CLASS II ONGOING Device recall · Reported 5 Feb 2025 · FDA Enforcement Report

Infinity Central Station recalled by Draeger Medical Systems

Draeger Medical Systems is recalling Infinity Central Station, distributed nationwide in the US. The recall was initiated on 14 Jan 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesInfinity CentralStation (ICS) Model Number: MS26800. UDI-DI Number: 04049098001878. Edifier Speaker Model Number: MS34036. UDI-DI Number: 04048675548027. Serial Numbers: 12946M00269, 12946M00314, 12946M00506, 12946M00519, 12946M00518, 12946M00553, 1OQ7WR-00103, 1OQ8CA-00249, 1OQ8CA-00254, 1OQ8CA-00229, 1OQ8CA-00221, 1OQ8CA-00282, 1OQ8CA-00314, 1OQ8CA-00193, 1OQ8CA-00200, 1OQ8CA-00194, 1OQ8CA-00181, 1OQ8CA-00178, 1OQ8CA-00248, 1OQ8CA-00125, 1OQ8CA-00071, 1OQ8CA-00032, 1OQ8CA-00171, 1OQ8CA-00164, 1OQ8CA-00149, 1OQ8CA-00069, 1OQ8CA-00057, 1OQ8CA-00153, 1OQ8CA-00150, 1OQ8CA-00148, 1OQ8CA-00165, 1OQ8CA-00155, 1OQ8CA-00097, 1OQ8CA-00113, 1OQ8CA-00106, 1OQ8CA-00075, 1OQ8CA-00052, 1OQ8CA-00067, 1OQ8CA-00128, 1OQ8CA-00132, 1OQ8CA-00268, 1OQ8CA-00198, 1OQ8CA-00255, 1OQ8CA-00151, 1OQ8CA-00342, 1OQ8CA-00391, 1OQ8CA-00321, 1OQ8CA-00402, 1OQ8CA-00395, 1OQ8CA-00297, 1OQ8CA-00240, 1OQ89M-00691, 1OQ89M-00690, 1OQ89M-00708, 1OQ89M-00693, 1OQ89M-00687, 1OQ89M-00707, 1OQ89M-00703, 1OQ89M-00679, 1OQ89M-00683, 1OQ89M-00692, 1OQ89M-00610, 1OQ89M-00613, 1OQ89M-00669, 1OQ89M-00635, 1OQ89M-00664, 1OQ89M-00697, P1OQ84F-00080, 1OQ89M-00684, 1OQ89M-00676, 1OQ89M-00695, 1OQ8CA-00086, 1OQ89M-00677, 1OQ8CA-00031, 1OQ8CA-00027, 1OQ8CA-00065, 1OQ8CA-00092, 1OQ8CA-00082, 1OQ89M-00602, 1OQ89M-00680, 1OQ89M-00675, 1OQ89M-00593, 1OQ89M-00686, 1OQ89M-00696, 1OQ89M-00681, 1OQ89M-00704, 1OQ89M-00674, 1OQ89M-00661, 1OQ89M-00663, P1OQ89M-00559, P1OQ89M-00540, 1OQ8CA-00134, 1OQ8CA-00095, 1OQ8CA-00043, 1OQ8CA-00074, 1OQ8CA-00762, 1OQ8CA-00688, 1OQ8CA-00670, 1OQ8CA-00637, 1OQ8CA-00633, 1OQ8E8-00037, 1OQ8E8-00053, 1OQ8CA-00773, 1OQ8CA-00683, 1OQ8CA-00500, 1OQ8CA-00476, 1OQ8CA-00449, 1OQ8CA-00420, 1OQ8CA-00460, 1OQ8CA-00463, 1OQ8CA-00507, 1OQ8CA-00350, 1OQ8CA-00442, 1OQ8CA-00433, 1OQ8CA-00447, 1OQ8CA-00601, 1OQ8CA-00604, 1OQ8CA-00695, 1OQ8CA-00513, 1OQ8CA-00575, 1OQ8CA-00577, 1OQ8CA-00591, 1OQ8CA-00197, 1OQ8CA-00516, 1OQ8CA-00541, 1OQ8CA-00542, 1OQ8CA-00568, 1OQ8CA-00499, 1OQ8CA-00498, P12051M00227, 1OQ8CA-00593, 12946M00532, 12946M00256, 12946M00278, 1OQ7WR-00203
Quantity recalled1,013 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1066-2025
FDA event ID
96057
Recalling firm
Draeger Medical Systems, Andover, MA
Classification
Class II
Status
Ongoing
Recall initiated
14 Jan 2025
FDA classified
30 Jan 2025
Reported
5 Feb 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction
Device type
Patient Monitors

Timeline

14 Jan 2025Recall initiated by company
30 Jan 2025Classified by FDA+16 days
5 Feb 2025Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Edifier speakers (shipped in or after April 2023) may not annunciate when configured in analog mode which can lead to delayed treatment to the patient.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Infinity Central Station. Model Number: MS26800 The Infinity CentralStation (ICS) is intended for centralized monitoring of adult, pediatric and neonatal patient data within the hospital or clinical environment. Centralized monitoring involves the display and management of data from networked patient monitors including the annunciation of visual and audible physiologic parameter alarms at a central monitoring workstation.

Where it was distributed

Worldwide - US Nationwide distribution in the states of AL, AS, AZ, CA, CO, FL, GA, IA, ID, IL, IN, KS, MA, NE, NY, OK, PA, SD, TN, TX, VA, WA and the countries of Argentina, Australia, Austria, Belgium, Botswana, Brazil, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Egypt, Finland, France, Germany, Greece, Hong Kong, India, Indonesia, Ireland, Italy, Japan, Kenya, Malaysia, Mexico, Monaco, Morocco, Netherlands, New Zealand, Norway, Peru, Philippines, Poland, Portugal, Romania, Samoa, American, Saudi Arabia, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkiye, United Kingdom.

Nationwide AlabamaArizonaCaliforniaColoradoFloridaGeorgiaIowaIdahoIllinoisIndianaKansasMassachusettsNebraskaNew YorkOklahomaPennsylvaniaSouth DakotaTennesseeTexasVirginia

Questions about this recall

Why was it recalled?

Edifier speakers (shipped in or after April 2023) may not annunciate when configured in analog mode which can lead to delayed treatment to the patient.

Which lots are affected?

Infinity CentralStation (ICS) Model Number: MS26800. UDI-DI Number: 04049098001878. Edifier Speaker Model Number: MS34036. UDI-DI Number: 04048675548027. Serial Numbers: 12946M00269, 12946M00314, 12946M00506, 12946M00519, 12946M00518, 12946M00553, 1OQ7WR-00103, 1OQ8CA-00249, 1OQ8CA-00254, 1OQ8CA-00229, 1OQ8CA-00221, 1OQ8CA-00282, 1OQ8CA-00314, 1OQ8CA-00193, 1OQ8CA-00200, 1OQ8CA-00194, 1OQ8CA-00181, 1OQ8CA-00178, 1OQ8CA-00248, 1OQ8CA-00125, 1OQ8CA-00071, 1OQ8CA-00032, 1OQ8CA-00171, 1OQ8CA-00164, 1OQ8CA-00149, 1OQ8CA-00069, 1OQ8CA-00057, 1OQ8CA-00153, 1OQ8CA-00150, 1OQ8CA-00148, 1OQ8CA-00165, 1OQ8CA-00155, 1OQ8CA-00097, 1OQ8CA-00113, 1OQ8CA-00106, 1OQ8CA-00075, 1OQ8CA-00052, 1OQ8CA-00067, 1OQ8CA-00128, 1OQ8CA-00132, 1OQ8CA-00268, 1OQ8CA-00198, 1OQ8CA-00255, 1OQ8CA-00151, 1OQ8CA-00342, 1OQ8CA-00391, 1OQ8CA-00321, 1OQ8CA-00402, 1OQ8CA-00395, 1OQ8CA-00297, 1OQ8CA-00240, 1OQ89M-00691, 1OQ89M-00690, 1OQ89M-00708, 1OQ89M-00693, 1OQ89M-00687, 1OQ89M-00707, 1OQ89M-00703, 1OQ89M-00679, 1OQ89M-00683, 1OQ89M-00692, 1OQ89M-00610, 1OQ89M-00613, 1OQ89M-00669, 1OQ89M-00635, 1OQ89M-00664, 1OQ89M-00697, P1OQ84F-00080, 1OQ89M-00684, 1OQ89M-00676, 1OQ89M-00695, 1OQ8CA-00086, 1OQ89M-00677, 1OQ8CA-00031, 1OQ8CA-00027, 1OQ8CA-00065, 1OQ8CA-00092, 1OQ8CA-00082, 1OQ89M-00602, 1OQ89M-00680, 1OQ89M-00675, 1OQ89M-00593, 1OQ89M-00686, 1OQ89M-00696, 1OQ89M-00681, 1OQ89M-00704, 1OQ89M-00674, 1OQ89M-00661, 1OQ89M-00663, P1OQ89M-00559, P1OQ89M-00540, 1OQ8CA-00134, 1OQ8CA-00095, 1OQ8CA-00043, 1OQ8CA-00074, 1OQ8CA-00762, 1OQ8CA-00688, 1OQ8CA-00670, 1OQ8CA-00637, 1OQ8CA-00633, 1OQ8E8-00037, 1OQ8E8-00053, 1OQ8CA-00773, 1OQ8CA-00683, 1OQ8CA-00500, 1OQ8CA-00476, 1OQ8CA-00449, 1OQ8CA-00420, 1OQ8CA-00460, 1OQ8CA-00463, 1OQ8CA-00507, 1OQ8CA-00350, 1OQ8CA-00442, 1OQ8CA-00433, 1OQ8CA-00447, 1OQ8CA-00601, 1OQ8CA-00604, 1OQ8CA-00695, 1OQ8CA-00513, 1OQ8CA-00575, 1OQ8CA-00577, 1OQ8CA-00591, 1OQ8CA-00197, 1OQ8CA-00516, 1OQ8CA-00541, 1OQ8CA-00542, 1OQ8CA-00568, 1OQ8CA-00499, 1OQ8CA-00498, P12051M00227, 1OQ8CA-00593, 12946M00532, 12946M00256, 12946M00278, 1OQ7WR-00203

Where was it sold?

Worldwide - US Nationwide distribution in the states of AL, AS, AZ, CA, CO, FL, GA, IA, ID, IL, IN, KS, MA, NE, NY, OK, PA, SD, TN, TX, VA, WA and the countries of Argentina, Australia, Austria, Belgium, Botswana, Brazil, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Egypt, Finland, France, Germany, Greece, Hong Kong, India, Indonesia, Ireland, Italy, Japan, Kenya, Malaysia, Mexico, Monaco, Morocco, Netherlands, New Zealand, Norway, Peru, Philippines, Poland, Portugal, Romania, Samoa, American, Saudi Arabia, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkiye, United Kingdom.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Draeger Medical Systems

All 12

Other Patient Monitors recalls

All 1,022