CLASS II ONGOING Device recall · Reported 10 Jul 2024 · FDA Enforcement Report

Horiba custom configured fluorescence instrument recalled

Horiba Instruments is recalling Horiba custom configured fluorescence instrument, modular Fluorolog-QM, distributed nationwide in the US. The recall was initiated on 28 Mar 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesModular Fluorolog-QM
Quantity recalled4 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2166-2024
FDA event ID
94839
Recalling firm
Horiba Instruments, Piscataway, NJ
Classification
Class II
Status
Ongoing
Recall initiated
28 Mar 2024
FDA classified
1 Jul 2024
Reported
10 Jul 2024
Type
FDA Mandated
Customers notified by
Letter
Reason category
Foreign Material

Timeline

28 Mar 2024Recall initiated by company
1 Jul 2024Classified by FDA+95 days
10 Jul 2024Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

HORIBA custom configured fluorescence instrument has a laser safety defect. This custom configuration of the modular Instrument removed the primary light source and the excitation monochromator. A plastic cap was installed to plug the optical port, where a more durable beam stop should have been used. The defect may allow the class 4 laser beam to burn (or transmit) through the plastic cap to exit the protective enclosure.

Foreign material recalls cover pieces of metal, plastic, glass, wood or other objects that got into the product during processing and could cause injury if eaten.

Product as listed by the FDA

HORIBA custom configured fluorescence instrument, modular Fluorolog-QM

Where it was distributed

US Nationwide and Worldwide Distribution

Nationwide

Questions about this recall

Why was it recalled?

HORIBA custom configured fluorescence instrument has a laser safety defect. This custom configuration of the modular Instrument removed the primary light source and the excitation monochromator. A plastic cap was installed to plug the optical port, where a more durable beam stop should have been used. The defect may allow the class 4 laser beam to burn (or transmit) through the plastic cap to exit the protective enclosure.

Which lots are affected?

Modular Fluorolog-QM

Where was it sold?

US Nationwide and Worldwide Distribution

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.