CLASS II ONGOING Device recall · Reported 3 Jul 2024 · FDA Enforcement Report

Infinity Acute Care System (IACS) Monitoring System recalled

Draeger Medical Systems is recalling Infinity Acute Care System (IACS) Monitoring System, Infinity M500 with Power, distributed nationwide in the US. The recall was initiated on 13 May 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesModel Number: MS20407. UDI-DI: 04049098054447. Rev 20, 21, and 23
Quantity recalled34,102 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2189-2024
FDA event ID
94644
Recalling firm
Draeger Medical Systems, Andover, MA
Classification
Class II
Status
Ongoing
Recall initiated
13 May 2024
FDA classified
25 Jun 2024
Reported
3 Jul 2024
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons
Device type
Patient Monitors

Timeline

13 May 2024Recall initiated by company
25 Jun 2024Classified by FDA+43 days
3 Jul 2024Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Device is not in full compliance to Type CF requirements of IEC 60601-1 and IEC 60601-2-34 standards. Therefore, there is a potential risk of electric shock. Health consequences may include electrical shock, burns, and cardiac arrhythmia.

Product as listed by the FDA

Infinity Acute Care System (IACS) Monitoring System, Infinity M500 with Power, Model No. MS20407. Rev 20, 21, and 23

Where it was distributed

US Nationwide distribution. International distribution to Afghanistan, Argentina, Armenia, Australia, Austria, Belarus, Belgium, Bolivia, Botswana, Brazil, Brunei Dar-es-S, Bulgaria, Canada, Chile, China, Columbia, Costa Rica, Croatia, Czech Republic, Egypt, Estonia, Finland, France, French Guiana, Georgia, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Kazakhstan, Kenya, Kuwait, Latvia, Liechtenstein, Lithuania, Malaysia, Malta, Martinique, Mexico, Monaco, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Fed., San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, T¿rkiye, Turkmenistan, Ukraine, United Kingdom, Uruguay, United Arab Emirates, Vietnam.

Nationwide Georgia

Questions about this recall

Why was it recalled?

Device is not in full compliance to Type CF requirements of IEC 60601-1 and IEC 60601-2-34 standards. Therefore, there is a potential risk of electric shock. Health consequences may include electrical shock, burns, and cardiac arrhythmia.

Which lots are affected?

Model Number: MS20407. UDI-DI: 04049098054447. Rev 20, 21, and 23

Where was it sold?

US Nationwide distribution. International distribution to Afghanistan, Argentina, Armenia, Australia, Austria, Belarus, Belgium, Bolivia, Botswana, Brazil, Brunei Dar-es-S, Bulgaria, Canada, Chile, China, Columbia, Costa Rica, Croatia, Czech Republic, Egypt, Estonia, Finland, France, French Guiana, Georgia, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Kazakhstan, Kenya, Kuwait, Latvia, Liechtenstein, Lithuania, Malaysia, Malta, Martinique, Mexico, Monaco, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Fed., San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, T¿rkiye, Turkmenistan, Ukraine, United Kingdom, Uruguay, United Arab Emirates, Vietnam.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.