CLASS II Device recall · Reported 30 Sep 2015 · FDA Enforcement Report
The Leica ASP6025 Tissue Processor recalled by Leica Microsystems
Leica Microsystems is recalling The Leica ASP6025 Tissue Processor, distributed in 34 states. The recall was initiated on 27 Jul 2015 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2804-2015
- FDA event ID
- 71928
- Recalling firm
- Leica Microsystems, Buffalo Grove, IL
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 27 Jul 2015
- FDA classified
- 22 Sep 2015
- Reported
- 30 Sep 2015
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Other Medical Devices
Timeline
| 27 Jul 2015 | Recall initiated by company | |
| 22 Sep 2015 | Classified by FDA | +57 days |
| 30 Sep 2015 | Published in enforcement report | +8 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
The ASP6025 Tissue Processor instrument is incorrectly getting into a state of bottle empty during retort filling, after passing the fill lever sensor 2 and prior to reaching level sensor 3, thus not completing a 5 litre fill.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
The Leica ASP6025 Tissue Processor. Pathology and Hematology Departments. The Leica ASP6025 Tissue Processor is an automated system used to process tissue specimens for examination through fixation, dehydration and infiltration.
Where it was distributed
Worldwide Distribution - US including AR, AZ, CA, CT, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, MN, MO, MT, NC, NE, NJ, NM, NV, NY, OH,PA, SC, TN, TX, VA, WA, WI, WV and District of Columbia. Foreign distribution to United Arab Emirates, Australia, Belgium, Brazil, Canada, Switzerland, Chile, China, Colombia, Czech Republic , Germany, Spain, France, England, Georgia, Indonesia, Ireland, Israel, India, Italy, Jordan, Japan, Korea, Kuwait, Lithuania, Myanmar, Mexico, Malaysia, Netherland, Norway, Peru, Philippines, Poland, Qatar, Reunion, Romania, Serbia, Saudi Arabia, Singapore, Slovenia, Thailand, Turkey, Taiwan, and South Africa.
ArkansasArizonaCaliforniaConnecticutDistrict of ColumbiaFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKentuckyLouisianaMassachusettsMarylandMinnesotaMissouriMontanaNorth Carolina
Questions about this recall
Why was it recalled?
The ASP6025 Tissue Processor instrument is incorrectly getting into a state of bottle empty during retort filling, after passing the fill lever sensor 2 and prior to reaching level sensor 3, thus not completing a 5 litre fill.
Which lots are affected?
Product Code: Leica ASP6025 Serial Numbers: 163, 169, 200-729, 731, 733, 735, 737, 739, 741, 743, 745, 747, 749, 751, 753, 755, 757, 759, 761, 763, 765, 767, 769, 771, 773, 775, 777, 779, 781, 783, 785, 787, 789, 791, 793, 795, 797, 799, 801, 803, 805, 807, 809, 811.
Where was it sold?
Worldwide Distribution - US including AR, AZ, CA, CT, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, MN, MO, MT, NC, NE, NJ, NM, NV, NY, OH,PA, SC, TN, TX, VA, WA, WI, WV and District of Columbia. Foreign distribution to United Arab Emirates, Australia, Belgium, Brazil, Canada, Switzerland, Chile, China, Colombia, Czech Republic , Germany, Spain, France, England, Georgia, Indonesia, Ireland, Israel, India, Italy, Jordan, Japan, Korea, Kuwait, Lithuania, Myanmar, Mexico, Malaysia, Netherland, Norway, Peru, Philippines, Poland, Qatar, Reunion, Romania, Serbia, Saudi Arabia, Singapore, Slovenia, Thailand, Turkey, Taiwan, and South Africa.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Leica Microsystems
All 40| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Leica Microsystems Proveo 8, Part Number 10449058, surgical recalled | Leica Microsystems | Other | Nationwide |
| Device | CLASS II | Provido, Part Numbers 10448950 (Premium), 10448976 (Standard) recalled | Leica Microsystems | Other | Nationwide |
| Device | CLASS II | M530 OHX recalled by Leica Microsystems | Leica Microsystems | Other | Nationwide |
| Device | CLASS II | Leica Surgical Operating Microscopes, PROvido Part Numbers recalled | Leica Microsystems | Labeling Errors | Nationwide |
| Device | CLASS II | Leica Surgical Operating Microscopes, Arveo recalled | Leica Microsystems | Labeling Errors | Nationwide |
Other Other Medical Devices recalls
All 13,884| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | NOxBOXi Nitric Oxide Delivery System recalled | Noxbox | Device Malfunction | Nationwide |
| Device | CLASS II | Palacos R pro recalled by Heraeus Medical GmbH (Dental | Heraeus Medical GmbH (Dental Division) | Packaging Defects | Nationwide |
| Device | CLASS II | Intrapulmonary Percussive Ventilation, IPV 1, Product Number: S00065 recalled | Percussionaire | Other | Nationwide |
| Device | CLASS II | Positron Emission Tomography and Computed Tomography System recalled | Shanghai United Imaging Healthcare | Device Malfunction | Nationwide |
| Device | CLASS II | Intrapulmonary Percussive Ventilation, IPV-1C recalled by Percussionaire | Percussionaire | Other | Nationwide |