CLASS II ONGOING Device recall · Reported 19 Nov 2025 · FDA Enforcement Report

GM85 Digital Mobile X-ray imaging System recalled by NeuroLogica

NeuroLogica is recalling GM85 Digital Mobile X-ray imaging System, distributed nationwide in the US. The recall was initiated on 7 Oct 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesModel Number: GM85; UDI-DI: 08806088582412; Serial Numbers: 514GM3IK400001L, 5143M3IK300001X, 5143M3IJ100010M, 5143M3IHC00014H, 5143M3IK300008R, 5143M3JJ400007E, 5143M3HJ900002V, 5143M3IK400004F, 513YM3IK500003Z, 5143M3HK600005V, 5143M3JJ200010J, 5143M3HK600008M, 5143M3HK200004H, 5143M3JK500003W, 5143M3IJ200002F, 5143M3IHC00007A, 5143M3JJ300001K, 5143M3IJ600014H, 5143M3JK500006X, 5143M3JJ200017Y, 5143M3JK400014E, 5143M3IJ600013B, 5143M3GK500004A, 5143M3IHC00011V, 5143M3JK500001M, 5143M3JJ300012N, 5143M3GK500005M, 5143M3IK400007W, 5143M3JK400008N, 5143M3JJ300011Z, 5143M3JK100003Z, 5143M3JK500007L, 5143M3IK500001F, 5143M3IHC00002L, 5143M3IK300006T, 5143M3IHC00010R, 5143M3IK600004K, 5143M3IJ300003E, 5143M3IHC00003Y, 5143M3JJ400006K, 5143M3HJ900001L, 5143M3IJ600003L, 5143M3IJC00001Y, 5143M3IJ600001H, 5143M3HJ500003A, 5149M3IK400003E, 5143M3IJ800008L, 5143M3JJ700002R, 5143M3HK300002V, 5143M3JJ200008B, 513YM3IK500002R, 5143M3JK100005D, 5143M3IJ600004F, 5143M3IHC00005J, 5143M3JJC00004Y, 5143M3HK600003R, 5143M3JK500004D, 5143M3HK600004A, 5143M3HK300012L, 5143M3HK300008M, 5143M3JJ200015M, 5143M3HK300007N, 5143M3JK100007F, 5143M3IHC00001H, 5143M3IJ400006V, 5143M3IJ100002P, 514GM3HK100003J, 5143M3HK600010W, 5143M3HJ500001Y, 5143M3JK400010W, 5143M3IJ300005Z, 5143M3HK400010M, 5149M3IK400002K, 5143M3IK400002J, 5143M3GK500007W, 5143M3IK400009H, 5143M3IJ300002Y, 5143M3JK100004A, 513YM3IK400001W, 5143M3HK600011N, 5143M3HK300003Y, 5143M3JJ200012Z, 5143M3JK400007B, 5143M3HJA00002K, 5149M3IK400001H, 5143M3IJ100005R, 5143M3IHC00012Z, 5143M3HK400009W, 514GM3HK100002R, 5143M3IJ100001Y, 5143M3HK700002K, 5143M3IJ400004B, 5143M3JK400004M, 5143M3HJ900006H, 5143M3HK300005W, 5143M3JJ700001V, 5143M3IJ600015P, 5143M3JJ500002M, 5143M3HJ900003W, 5143M3IK600001Y, 5143M3JK100015B, 5143M3HJ600003K, 5143M3IJ200001D, 5143M3RJ100001M, 5143M3HK600009X, 5143M3IK600003N, 5143M3JJ300007T, 5143M3IJ400003R, 5143M3IK600002W, 5143M3HK200006K, 5143M3HJ900007R, 5143M3JK100002B, 5143M3IJ300006N, 5143M3IHC00017N, 5143M3IK300009L, 5143M3IJ500001B, 5143M3JJC00003L, 5143M3HJ500004M, 5143M3JK100006N, 5143M3HK700001D, 514GM3JK300001A, 5143M3IJ800005X, 5143M3HK400003P, 5143M3IK400005T, 5143M3HK300009K, 5143M3IJ600012X, 5143M3IHC00015L, 5143M3HK300010P, 5143M3RJ100003T, 5143M3DHB00001M, 5143M3IJ100003X, 5143M3IJ600006W, 5143M3IK400008A, 5143M3IHC00019T, 5143M3JK100014L, 513YM3IK400002A, 513YM3IK500005Y, 5143M3JK500002P, 5143M3HK400004J, 5143M3HJ500002J, 5143M3IJ400005N, 5143M3JJ400002X, 5143M3JJ200014F, 5143M3JJ700004K, 5143M3HK300017E, 5143M3IJ800011E, 5143M3HK300013A, 5143M3HK600007Z, 5143M3HK300004E, 5143M3GK500003J, 5143M3GK500001H, 5143M3GK500002Y, 5143M3GK500006P, 5143M3HK600002T, 5143M3HK600001D, 5143M3JK400013K, 5143M3HK400006F, 5143M3HK400007T, 5143M3HK400005Z, 5143M3HK400008X, 5143M3JK500005V, 513YM3IK500001K, 513YM3IK500004H, 5143M3HK300019Z, 5143M3JK400011A, 5143M3JK400003D, 5143M3JK400006R, 5143M3JK400009V, 5143M3JK400012H, 5143M3IK400003Z, 5143M3IK400006X, 5143M3HK300016Y, 5143M3IK300005H, 5143M3IK300007P, 5143M3HK300006Z, 5143M3HK300001F, 5143M3IK300002J, 5143M3HK200005V, 513YM3IK100002D, 513YM3IK100003N, 5143M3JK100013W, 514GM3HK100001E, 514GM3HK100004W, 5143M3JK100009M, 5143M3JK100008H, 5143M3JJC00006J, 5143M3JJC00002H, 5143M3HJA00010W, 5143M3HJA00008X, 513YM3IJA00001X, 5143M3HJA00007D, 5143M3HJA00001B, 5143M3HJ900010Z, 5143M3HJ900009E, 5143M3IJ800012D, 5143M3IJ800014B, 5143M3IJ800009M, 5143M3IJ800013Y, 514GM3JJ900001R, 5143M3IJ800007K, 5143M3IJ800001B, 5143M3IJ800002V, 5143M3IJ800003J, 5143M3IJ800004F, 5143M3JJ700003D, 5143M3JJ700005W, 5143M3HJ600005D, 5143M3IJ600009J, 5143M3IJ600008K, 5143M3IJ600007N, 5143M3IJ600005Y, 5143M3IJ600011M, 5143M3HJ500006W, 5143M3IJ600002P, 5143M3IJ400007P, 5143M3JJ400005H, 5143M3IJ300004W, 5143M3JJ300009R, 5143M3JJ300004B, 5143M3JJ300005D, 5143M3JJ200018X, 5143M3JJ300002X, 5143M3JJ300003J, 5143M3JJ200013D, 5143M3JJ200011L, 5143M3JJ200002M, 5143M3JJ200005X, 5143M3JJ200009A, 5143M3JJ200006R, 5143M3RJ100002V, 5143M3IJ100011V, 5143M3IJ100004E, 5143M3IHC00016Y, 5143M3IHC00009M, 5143M3IHC00008E, 5143M3IHC00006T;
Quantity recalled236 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0480-2026
FDA event ID
97721
Recalling firm
NeuroLogica, Danvers, MA
Classification
Class II
Status
Ongoing
Recall initiated
7 Oct 2025
FDA classified
7 Nov 2025
Reported
19 Nov 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Foreign Material

Timeline

7 Oct 2025Recall initiated by company
7 Nov 2025Classified by FDA+31 days
19 Nov 2025Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The four (4) M5 bolts that secure the Tube Head Unit to the end of the arm may loosen or fracture due to metal fatigue.

Foreign material recalls cover pieces of metal, plastic, glass, wood or other objects that got into the product during processing and could cause injury if eaten.

Product as listed by the FDA

GM85 Digital Mobile X-ray imaging System; Model Number: GM85;

Where it was distributed

Domestic: US Nationwide Distribution. International: Australia, Belgium, Canada, Chile, China Mainland, Czech Republic, Denmark, Ecuador, Finland, France, Germany, GHANA, Hungary, India, Italy, Korea, Malaysia, Mexico, Netherlands, New Zealand, Norway, Peru, Philippines, Poland, Portugal, Saudi Arabia, Serbia, Singapore, Spain, Sweden, Switzerland, Thailand, T¿rkiye, United Kingdom, Utd.Arab Emir., Vietnam,

Nationwide

Questions about this recall

Why was it recalled?

The four (4) M5 bolts that secure the Tube Head Unit to the end of the arm may loosen or fracture due to metal fatigue.

Which lots are affected?

Model Number: GM85; UDI-DI: 08806088582412; Serial Numbers: 514GM3IK400001L, 5143M3IK300001X, 5143M3IJ100010M, 5143M3IHC00014H, 5143M3IK300008R, 5143M3JJ400007E, 5143M3HJ900002V, 5143M3IK400004F, 513YM3IK500003Z, 5143M3HK600005V, 5143M3JJ200010J, 5143M3HK600008M, 5143M3HK200004H, 5143M3JK500003W, 5143M3IJ200002F, 5143M3IHC00007A, 5143M3JJ300001K, 5143M3IJ600014H, 5143M3JK500006X, 5143M3JJ200017Y, 5143M3JK400014E, 5143M3IJ600013B, 5143M3GK500004A, 5143M3IHC00011V, 5143M3JK500001M, 5143M3JJ300012N, 5143M3GK500005M, 5143M3IK400007W, 5143M3JK400008N, 5143M3JJ300011Z, 5143M3JK100003Z, 5143M3JK500007L, 5143M3IK500001F, 5143M3IHC00002L, 5143M3IK300006T, 5143M3IHC00010R, 5143M3IK600004K, 5143M3IJ300003E, 5143M3IHC00003Y, 5143M3JJ400006K, 5143M3HJ900001L, 5143M3IJ600003L, 5143M3IJC00001Y, 5143M3IJ600001H, 5143M3HJ500003A, 5149M3IK400003E, 5143M3IJ800008L, 5143M3JJ700002R, 5143M3HK300002V, 5143M3JJ200008B, 513YM3IK500002R, 5143M3JK100005D, 5143M3IJ600004F, 5143M3IHC00005J, 5143M3JJC00004Y, 5143M3HK600003R, 5143M3JK500004D, 5143M3HK600004A, 5143M3HK300012L, 5143M3HK300008M, 5143M3JJ200015M, 5143M3HK300007N, 5143M3JK100007F, 5143M3IHC00001H, 5143M3IJ400006V, 5143M3IJ100002P, 514GM3HK100003J, 5143M3HK600010W, 5143M3HJ500001Y, 5143M3JK400010W, 5143M3IJ300005Z, 5143M3HK400010M, 5149M3IK400002K, 5143M3IK400002J, 5143M3GK500007W, 5143M3IK400009H, 5143M3IJ300002Y, 5143M3JK100004A, 513YM3IK400001W, 5143M3HK600011N, 5143M3HK300003Y, 5143M3JJ200012Z, 5143M3JK400007B, 5143M3HJA00002K, 5149M3IK400001H, 5143M3IJ100005R, 5143M3IHC00012Z, 5143M3HK400009W, 514GM3HK100002R, 5143M3IJ100001Y, 5143M3HK700002K, 5143M3IJ400004B, 5143M3JK400004M, 5143M3HJ900006H, 5143M3HK300005W, 5143M3JJ700001V, 5143M3IJ600015P, 5143M3JJ500002M, 5143M3HJ900003W, 5143M3IK600001Y, 5143M3JK100015B, 5143M3HJ600003K, 5143M3IJ200001D, 5143M3RJ100001M, 5143M3HK600009X, 5143M3IK600003N, 5143M3JJ300007T, 5143M3IJ400003R, 5143M3IK600002W, 5143M3HK200006K, 5143M3HJ900007R, 5143M3JK100002B, 5143M3IJ300006N, 5143M3IHC00017N, 5143M3IK300009L, 5143M3IJ500001B, 5143M3JJC00003L, 5143M3HJ500004M, 5143M3JK100006N, 5143M3HK700001D, 514GM3JK300001A, 5143M3IJ800005X, 5143M3HK400003P, 5143M3IK400005T, 5143M3HK300009K, 5143M3IJ600012X, 5143M3IHC00015L, 5143M3HK300010P, 5143M3RJ100003T, 5143M3DHB00001M, 5143M3IJ100003X, 5143M3IJ600006W, 5143M3IK400008A, 5143M3IHC00019T, 5143M3JK100014L, 513YM3IK400002A, 513YM3IK500005Y, 5143M3JK500002P, 5143M3HK400004J, 5143M3HJ500002J, 5143M3IJ400005N, 5143M3JJ400002X, 5143M3JJ200014F, 5143M3JJ700004K, 5143M3HK300017E, 5143M3IJ800011E, 5143M3HK300013A, 5143M3HK600007Z, 5143M3HK300004E, 5143M3GK500003J, 5143M3GK500001H, 5143M3GK500002Y, 5143M3GK500006P, 5143M3HK600002T, 5143M3HK600001D, 5143M3JK400013K, 5143M3HK400006F, 5143M3HK400007T, 5143M3HK400005Z, 5143M3HK400008X, 5143M3JK500005V, 513YM3IK500001K, 513YM3IK500004H, 5143M3HK300019Z, 5143M3JK400011A, 5143M3JK400003D, 5143M3JK400006R, 5143M3JK400009V, 5143M3JK400012H, 5143M3IK400003Z, 5143M3IK400006X, 5143M3HK300016Y, 5143M3IK300005H, 5143M3IK300007P, 5143M3HK300006Z, 5143M3HK300001F, 5143M3IK300002J, 5143M3HK200005V, 513YM3IK100002D, 513YM3IK100003N, 5143M3JK100013W, 514GM3HK100001E, 514GM3HK100004W, 5143M3JK100009M, 5143M3JK100008H, 5143M3JJC00006J, 5143M3JJC00002H, 5143M3HJA00010W, 5143M3HJA00008X, 513YM3IJA00001X, 5143M3HJA00007D, 5143M3HJA00001B, 5143M3HJ900010Z, 5143M3HJ900009E, 5143M3IJ800012D, 5143M3IJ800014B, 5143M3IJ800009M, 5143M3IJ800013Y, 514GM3JJ900001R, 5143M3IJ800007K, 5143M3IJ800001B, 5143M3IJ800002V, 5143M3IJ800003J, 5143M3IJ800004F, 5143M3JJ700003D, 5143M3JJ700005W, 5143M3HJ600005D, 5143M3IJ600009J, 5143M3IJ600008K, 5143M3IJ600007N, 5143M3IJ600005Y, 5143M3IJ600011M, 5143M3HJ500006W, 5143M3IJ600002P, 5143M3IJ400007P, 5143M3JJ400005H, 5143M3IJ300004W, 5143M3JJ300009R, 5143M3JJ300004B, 5143M3JJ300005D, 5143M3JJ200018X, 5143M3JJ300002X, 5143M3JJ300003J, 5143M3JJ200013D, 5143M3JJ200011L, 5143M3JJ200002M, 5143M3JJ200005X, 5143M3JJ200009A, 5143M3JJ200006R, 5143M3RJ100002V, 5143M3IJ100011V, 5143M3IJ100004E, 5143M3IHC00016Y, 5143M3IHC00009M, 5143M3IHC00008E, 5143M3IHC00006T;

Where was it sold?

Domestic: US Nationwide Distribution. International: Australia, Belgium, Canada, Chile, China Mainland, Czech Republic, Denmark, Ecuador, Finland, France, Germany, GHANA, Hungary, India, Italy, Korea, Malaysia, Mexico, Netherlands, New Zealand, Norway, Peru, Philippines, Poland, Portugal, Saudi Arabia, Serbia, Singapore, Spain, Sweden, Switzerland, Thailand, T¿rkiye, United Kingdom, Utd.Arab Emir., Vietnam,

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.