CLASS I ONGOING Device recall · Reported 20 Sep 2023 · FDA Enforcement Report

The implantable pulse generator recalled by Abbott Medical

Abbott Medical is recalling The implantable pulse generator (IPG) is an electronic device designed to be connected to, distributed nationwide in the US. The recall was initiated on 18 Jul 2023 and the FDA has classified it as Class I.

Check your package

Lot, batch or date codesModel Number: 3672 UDI-DI Codes: 05415067046406 Serial Numbers: US CBW328.1 CCA167.1 CCA177.1 CCU613.1 CCV760.1 CCX700.1 CCY890.1 CDR307.1 CDW220.1 CEC548.1 CEC719.1 CEE095.1 CEH111.1 CEH156.1 CEH200.1 CEH272.1 CEH293.1 CEH334.1 CEH964.1 CEH975.1 CEH993.1 CEK127.1 CEK258.1 CEK265.1 CEK271.1 CEK310.1 CEK432.1 CEM232.1 CEM340.1 CEM349.1 CEM375.1 CEM383.1 CEM396.1 CEN429.1 CEN444.1 CEP494.1 CEP497.1 CEP537.1 CEP540.1 CEP541.1 CEP553.1 CEP554.1 CEP562.1 CEP585.1 CEP645.1 CEP658.1 CEP663.1 CEP671.1 CEP685.1 CEP725.1 CEP730.1 CEP838.1 CEP847.1 CEP867.1 CER393.1 CER395.1 CER462.1 CER482.1 CER509.1 CER576.1 CET134.1 CET147.1 CET298.1 CET336.1 CET343.1 CET649.1 CEY608.1 CEY648.1 CFL299.1 CFL403.1 CFL540.1 CFL582.1 CFP963.1 CFP987.1 CFP991.1 CFR079.1 CFR100.1 CFR121.1 CFR392.1 CFR446.1 CFR478.1 CFR585.1 CFU079.1 CFU082.1 CFU091.1 CFU097.1 CFU098.1 CFU099.1 CFU121.1 CFU126.1 CFU132.1 CFU145.1 CFU333.1 CFU401.1 CFU426.1 CFU441.1 CFU460.1 CFU467.1 CFU472.1 CFU483.1 CFU526.1 CFU547.1 CFU569.1 CFU646.1 CFU687.1 CFV457.1 CFV588.1 CFV718.1 CFX211.1 CGA013.1 CGA029.1 CGA065.1 CGA097.1 CGA145.1 CGA331.1 CGA365.1 CGA406.1 CGA426.1 CGA451.1 CGA491.1 CGA500.1 CGA515.1 CGA581.1 CGC898.1 CGC908.1 CGC916.1 CGC921.1 CGC926.1 CGC955.1 CGC957.1 CGD063.1 CGD070.1 CGD076.1 CGD085.1 CGD101.1 CGD111.1 CGD119.1 CGD123.1 CGD192.1 CGD193.1 CGD220.1 CGD231.1 CGD240.1 CGD248.1 CGD251.1 CGD277.1 CGD278.1 CGD283.1 CGD308.1 CGD342.1 CGD343.1 CGD362.1 CGD694.1 CGD719.1 CGD739.1 CGD777.1 CGD790.1 CGD832.1 CGD892.1 CGG890.1 CGH131.1 CGH175.1 CGH183.1 CGH199.1 CGH245.1 CGH346.1 CGH413.1 CGL108.1 CGL207.1 CGL321.1 CGL863.1 CGL866.1 CGL889.1 CGP051.1 CGP078.1 CGP326.1 CGP615.1 CGS914.1 CGW231.1 CHB011.1 CHE598.1 CHE845.1 CHF161.1 CHF178.1 CHH717.1 CHH744.1 CHH773.1 CHJ032.1 CHK331.1 CHK354.1 CHK434.1 CHK648.1 CHM211.1 CHM389.1 CHM509.1 CHP603.1 CHP605.1 CHP621.1 CHR155.1 CHR240.1 CHT217.1 CHT238.1 CHT240.1 CHT241.1 CHT243.1 CHT245.1 CHT272.1 CHT356.1 CHT361.1 CHT373.1 CHT469.1 CHT478.1 CHT539.1 CHT553.1 CHT561.1 CHT589.1 CHT609.1 CHT624.1 CHT758.1 CHT815.1 CHT834.1 CHV674.1 CHV680.1 CHV710.1 CHV713.1 CHV794.1 CHV901.1 CHV989.1 CHW217.1 CHW262.1 CHW263.1 CHW306.1 CHW350.1 CHW362.1 CHW526.1 CHW560.1 CHW590.1 CHW591.1 CHW640.1 CHZ471.1 CHZ563.1 CHZ576.1 CJA246.1 CJG129.1 CJG177.1 CJM932.1 CJM940.1 CJN773.1 CJN780.1 CJN879.1 CJN987.1 CJP047.1 CJW418.1 CJW475.1 CJW534.1 CJW548.1 CJW586.1 CJX925.1 CJX962.1 CJX982.1 CJX986.1 CJY012.1 CJY075.1 CJY111.1 CJY226.1 CJY238.1 CJY304.1 CJY323.1 CJY360.1 CJY409.1 CJZ061.1 CJZ103.1 CJZ135.1 CJZ157.1 CJZ224.1 CJZ241.1 CJZ253.1 CJZ256.1 CJZ283.1 CJZ293.1 CJZ300.1 CJZ330.1 CJZ420.1 CJZ444.1 CJZ450.1 CKB402.1 CKB428.1 CKB439.1 CKB447.1 CKB487.1 CKB507.1 CKB608.1 CKB624.1 CKB654.1 CKB696.1 CKB725.1 CKB767.1 CKB887.1 CKD072.1 CKF569.1 CKF587.1 CKG394.1 CKG550.1 CKG586.1 CKG607.1 CKG915.1 CKG954.1 CKG956.1 CKG962.1 CKH467.1 CKH606.1 CKH692.1 CKH719.1 CKH736.1 CKH765.1 CKH774.1 CKH786.1 CKH815.1 CKH859.1 CKH923.1 CKL195.1 CKL217.1 CKL236.1 CKL295.1 CKL345.1 CKL464.1 CKL487.1 CKP251.1 CKP255.1 CKP277.1 CKP291.1 CKP393.1 CKP435.1 CKP483.1 CKP495.1 CKP563.1 CKP656.1 CKP678.1 CKP682.1 CKP717.1 CKP733.1 CKP803.1 CKP939.1 CKS483.1 CKS519.1 CKS531.1 CKS552.1 CKS588.1 CKS590.1 CKS616.1 CKS641.1 CKS662.1 CKS673.1 CKS684.1 CKS692.1 CKS698.1 CKS706.1 CKS827.1 CKS856.1 CKS869.1 CKS954.1 CKS986.1 CKS992.1 CKS994.1 CKS997.1 CKT159.1 CKY723.1 CKY781.1 CLC604.1 CLD150.1 CLD293.1 CLF739.1 CLF855.1 CLF880.1 CLF888.1 CLF901.1 CLG005.1 CLG094.1 CLG124.1 CLG391.1 CLG393.1 CLG395.1 CLG396.1 CLG423.1 CLG457.1 CLG534.1 CLK164.1
Quantity recalled387 devices (US only)

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2508-2023
FDA event ID
92688
Recalling firm
Abbott Medical, Plano, TX
Classification
Class I
Status
Ongoing
Recall initiated
18 Jul 2023
FDA classified
11 Sep 2023
Reported
20 Sep 2023
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Packaging Defects
Device type
Spinal Implants

Timeline

18 Jul 2023Recall initiated by company
11 Sep 2023Classified by FDA+55 days
20 Sep 2023Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Firm has received complaints from patients who are unable to exit MRI (Magnetic Resonance Imaging) mode on their IPGs (Implantable Pulse Generators).

Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.

Product as listed by the FDA

Proclaim Plus 7 Implantable Pulse Generator REF 3672 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensions or leads with up to 16 electrodes total. It is powered by a hermetically sealed battery within a titanium case and uses microelectronic circuitry to generate constant-current electrical stimulation. The IPG can deliver stimulation with a single program or with multiple programs. Each program can provide stimulation to a single anatomical area or to multiple areas. The IPG communicates wirelessly with system programmers and controllers, and the IPGs are available in small and large sizes to accommodate different power needs. The Proclaim" XR and Proclaim" Plus IPGs are used for Spinal Cord Stimulation (SCS). The Proclaim" DRG IPG is used for Dorsal Root Ganglion (DRG) stimulation. The Infinity" IPG is used for Deep Brain Stimulation (DBS).

Where it was distributed

Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Algeria, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, Colombia, Costa Rica, Czech Republic, Denmark, Ecuador, Estonia, Finland, France, Germany, Greece, Hungary, India, Iraq, Ireland, Israel, Italy, Japan, Korean, Kuwait, Lithuania, Luxembourg, Mexico, Netherlands, New Zealand, Norway, Pakistan, Poland, Portugal, Reunion, Russian Federation, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom.

Nationwide AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaDelawareFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusetts

Questions about this recall

Why was it recalled?

Firm has received complaints from patients who are unable to exit MRI (Magnetic Resonance Imaging) mode on their IPGs (Implantable Pulse Generators).

Which lots are affected?

Model Number: 3672 UDI-DI Codes: 05415067046406 Serial Numbers: US CBW328.1 CCA167.1 CCA177.1 CCU613.1 CCV760.1 CCX700.1 CCY890.1 CDR307.1 CDW220.1 CEC548.1 CEC719.1 CEE095.1 CEH111.1 CEH156.1 CEH200.1 CEH272.1 CEH293.1 CEH334.1 CEH964.1 CEH975.1 CEH993.1 CEK127.1 CEK258.1 CEK265.1 CEK271.1 CEK310.1 CEK432.1 CEM232.1 CEM340.1 CEM349.1 CEM375.1 CEM383.1 CEM396.1 CEN429.1 CEN444.1 CEP494.1 CEP497.1 CEP537.1 CEP540.1 CEP541.1 CEP553.1 CEP554.1 CEP562.1 CEP585.1 CEP645.1 CEP658.1 CEP663.1 CEP671.1 CEP685.1 CEP725.1 CEP730.1 CEP838.1 CEP847.1 CEP867.1 CER393.1 CER395.1 CER462.1 CER482.1 CER509.1 CER576.1 CET134.1 CET147.1 CET298.1 CET336.1 CET343.1 CET649.1 CEY608.1 CEY648.1 CFL299.1 CFL403.1 CFL540.1 CFL582.1 CFP963.1 CFP987.1 CFP991.1 CFR079.1 CFR100.1 CFR121.1 CFR392.1 CFR446.1 CFR478.1 CFR585.1 CFU079.1 CFU082.1 CFU091.1 CFU097.1 CFU098.1 CFU099.1 CFU121.1 CFU126.1 CFU132.1 CFU145.1 CFU333.1 CFU401.1 CFU426.1 CFU441.1 CFU460.1 CFU467.1 CFU472.1 CFU483.1 CFU526.1 CFU547.1 CFU569.1 CFU646.1 CFU687.1 CFV457.1 CFV588.1 CFV718.1 CFX211.1 CGA013.1 CGA029.1 CGA065.1 CGA097.1 CGA145.1 CGA331.1 CGA365.1 CGA406.1 CGA426.1 CGA451.1 CGA491.1 CGA500.1 CGA515.1 CGA581.1 CGC898.1 CGC908.1 CGC916.1 CGC921.1 CGC926.1 CGC955.1 CGC957.1 CGD063.1 CGD070.1 CGD076.1 CGD085.1 CGD101.1 CGD111.1 CGD119.1 CGD123.1 CGD192.1 CGD193.1 CGD220.1 CGD231.1 CGD240.1 CGD248.1 CGD251.1 CGD277.1 CGD278.1 CGD283.1 CGD308.1 CGD342.1 CGD343.1 CGD362.1 CGD694.1 CGD719.1 CGD739.1 CGD777.1 CGD790.1 CGD832.1 CGD892.1 CGG890.1 CGH131.1 CGH175.1 CGH183.1 CGH199.1 CGH245.1 CGH346.1 CGH413.1 CGL108.1 CGL207.1 CGL321.1 CGL863.1 CGL866.1 CGL889.1 CGP051.1 CGP078.1 CGP326.1 CGP615.1 CGS914.1 CGW231.1 CHB011.1 CHE598.1 CHE845.1 CHF161.1 CHF178.1 CHH717.1 CHH744.1 CHH773.1 CHJ032.1 CHK331.1 CHK354.1 CHK434.1 CHK648.1 CHM211.1 CHM389.1 CHM509.1 CHP603.1 CHP605.1 CHP621.1 CHR155.1 CHR240.1 CHT217.1 CHT238.1 CHT240.1 CHT241.1 CHT243.1 CHT245.1 CHT272.1 CHT356.1 CHT361.1 CHT373.1 CHT469.1 CHT478.1 CHT539.1 CHT553.1 CHT561.1 CHT589.1 CHT609.1 CHT624.1 CHT758.1 CHT815.1 CHT834.1 CHV674.1 CHV680.1 CHV710.1 CHV713.1 CHV794.1 CHV901.1 CHV989.1 CHW217.1 CHW262.1 CHW263.1 CHW306.1 CHW350.1 CHW362.1 CHW526.1 CHW560.1 CHW590.1 CHW591.1 CHW640.1 CHZ471.1 CHZ563.1 CHZ576.1 CJA246.1 CJG129.1 CJG177.1 CJM932.1 CJM940.1 CJN773.1 CJN780.1 CJN879.1 CJN987.1 CJP047.1 CJW418.1 CJW475.1 CJW534.1 CJW548.1 CJW586.1 CJX925.1 CJX962.1 CJX982.1 CJX986.1 CJY012.1 CJY075.1 CJY111.1 CJY226.1 CJY238.1 CJY304.1 CJY323.1 CJY360.1 CJY409.1 CJZ061.1 CJZ103.1 CJZ135.1 CJZ157.1 CJZ224.1 CJZ241.1 CJZ253.1 CJZ256.1 CJZ283.1 CJZ293.1 CJZ300.1 CJZ330.1 CJZ420.1 CJZ444.1 CJZ450.1 CKB402.1 CKB428.1 CKB439.1 CKB447.1 CKB487.1 CKB507.1 CKB608.1 CKB624.1 CKB654.1 CKB696.1 CKB725.1 CKB767.1 CKB887.1 CKD072.1 CKF569.1 CKF587.1 CKG394.1 CKG550.1 CKG586.1 CKG607.1 CKG915.1 CKG954.1 CKG956.1 CKG962.1 CKH467.1 CKH606.1 CKH692.1 CKH719.1 CKH736.1 CKH765.1 CKH774.1 CKH786.1 CKH815.1 CKH859.1 CKH923.1 CKL195.1 CKL217.1 CKL236.1 CKL295.1 CKL345.1 CKL464.1 CKL487.1 CKP251.1 CKP255.1 CKP277.1 CKP291.1 CKP393.1 CKP435.1 CKP483.1 CKP495.1 CKP563.1 CKP656.1 CKP678.1 CKP682.1 CKP717.1 CKP733.1 CKP803.1 CKP939.1 CKS483.1 CKS519.1 CKS531.1 CKS552.1 CKS588.1 CKS590.1 CKS616.1 CKS641.1 CKS662.1 CKS673.1 CKS684.1 CKS692.1 CKS698.1 CKS706.1 CKS827.1 CKS856.1 CKS869.1 CKS954.1 CKS986.1 CKS992.1 CKS994.1 CKS997.1 CKT159.1 CKY723.1 CKY781.1 CLC604.1 CLD150.1 CLD293.1 CLF739.1 CLF855.1 CLF880.1 CLF888.1 CLF901.1 CLG005.1 CLG094.1 CLG124.1 CLG391.1 CLG393.1 CLG395.1 CLG396.1 CLG423.1 CLG457.1 CLG534.1 CLK164.1

Where was it sold?

Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Algeria, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, Colombia, Costa Rica, Czech Republic, Denmark, Ecuador, Estonia, Finland, France, Germany, Greece, Hungary, India, Iraq, Ireland, Israel, Italy, Japan, Korean, Kuwait, Lithuania, Luxembourg, Mexico, Netherlands, New Zealand, Norway, Pakistan, Poland, Portugal, Reunion, Russian Federation, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class I mean?

Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

1 other product · FDA event 92688

This product is part of a recall event; the main page for the event is The implantable pulse generator recalled by Abbott Medical.

ReportedClassRecallCompanyReasonWhere
CLASS IThe implantable pulse generator recalled by Abbott MedicalAbbott MedicalPackaging DefectsNationwide

More recalls from Abbott Medical

All 33

Other Spinal Implants recalls

All 1,161