CLASS I ONGOING Device recall · Reported 10 Jun 2026 · FDA Enforcement Report

Ivenix Infusion System (IIS), Large Volume Pump recalled

Fresenius Kabi USA is recalling Ivenix Infusion System (IIS), Large Volume Pump, distributed in 21 states. The recall was initiated on 6 May 2026 and the FDA has classified it as Class I.

Check your package

Lot, batch or date codesModel Number: LVP-0004; UDI-DI: 00811505030320; All Serial/Lot Numbers:
Quantity recalled18,444

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2224-2026
FDA event ID
98945
Recalling firm
Fresenius Kabi USA, North Andover, MA
Classification
Class I
Status
Ongoing
Recall initiated
6 May 2026
FDA classified
29 May 2026
Reported
10 Jun 2026
Reason category
Device Malfunction

Timeline

6 May 2026Recall initiated by company
29 May 2026Classified by FDA+23 days
10 Jun 2026Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Fresenius Kabi has been made aware via customer complaints that dropping or severe jarring of the Ivenix LVP may cause damage to internal components that may not be apparent or readily visible and the devices therefore remained in service. More specifically, drops or jarring on a hard surface can cause the touchscreen input cable to loosen, resulting in unintended touchscreen behavior on the LVP, including random screen touches and/or lack of screen responsiveness to user input. Under the reported failure condition, unintended touch inputs may register at arbitrary screen locations without user interaction. These inputs occur intermittently and without a predictable pattern.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Ivenix Infusion System (IIS), Large Volume Pump, Model Number: LVP-0004

Where it was distributed

Domestic: CA, CO, FL, GA, ID, IL, MD, MI, MI , MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI, WI;

CaliforniaColoradoFloridaGeorgiaIdahoIllinoisMarylandMichiganMinnesotaMissouriMississippiNebraskaNew JerseyNevadaOklahomaOregonSouth CarolinaTexasVirginiaWashington

Questions about this recall

Why was it recalled?

Fresenius Kabi has been made aware via customer complaints that dropping or severe jarring of the Ivenix LVP may cause damage to internal components that may not be apparent or readily visible and the devices therefore remained in service. More specifically, drops or jarring on a hard surface can cause the touchscreen input cable to loosen, resulting in unintended touchscreen behavior on the LVP, including random screen touches and/or lack of screen responsiveness to user input. Under the reported failure condition, unintended touch inputs may register at arbitrary screen locations without user interaction. These inputs occur intermittently and without a predictable pattern.

Which lots are affected?

Model Number: LVP-0004; UDI-DI: 00811505030320; All Serial/Lot Numbers:

Where was it sold?

Domestic: CA, CO, FL, GA, ID, IL, MD, MI, MI , MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI, WI;

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class I mean?

Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.