CLASS II ONGOING Device recall · Reported 22 Jan 2025 · FDA Enforcement Report
Implant Epikut S Plus CM 16-Degrees recalled by Sin Sistema De Implante
SIN Sistema De Implante Nacional is recalling Implant Epikut S Plus CM 16-Degrees, distributed in 10 states. The recall was initiated on 1 Sep 2024 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0920-2025
- FDA event ID
- 95983
- Recalling firm
- SIN Sistema De Implante Nacional, Sao Paulo
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 1 Sep 2024
- FDA classified
- 15 Jan 2025
- Reported
- 22 Jan 2025
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Labeling Errors
- Device type
- Other Medical Devices
Timeline
| 1 Sep 2024 | Recall initiated by company | |
| 15 Jan 2025 | Classified by FDA | +136 days |
| 22 Jan 2025 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Dental implant box incorrectly labels the diameter as 4,5 X 18,0 mm, but the internal label correctly labels the diameter as 4,0 X 18,0 mm; if the dentist drills the bone with the larger diameter stated on the box, the smaller diameter implant, will become loose and will not fit, which may necessitate an additional surgery.
Product as listed by the FDA
Implant Epikut S Plus CM 16-Degrees 4,5 X 18,0 mm, REF: ILM 4518N
Where it was distributed
US distribution to states of: KY, CO, CA, NY, ID, NJ, FL, TX, KY, IL, OH; AND OUS (foreign) country of Brazil.
CaliforniaColoradoFloridaIdahoIllinoisKentuckyNew JerseyNew YorkOhioTexas
Questions about this recall
Why was it recalled?
Dental implant box incorrectly labels the diameter as 4,5 X 18,0 mm, but the internal label correctly labels the diameter as 4,0 X 18,0 mm; if the dentist drills the bone with the larger diameter stated on the box, the smaller diameter implant, will become loose and will not fit, which may necessitate an additional surgery.
Which lots are affected?
UDI-DI: 07899995273437, Lot: X010469757, Expiration: 2028/01
Where was it sold?
US distribution to states of: KY, CO, CA, NY, ID, NJ, FL, TX, KY, IL, OH; AND OUS (foreign) country of Brazil.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other Other Medical Devices recalls
All 13,986| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | ANNOUNCED | Luna Bilevel Positive Airway Pressure (BPAP) System recalledLuna | React Health | Other | VT |
| Device | ANNOUNCED | Beds, Rails, Mattresses, and accessories recalled by MedlineMedline | Medline | Other | |
| Device | CLASS II | Hemochron Jr., Whole Blood Microcoagulation System recalled | Accriva Diagnostics | Other | Nationwide |
| Device | CLASS II | Senographe Pristina 3D recalled by GE Medical Systems | GE Medical Systems | Other | WI |
| Device | CLASS II | Mobile column TruSystem recalled by Baxter Healthcare | Baxter Healthcare | Device Malfunction | Nationwide |