CLASS II Device recall · Reported 15 Sep 2021 · FDA Enforcement Report
Ingenia 1.5T CX-a Magnetic Resonance Medical Electrical recalled
Philips North America is recalling Ingenia 1.5T CX-a Magnetic Resonance Medical Electrical Systems indicated for use as a, distributed nationwide in the US. The recall was initiated on 11 Aug 2021 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2414-2021
- FDA event ID
- 88491
- Recalling firm
- Philips North America, Cambridge, MA
- Classification
- Class II
- Status
- Completed
- Recall initiated
- 11 Aug 2021
- FDA classified
- 3 Sep 2021
- Reported
- 15 Sep 2021
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Test Kits & Diagnostics
Timeline
| 11 Aug 2021 | Recall initiated by company | |
| 3 Sep 2021 | Classified by FDA | +23 days |
| 15 Sep 2021 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
When using manual mode, the tabletop may not completely move in or out and result in potential delay in patient evacuation from the bore, which could result in a delay of treatment
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
Ingenia 1.5T CX-a Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body, or extremities Models: 781261 781262
Where it was distributed
Worldwide distribution - US Nationwide distribution and the countries of Albania, Andorra, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, French Guiana, Georgia, Germany, Ghana, Greece, Guadeloupe, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of Kosovo, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malaysia, Maldives, Malta, Martinique, Mauritius, Mayotte, Mexico, Moldova, Monaco, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Norway, Oman, Pakistan Palestine, State of Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Viet Nam.
Questions about this recall
Why was it recalled?
When using manual mode, the tabletop may not completely move in or out and result in potential delay in patient evacuation from the bore, which could result in a delay of treatment
Which lots are affected?
Model: 781261 Serial Number Unique Device Identifier (UDI#) 77008 77000 77004 77007 77003 77001 77002 77018 77016 77019 77010 77014 Model:781262 77131 (01)00884838068445(21)77131 77031 N/A 77046 N/A 77070 N/A 77124 (01)00884838068445(21)77124 77144 (01)00884838068445(21)77144 77172 (01)00884838068445(21)77172 77074 (01)00884838068445(21)77074 77058 N/A 77114 N/A 77127 (01)00884838068445(21)77127 77160 (01)00884838068445(21)77160 77159 (01)00884838068445(21)77159 77182 (01)00884838068445(21)77182 77180 (01)00884838068445(21)77180 77171 (01)00884838068445(21)77171 77179 (01)00884838068445(21)77179 77048 N/A 77040 N/A 77034 N/A 77039 N/A 77035 N/A 77092 N/A 77140 (01)00884838068445(21)77140 77177 (01)00884838068445(21)77177 77181 (01)00884838068445(21)77181 77178 (01)00884838068445(21)77178 77041 N/A 77103 N/A 77193 (01)00884838068445(21)77193 77203 (01)00884838068445(21)77203 77134 (01)00884838068445(21)77134 77047 N/A 77051 N/A 77050 N/A 77060 N/A 77194 (01)00884838068445(21)77194 77111 N/A 77119 (01)00884838068445(21)77119 77167 (01)00884838068445(21)77167 77183 (01)00884838068445(21)77183 77185 (01)00884838068445(21)77185 77187 (01)00884838068445(21)77187 77188 (01)00884838068445(21)77188 77198 (01)00884838068445(21)77198 77195 (01)00884838068445(21)77195 77196 (01)00884838068445(21)77196 77199 N/A 77202 N/A 77037 N/A 77057 N/A 77072 N/A 77075 N/A 77084 N/A 77085 N/A 77078 N/A 77086 N/A 77101 N/A 77110 (01)00884838068445(21)77110 77038 (01)00884838068445(21)77038 77157 (01)00884838068445(21)77157 77156 (01)00884838068445(21)77156 77056 N/A 77052 N/A 77054 N/A 77053 N/A 77055 N/A 77064 N/A 77065 N/A 77067 N/A 77066 N/A 77068 N/A 77069 N/A 77105 N/A 77126 (01)00884838068445(21)77126 77125 (01)00884838068445(21)77125 77137 (01)00884838068445(21)77137 77139 (01)00884838068445(21)77139 77136 (01)00884838068445(21)77136 77135 (01)00884838068445(21)77135 77138 (01)00884838068445(21)77138 77148 (01)00884838068445(21)77148 77141 (01)00884838068445(21)77141 77030 N/A 77033 N/A 77036 N/A 77042 N/A 77044 N/A 77043 N/A 77032 N/A 77049 N/A 77061 N/A 77059 N/A 77071 N/A 77090 N/A 77088 N/A 77093 N/A 77091 N/A 77087 N/A 77097 N/A 77083 N/A 77106 N/A 77112 N/A 77107 (01)00884838068445(21)77107 77115 N/A 77100 (01)00884838068445(21)77100 77118 (01)00884838068445(21)77118 77121 (01)00884838068445(21)77121 77129 (01)00884838068445(21)77129 77120 (01)00884838068445(21)77120 77128 (01)00884838068445(21)77128 77151 (01)00884838068445(21)77151 77143 (01)00884838068445(21)77143 77158 (01)00884838068445(21)77158 77162 (01)00884838068445(21)77162 77161 (01)00884838068445(21)77161 77155 (01)00884838068445(21)77155 77190 (01)00884838068445(21)77190 77191 (01)00884838068445(21)77191 77150 (01)00884838068445(21)77150 77081 N/A 77117 (01)00884838068445(21)77117 77123 (01)00884838068445(21)77123 77166 (01)00884838068445(21)77166 77045 N/A 77063 N/A 77109 N/A 77201 (01)00884838068445(21)77201 77113 N/A 77147 (01)00884838068445(21)77147 77122 (01)00884838068445(21)77122 77154 (01)00884838068445(21)77154 77169 (01)00884838068445(21)77169 77168 (01)00884838068445(21)77168 77130 (01)00884838068445(21)77130 77094 N/A 77062 N/A
Where was it sold?
Worldwide distribution - US Nationwide distribution and the countries of Albania, Andorra, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, French Guiana, Georgia, Germany, Ghana, Greece, Guadeloupe, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of Kosovo, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malaysia, Maldives, Malta, Martinique, Mauritius, Mayotte, Mexico, Moldova, Monaco, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Norway, Oman, Pakistan Palestine, State of Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Viet Nam.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other products in this recall event
11 other products · FDA event 88491More recalls from Philips North America
All 531| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Philips Brilliance recalled by Philips North America | Philips North America | Device Malfunction | MA |
| Device | CLASS II | Philips IQon Spectral CT recalled by Philips North America | Philips North America | Device Malfunction | MA |
| Device | CLASS II | Philips Ingenuity CT recalled by Philips North America | Philips North America | Device Malfunction | MA |
| Device | CLASS II | Philips Brilliance CT 64: Brilliance CT 64 Channel recalled | Philips North America | Device Malfunction | MA |
| Device | CLASS II | Cardiac Workstation recalled by Philips North America | Philips North America | Labeling Errors | Nationwide |
Other Test Kits & Diagnostics recalls
All 2,656| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | STA Staclot Protein recalled by Diagnostica Stago | Diagnostica Stago | Potency & Assay | Nationwide |
| Device | CLASS II | DxC 500i Clinical Analyzer Modules using recalled by Beckman Coulter | Beckman Coulter | Packaging Defects | Nationwide |
| Device | CLASS II | DxC 500 AU Clinical Analyzer Modules using recalled by Beckman Coulter | Beckman Coulter | Packaging Defects | Nationwide |
| Device | CLASS II | Syva EMIT 2000 Quinidine Assay recalled | Siemens Healthcare Diagnostics | Potency & Assay | Nationwide |
| Device | CLASS II | Beckman Coulter, Emit 2000 Quinidine Assay recalled | Siemens Healthcare Diagnostics | Potency & Assay | Nationwide |