CLASS II Device recall · Reported 18 Apr 2018 · FDA Enforcement Report

iQ200 Series Urine Microscopy Analyzer recalled by Beckman Coulter

Beckman Coulter is recalling iQ200 Series Urine Microscopy Analyzer. The recall was initiated on 2 Mar 2018 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesAll
Quantity recalled5247 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1362-2018
FDA event ID
79566
Recalling firm
Beckman Coulter, Brea, CA
Classification
Class II
Status
Terminated
Recall initiated
2 Mar 2018
FDA classified
11 Apr 2018
Reported
18 Apr 2018
Type
Voluntary: Firm initiated
Customers notified by
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Reason category
Device Malfunction

Timeline

2 Mar 2018Recall initiated by company
11 Apr 2018Classified by FDA+40 days
18 Apr 2018Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Beckman Coulter has determined that there is a potential for under-reporting casts. This can occur if per high-power field (/HPF) units of measurement for casts are selected in the iQ200 software, but the abnormal threshold and/or grading format is set up based on reporting per low-power field (/LPF)* or *per microliter. This may occur during initial method validation or if settings are altered after the initial validation.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

BECKMAN COULTER iQ200 Series Urine Microscopy Analyzer, Catalog Numbers: 700-3345, C10684, 700-3375, C10683, 700-3325, 700-3347, 700-3320, 700-3322, 700-3370

Where it was distributed

Worldwide

Questions about this recall

Why was it recalled?

Beckman Coulter has determined that there is a potential for under-reporting casts. This can occur if per high-power field (/HPF) units of measurement for casts are selected in the iQ200 software, but the abnormal threshold and/or grading format is set up based on reporting per low-power field (/LPF)* or *per microliter. This may occur during initial method validation or if settings are altered after the initial validation.

Which lots are affected?

All

Where was it sold?

Worldwide

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.