CLASS II Device recall · Reported 18 Jun 2014 · FDA Enforcement Report
IntelliVue MX40 Wearable Patient Monitor in use recalled
Philips Medical Systems is recalling IntelliVue MX40 Wearable Patient Monitor in use with the Philips IntelliVue Information, distributed nationwide in the US. The recall was initiated on 22 May 2014 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1746-2014
- FDA event ID
- 68349
- Recalling firm
- Philips Medical Systems, Andover, MA
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 22 May 2014
- FDA classified
- 10 Jun 2014
- Reported
- 18 Jun 2014
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Patient Monitors
Timeline
| 22 May 2014 | Recall initiated by company | |
| 10 Jun 2014 | Classified by FDA | +19 days |
| 18 Jun 2014 | Published in enforcement report | +8 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
IntelliVue MX40 (part numbers 865350, 865352) with software revisions B.00 or B.01 may not resume alarming when the alarms are paused and the pause timer ends.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
IntelliVue MX40 Wearable Patient Monitor in use with the Philips IntelliVue Information Center iX (PIIC iX); 865350, 865351, 865352
Where it was distributed
United States: Nationwide Foreign Countries: Germany, Austria, Finland, Sweden, Australia, Qatar, Canada, Netherlands, New Zealand, Portugal, United Kingdom, Spain, Belgium, France, Switzerland, Norway, Czech Republic, Poland, Italy, Singapore, Denmark, Iceland, India, Saudi Arabia, Ireland, Oman, Bahrain, Japan, Israel, Malaysia, South Africa, Turkey, Thailand, Latvia, Bulgaria, Indonesia, Slovenia, United Arab Emirates, Hong Kong, Aruba, Chile, Lebanon, Korea, Kuwait, Taiwan, Argentina, Hungary, Luxembourg, Mexico, Gabon
Questions about this recall
Why was it recalled?
IntelliVue MX40 (part numbers 865350, 865352) with software revisions B.00 or B.01 may not resume alarming when the alarms are paused and the pause timer ends.
Which lots are affected?
Serial numbers within the following ranges that have the MX40 interfaces: 00700001 - US01443745 and US014Z1301 - US014Z1431
Where was it sold?
United States: Nationwide Foreign Countries: Germany, Austria, Finland, Sweden, Australia, Qatar, Canada, Netherlands, New Zealand, Portugal, United Kingdom, Spain, Belgium, France, Switzerland, Norway, Czech Republic, Poland, Italy, Singapore, Denmark, Iceland, India, Saudi Arabia, Ireland, Oman, Bahrain, Japan, Israel, Malaysia, South Africa, Turkey, Thailand, Latvia, Bulgaria, Indonesia, Slovenia, United Arab Emirates, Hong Kong, Aruba, Chile, Lebanon, Korea, Kuwait, Taiwan, Argentina, Hungary, Luxembourg, Mexico, Gabon
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Philips Medical Systems
All 81| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Ingenuity CT CT5000 Ingenuity Plus CT5000 Ingenuity Pro recalled | Philips Medical Systems | Device Malfunction | Nationwide |
| Device | CLASS II | Brilliance iCT Brilliance iCT Upgrades Brilliance iCT SP recalled | Philips Medical Systems | Device Malfunction | Nationwide |
| Device | CLASS II | IQon Spectral CT IQon Spectral CT Upgrades recalled | Philips Medical Systems | Device Malfunction | Nationwide |
| Device | CLASS II | Philips Spectral CT on Rails recalled by Philips Medical Systems | Philips Medical Systems | Device Malfunction | Nationwide |
| Device | CLASS II | Model: Incisive CT; Product Code (REF): (1) 728143, (2) recalled | Philips Medical Systems | Device Malfunction | 18 states |
Other Patient Monitors recalls
All 1,013| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | Bravo CF capsule delivery device, Product Number FGS recalled | Given Imaging | Device Malfunction | Nationwide |
| Device | CLASS II | Lux-Dx, Lux-Dx II, Lux-Dx II+ Insertable Cardiac Monitor recalled | Boston Scientific | Device Malfunction | Nationwide |
| Device | CLASS II | Lux-Dx, Lux-Dx II, Lux-Dx II+ Insertable Cardiac Monitor recalled | Boston Scientific | Device Malfunction | Nationwide |
| Device | CLASS II | Lux-Dx, Lux-Dx II, Lux-Dx II+ Insertable Cardiac Monitor recalled | Boston Scientific | Device Malfunction | Nationwide |
| Device | CLASS II | IQecg Digital ECG recalled by Midmark | Midmark | Labeling Errors | Nationwide |