type
- Other Medical Devices
Timeline
| 9 Apr 2015 | Recall initiated by company | |
| 27 Apr 2015 | Classified by FDA | +18 days |
| 6 May 2015 | Published in enforcement report | +9 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Several lots of the 9.5 mm Genesis Malleable Penile Prosthesis were mislabeled. The Retail Box and Patient Labels indicate a size 9 mm, instead of 9.5 mm. The device size is printed directly on the individual prosthesis tray during the packaging process and states the correct size, 9.5 mm.
Product as listed by the FDA
Genesis Malleable Penile Prosthesis, 9.5 mm, Catalog /Item Number 91-9509SC/519250. Product Usage: The Coloplast Genesis Penile prosthesis is a flexible silicone elastomer device designed to be implanted into the penis for the management of erectile dysfunction.
Where it was distributed
Worldwide - US Nationwide Distribution in the states of FL, NY, TX and the country of FRANCE
Questions about this recall
Why was it recalled?
Several lots of the 9.5 mm Genesis Malleable Penile Prosthesis were mislabeled. The Retail Box and Patient Labels indicate a size 9 mm, instead of 9.5 mm. The device size is printed directly on the individual prosthesis tray during the packaging process and states the correct size, 9.5 mm.
Which lots are affected?
4336612, 4364589 , 4413811, 4413812, 4336613, 4413810.
Where was it sold?
Worldwide - US Nationwide Distribution in the states of FL, NY, TX and the country of FRANCE
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Coloplast Manufacturing US
All 28| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Folysil Silicone Foley Catheters recalled by Coloplast Manufacturing US | Coloplast Manufacturing US | Sterility | Nationwide |
| Device | CLASS II | X-Flow prostatectomy catheter recalled by Coloplast Manufacturing US | Coloplast Manufacturing US | Sterility | Nationwide |
| Device | CLASS II | Ureteral dilator recalled by Coloplast Manufacturing US | Coloplast Manufacturing US | Sterility | 9 states |
| Device | CLASS II | Titan NB Infra Zero Ang recalled by Coloplast Manufacturing US | Coloplast Manufacturing US | Device Malfunction | Nationwide |
| Device | CLASS II | Titan Touch Pump recalled by Coloplast Manufacturing US | Coloplast Manufacturing US | Device Malfunction | Nationwide |
Other Other Medical Devices recalls
All 13,986| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | ANNOUNCED | Luna Bilevel Positive Airway Pressure (BPAP) System recalledLuna | React Health | Other | VT |
| Device | ANNOUNCED | Beds, Rails, Mattresses, and accessories recalled by MedlineMedline | Medline | Other | |
| Device | CLASS II | Hemochron Jr., Whole Blood Microcoagulation System recalled | Accriva Diagnostics | Other | Nationwide |
| Device | CLASS II | Senographe Pristina 3D recalled by GE Medical Systems | GE Medical Systems | Other | WI |
| Device | CLASS II | Mobile column TruSystem recalled by Baxter Healthcare | Baxter Healthcare | Device Malfunction | Nationwide |