CLASS II ONGOING Device recall · Reported 12 Jul 2023 · FDA Enforcement Report

OmniTom/OmniTom Elite- X-ray computed tomography recalled by NeuroLogica

NeuroLogica is recalling OmniTom/OmniTom Elite- X-ray computed tomography applications for anatomy that can be, distributed nationwide in the US. The recall was initiated on 6 Jun 2023 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesGTIN: 10815411020335 10815411020663 Serial Numbers: 10 14 23 25 26 36 38 40 41 42 44 46 48 49 50 51 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72 73 74 75 76 77 78 79 80 81 82 83 15R
Quantity recalled48 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2068-2023
FDA event ID
92506
Recalling firm
NeuroLogica, Danvers, MA
Classification
Class II
Status
Ongoing
Recall initiated
6 Jun 2023
FDA classified
3 Jul 2023
Reported
12 Jul 2023
Type
Voluntary: Firm initiated
Customers notified by
E-Mail
Reason category
Other Reasons

Timeline

6 Jun 2023Recall initiated by company
3 Jul 2023Classified by FDA+27 days
12 Jul 2023Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

1. Wheels loosening over time with use of the device could lead to the inability to effectively drive the device and could lead to delay in patient treatment. 2.Batteries used to power the NL5000 device contain contaminated boards, which may impact the ability to power on the device, this could lead to delay for patient treatment

Product as listed by the FDA

OmniTom/OmniTom Elite- X-ray computed tomography applications for anatomy that can be imaged in the 40 cm aperture, primarily head and neck Model Number: NL5000

Where it was distributed

Worldwide distribution - US Nationwide and the countries of Australia, Canada, Grand Bahama, India, Italy, Korea, Portugal, Saudi, Scotland, Thailand.

Nationwide

Questions about this recall

Why was it recalled?

1. Wheels loosening over time with use of the device could lead to the inability to effectively drive the device and could lead to delay in patient treatment. 2.Batteries used to power the NL5000 device contain contaminated boards, which may impact the ability to power on the device, this could lead to delay for patient treatment

Which lots are affected?

GTIN: 10815411020335 10815411020663 Serial Numbers: 10 14 23 25 26 36 38 40 41 42 44 46 48 49 50 51 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72 73 74 75 76 77 78 79 80 81 82 83 15R

Where was it sold?

Worldwide distribution - US Nationwide and the countries of Australia, Canada, Grand Bahama, India, Italy, Korea, Portugal, Saudi, Scotland, Thailand.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from NeuroLogica

All 13

Other Imaging Equipment recalls

All 2,349