CLASS II Device recall · Reported 4 Jul 2018 · FDA Enforcement Report

ABX Pentra Micro Albumin 2 CP, Serum/Plasma recalled

Horiba Instruments is recalling ABX Pentra Micro Albumin 2 CP, Serum/Plasma, distributed nationwide in the US. The recall was initiated on 26 Apr 2018 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesAll lot numbers; Software Version(s): Application versions below (V5.14 & V2.12) for Pentra 400 and Versions below (V1.07) for Pentra C400
Quantity recalled686 devices (233 devices US, 453 devices OUS)

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2286-2018
FDA event ID
80022
Recalling firm
Horiba Instruments, Irvine, CA
Classification
Class II
Status
Terminated
Recall initiated
26 Apr 2018
FDA classified
25 Jun 2018
Reported
4 Jul 2018
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

26 Apr 2018Recall initiated by company
25 Jun 2018Classified by FDA+60 days
4 Jul 2018Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

When the ABX Pentra Micro ALBUMIN 2 CP (ALBT2) Reagent 2 is sampled and dispensed immediately after an ABX Pentra Creatinine 120 CP (CREA3) Reagent 2 dispense, the final result of the microalbumin may be elevated more than 10%. This increase may affect serum, plasma and urine sample types.

Product as listed by the FDA

ABX Pentra Micro ALBUMIN 2 CP, Serum/Plasma, REF 1300032563 The device is intended for use in conjunction with certain materials to measure a variety of analytes.

Where it was distributed

Worldwide Distribution: US (nationwide) and to countries of: Bulgaria, Burma, Cameroon, Djibouti, France, Gabon, Germany, Italy, Lebanon, Mali, Mauritania, Poland, Portugal, Romania, Saudi Arabia, South Africa and Thailand.

Nationwide

Questions about this recall

Why was it recalled?

When the ABX Pentra Micro ALBUMIN 2 CP (ALBT2) Reagent 2 is sampled and dispensed immediately after an ABX Pentra Creatinine 120 CP (CREA3) Reagent 2 dispense, the final result of the microalbumin may be elevated more than 10%. This increase may affect serum, plasma and urine sample types.

Which lots are affected?

All lot numbers; Software Version(s): Application versions below (V5.14 & V2.12) for Pentra 400 and Versions below (V1.07) for Pentra C400

Where was it sold?

Worldwide Distribution: US (nationwide) and to countries of: Bulgaria, Burma, Cameroon, Djibouti, France, Gabon, Germany, Italy, Lebanon, Mali, Mauritania, Poland, Portugal, Romania, Saudi Arabia, South Africa and Thailand.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.