CLASS II ONGOING Device recall · Reported 12 Oct 2022 · FDA Enforcement Report

Hammertoe Continuous Compression Implant (CCI) Kit recalled

Synthes (USA) Products is recalling Hammertoe Continuous Compression Implant (CCI) Kit, distributed nationwide in the US. The recall was initiated on 30 Aug 2022 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI: 00810633022047 Lot Numbers: MHC220119, MHC220120 Expiry Date: 1 June 2027
Quantity recalled191 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0008-2023
FDA event ID
90842
Recalling firm
Synthes (USA) Products, West Chester, PA
Classification
Class II
Status
Ongoing
Recall initiated
30 Aug 2022
FDA classified
6 Oct 2022
Reported
12 Oct 2022
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Foreign Material

Timeline

30 Aug 2022Recall initiated by company
6 Oct 2022Classified by FDA+37 days
12 Oct 2022Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Drill Pin (drill bit) is not compatible with the Drill Template included in the kit due to a packaging error, result in a surgical delay and may cause soft tissue damage if a larger incision is required to accommodate an alternative form of fixation. Additionally, there is a potential for adverse tissue reaction if the surgeon attempts to use the guide regardless of tight fit and plastic debris is created and ends up in the wound

Foreign material recalls cover pieces of metal, plastic, glass, wood or other objects that got into the product during processing and could cause injury if eaten.

Product as listed by the FDA

Hammertoe Continuous Compression Implant (CCI) Kit Part Number: 46.239.001

Where it was distributed

US Nationwide distribution.

Nationwide

Questions about this recall

Why was it recalled?

Drill Pin (drill bit) is not compatible with the Drill Template included in the kit due to a packaging error, result in a surgical delay and may cause soft tissue damage if a larger incision is required to accommodate an alternative form of fixation. Additionally, there is a potential for adverse tissue reaction if the surgeon attempts to use the guide regardless of tight fit and plastic debris is created and ends up in the wound

Which lots are affected?

UDI-DI: 00810633022047 Lot Numbers: MHC220119, MHC220120 Expiry Date: 1 June 2027

Where was it sold?

US Nationwide distribution.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Synthes (USA) Products