CLASS II Device recall · Reported 7 Feb 2018 · FDA Enforcement Report

Reagent Cup, B1037307, Sap 1221037307 for use recalled

Horiba Instruments is recalling Reagent Cup, B1037307, Sap 1221037307 for use with Abx Pentra 400 / Pentra C400, distributed nationwide in the US. The recall was initiated on 21 Nov 2017 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesDATE - MVT 25.08.2017
Quantity recalled10 bags of 100 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0494-2018
FDA event ID
78692
Recalling firm
Horiba Instruments, Irvine, CA
Classification
Class II
Status
Terminated
Recall initiated
21 Nov 2017
FDA classified
31 Jan 2018
Reported
7 Feb 2018
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

21 Nov 2017Recall initiated by company
31 Jan 2018Classified by FDA+71 days
7 Feb 2018Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The ABX PENTRA Reagent Containers do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent volume being pipetted and could lead to incorrect results being reported without an alarm.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

15 ML REAGENT CUP, B1037307, SAP 1221037307 for use with ABX PENTRA 400 / PENTRA C400

Where it was distributed

Nationwide

Nationwide

Questions about this recall

Why was it recalled?

The ABX PENTRA Reagent Containers do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent volume being pipetted and could lead to incorrect results being reported without an alarm.

Which lots are affected?

DATE - MVT 25.08.2017

Where was it sold?

Nationwide

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

1 other product · FDA event 78692

This product is part of a recall event; the main page for the event is Reagent Cup, B1034626, Sap 1221034626 for use recalled.

ReportedClassRecallCompanyReasonWhere
CLASS IIReagent Cup, B1034626, Sap 1221034626 for use recalledHoriba InstrumentsDevice MalfunctionNationwide

More recalls from Horiba Instruments

All 28

Other Test Kits & Diagnostics recalls

All 2,671