CLASS II Device recall · Reported 29 Jan 2020 · FDA Enforcement Report

Henry Schein CardioChek Starter Kit recalled

Polymer Technology Systems is recalling Henry Schein CardioChek Starter Kit, Reference Number 4060 HS, 570-1036, distributed nationwide in the US. The recall was initiated on 2 Dec 2019 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesAll Serial Numbers shipped during the 11/1/2016 to 10/25/2019 timeframe. UDI: 00304040040539
Quantity recalled44,448

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0834-2020
FDA event ID
84274
Recalling firm
Polymer Technology Systems, Indianapolis, IN
Classification
Class II
Status
Terminated
Recall initiated
2 Dec 2019
FDA classified
21 Jan 2020
Reported
29 Jan 2020
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

2 Dec 2019Recall initiated by company
21 Jan 2020Classified by FDA+50 days
29 Jan 2020Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

One brand of batteries have a small dimensional difference in the negative button terminal, which, when the batteries are inserted incorrectly, can create the possibility for the battery terminals to make reverse contact. This reverse contact can cause a potential for overheating.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Henry Schein CardioChek Starter Kit, Reference Number 4060 HS, 570-1036

Where it was distributed

The products were distributed US Nationwide. The products were distributed to the following foreign countries: Australia, Austria, Belgium, Bermuda, Canada, China, Chile, Costa Rica, Cayman Islands, Czech Republic, England, Finland, France, Germany, Greece, Hong Kong, India, Iraq, Italy, Jordan, Lebanon, Lithuania, Mexico, Morocco, Panama, Philippines, Poland, Qatar, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Trinidad, UAE.

Nationwide

Questions about this recall

Why was it recalled?

One brand of batteries have a small dimensional difference in the negative button terminal, which, when the batteries are inserted incorrectly, can create the possibility for the battery terminals to make reverse contact. This reverse contact can cause a potential for overheating.

Which lots are affected?

All Serial Numbers shipped during the 11/1/2016 to 10/25/2019 timeframe. UDI: 00304040040539

Where was it sold?

The products were distributed US Nationwide. The products were distributed to the following foreign countries: Australia, Austria, Belgium, Bermuda, Canada, China, Chile, Costa Rica, Cayman Islands, Czech Republic, England, Finland, France, Germany, Greece, Hong Kong, India, Iraq, Italy, Jordan, Lebanon, Lithuania, Mexico, Morocco, Panama, Philippines, Poland, Qatar, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Trinidad, UAE.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

3 other products · FDA event 84274

This product is part of a recall event; the main page for the event is PTS Diagnostics CardioChek Plus Professional Analyzer recalled.

More recalls from Polymer Technology Systems

All 21

Other Other Medical Devices recalls

All 13,857