CLASS II ONGOING Device recall · Reported 8 Oct 2025 · FDA Enforcement Report
Implant Epikut Plus Mt16 recalled by S.I.N. Implant System Ltda
S.I.N. Implant System Ltda is recalling Implant Epikut Plus Mt16, distributed nationwide in the US. The recall was initiated on 30 Jul 2025 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0017-2026
- FDA event ID
- 97553
- Recalling firm
- S.I.N. Implant System Ltda, Sao Paulo, N/A
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 30 Jul 2025
- FDA classified
- 1 Oct 2025
- Reported
- 8 Oct 2025
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Packaging Defects
- Device type
- Other Medical Devices
Timeline
| 30 Jul 2025 | Recall initiated by company | |
| 1 Oct 2025 | Classified by FDA | +63 days |
| 8 Oct 2025 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Dental implants were imported under temperature and storage conditions that do not comply with the instructions for use and product labeling so there may be packaging damage, which could compromise product quality and performance.
Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.
Product as listed by the FDA
IMPLANT EPIKUT PLUS MT16 3,5X10,0MM, REF: ILM 3510N; IMPLANT EPIKUT PLUS MT16 3.8X11.5MM, REF: ILM 3511N; IMPLANT EPIKUT PLUS MT16 3,5X13,0MM, REF: ILM 3513N; IMPLANT EPIKUT PLUS MT16 3,5X15,0MM, REF: ILM 3515N; IMPLANT EPIKUT PLUS MT16 3,8X10,0MM, REF: ILM 3810N; IMPLANT CM16 3,8 X 11MM, REF: ILM 3811N; IMPLANT EPIKUT PLUS MT16 3,8X15,0MM, REF: ILM 3815N; IMPLANT EPIKUT PLUS MT16 4,0X10,0MM, REF: ILM 4010N; IMPLANT EPIKUT PLUS MT16 4,0X11,5MM, REF: ILM 4011N; IMPLANT EPIKUT PLUS MT16 4,0X13,0MM, REF: ILM 4013N; IMP EPIKUT PLUS LONG MT16 4,0X20,0MM, RES: ILM 4020N; IMPLANT EPIKUT PLUS MT16 4,0X8,5MM, RES: ILM 4085N; IMPLANT EPIKUT PLUS MT16 4,5X13,0MM, REF: ILM 4513N; IMPLANT EPIKUT PLUS MT16 4,5X15,0MM, REF: ILM 4515N; IMPLANT EPIKUT PLUS MT16 4,5X8,5MM, REF: ILM 4585N; IMPLANT EPIKUT PLUS MT16 5,0X10,0MM, REF: ILM 5010N; IMPLANT EPIKUT PLUS MT16 5,0X11,5MM, REF: ILM 5011N; IMPLANT EPIKUT PLUS MT16 4,5X10,0MM, REF: ILM 4510N; IMPLANT EPIKUT PLUS MT16 5,0X13,0MM, RES: ILM 5013N; IMPLANT EPIKUT PLUS MT16 5,0X15,0MM, REF: ILM 5015N; IMPLANT EPIKUT PLUS MT16 5,0X8,5MM, REF: ILM 5085N; IMP EPIKUT PLUS LONG MT16 4,0X18,0MM, REF: ILM 4018N; IMP EPIKUT PLUS LONG MT16 3,8X18,0MM, REF: ILM 3818N; IMPLANT EPIKUT PLUS MT16 4,5X11,5MM, REF: ILM 4511N; IMPLANT EPIKUT MT16 4,5X15,0MM, REF: ILM 4515; IMPLANT EPIKUT MT16 4,5X8,5MM, REF: ILM 4585; IMPLANT EPIKUT MT16 3,8X13,0MM, REF: ILM 3813; IMPLANT EPIKUT MT16 3,8X15,0MM, REF: ILM 3815; IMPLANT EPIKUT MT16 5,0X8,5MM, REF: ILM 5085; IMPLANT EPIKUT MT16 3,8X11,5MM, REF: ILM 3811; IMPLANT EPIKUT PLUS MT16 3,8X20,0MM, REF: ILM 3820N; IMPLANT EPIKUT PLUS MT16 4,5X18,0MM, REF: ILM 4518N; IMPLANT UNITITE MORSE COMPACT 6.0X7MM, REF: UCMC 6007N
Where it was distributed
US Nationwide distribution including in the states of TX, IL, KY, NC, TN, CO, CA, MO, MD, AL, FL, PA, UT, NE, ID, AZ, WA, OK, OR, NV, NJ, VA, NY, LA, CT, SD, SC, HI, KS, MA, MI, NM, OH, MT, WI, MN, ME.
Nationwide AlabamaArizonaCaliforniaColoradoConnecticutFloridaHawaiiIdahoIllinoisKansasKentuckyLouisianaMassachusettsMarylandMaineMichiganMinnesotaMissouriMontanaNorth Carolina
Questions about this recall
Why was it recalled?
Dental implants were imported under temperature and storage conditions that do not comply with the instructions for use and product labeling so there may be packaging damage, which could compromise product quality and performance.
Which lots are affected?
REF/UDI-DI/Lot(Expiration): ILM 3510N/7899995277558/X010469795(9/12/2026); ILM 3511N/7899995274342/X020477141(7/11/2027), X040490928(1/3/2028), X030483330(1/2/2028), X030482747(9/2/2027), X030482813(10/10/2027), W100457486(2/9/2027), X030482750(9/2/2027), W020415041(10/12/2026), X020477137(3/4/2027); ILM 3513N/7899995274335/W080447571(2/4/2028), X010469706(3/4/2028), X010471894(4/11/2028), W020415045(1/3/2028), X120531584(6/5/2028), X040490953(5/11/2028), Y010537386(6/12/2028); ILM 3515N/7899995280893/W020420816(6/12/2028), X030484303(10/2/2028), X010472646(9/12/2028), X010472644(8/10/2028), X120531182(10/5/2028); ILM 3810N/7899995274274/X120531775(10/5/2028); ILM 3811N/7899995274229/X010471364(10/12/2028), X080513064(12/2/2028), X100521415(12/2/2028), X120531368(2/2/2029), X070504152(11/5/2028), X020478981(10/12/2028), X120531376(3/2/2029), X020478983(11/2/202), X010471356(10/12/2028), X100523193(1/2/2029); ILM 3815N/7899995274168/X120531820(11/2/2029), X120531267(11/2/2029), X010473617(11/2/2029), V110406564(10/2/2029); ILM 4010N/7899995274137/Y020538716(14/12/2028), X120530744(14/12/2028), X120530266(14/01/2029); ILM 4011N/7899995274113/W100458406(15/12/2028), V110406330(15/12/2028), X120531292(16/05/2028), V110404633(15/12/2028), X090516009(16/01/2029); ILM 4013N/7899995274106/W080449448(16/08/2027), X120531842(17/12/2028), X120530728(17/12/2028), X120530726(17/01/2029), X120530724(16/12/2028), X120530730(17/12/2028), Y010537900(17/12/2028), W090454924(16/09/2028), X010473946(16/12/2028), W100458876(16/09/2028), X120531618(17/12/2028), X090517847(16/12/2028), X120530300(16/12/2028), X120531930(17/12/2028), X010472967(16/12/2028), X090517840(16/12/2028), X090515626(16/12/2028), X090515638(16/12/2028); ILM 4020N/7899995273499/X120530786(17/12/2028); ILM 4085N/7899995274144/X120531822(18/06/2028); ILM 4513N/7899995274038/X120530720(20/03/2028), Y010538078(22/09/2028), X120530722(20/10/2027), X120530849(22/06/2028); ILM 4515N/7899995274021/X120530902(23/10/2028), X120530716(23/07/2028), X120530340(23/03/2028); ILM 4585N/7899995274076/X110529195(25/11/2028), X110529193(25/10/2027); ILM 5010N/7899995273970/X040490644(26/12/2028), X040490886(26/12/2028); ILM 5011N/7899995273963/X120530851(26/12/2028); ILM 4510N/7899995274069/Y010537208(18/12/2028); ILM 5013N/7899995273956/X100522628(27/06/2027); ILM 5015N/7899995273932/X120530844(28/12/2028), X040490872(27/12/2028); ILM 5085N/7899995274007/X120530511(28/12/2028), X120530714(29/01/2029), X120530338(28/12/2028), Y010536031(30/05/2028); ILM 4018N/7899995273529/Y010537853(17/12/2028); ILM 3818N/7899995280886/Y010536867(13/12/2028); ILM 4511N/7899995274045/X120531860(18/12/2028), X120531935(18/12/2028), X120530732(18/12/2028); ILM 4515/7899995280169/X060497920(23/02/2028), X060497915(23/01/2028); ILM 4585/7899995280251/X040490690(24/01/2028), X120529433(24/01/2029), X120529450(24/01/2029); ILM 3813/7899995280527/Y020539359(6/2/2029); ILM 3815/7899995280473/Y010535029(7/2/2029); ILM 5085/7899995280091/X090516829(28/12/2028); ILM 3811/7899995280534/X110528601(10/10/2028); ILM 3820N/7899995273642/V110406210(13/12/2028), X020479077(X020479077); ILM 4518N/7899995273437/W060438138(24/01/2028), X120531278(24/01/2028); UCMC 6007N/7899995215895/Y010536358(30/10/2028);
Where was it sold?
US Nationwide distribution including in the states of TX, IL, KY, NC, TN, CO, CA, MO, MD, AL, FL, PA, UT, NE, ID, AZ, WA, OK, OR, NV, NJ, VA, NY, LA, CT, SD, SC, HI, KS, MA, MI, NM, OH, MT, WI, MN, ME.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other Other Medical Devices recalls
All 13,908| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Plastics Tray kits recalled by Avid Medical | Avid Medical | Foreign Material | Nationwide |
| Device | CLASS II | Heart Prep Tray kits recalled by Avid Medical | Avid Medical | Other | Nationwide |
| Device | CLASS II | Pack Open Heart Cfh kits recalled by Avid Medical | Avid Medical | Other | Nationwide |
| Device | CLASS II | Sperm Ab Elisa (96 well plate test) recalled by DRG International | DRG International | Other | Nationwide |
| Device | CLASS III | BeneView Series recalled by Mindray DS USA | Mindray DS USA | Device Malfunction | Nationwide |