CLASS II ONGOING Device recall · Reported 10 Nov 2021 · FDA Enforcement Report

Lisendo 880 Ultrasound recalled by FujiFilm Healthcare Americas

FujiFilm Healthcare Americas is recalling Lisendo 880 Ultrasound, distributed nationwide in the US. The recall was initiated on 14 Oct 2021 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSerial Numbers/UDI: 20514657 (01)04562122958868(11)161220(21)20514657; 20659147 (01)04562122958868(11)20181124(21)20659147; 20659148 (01)04562122958868(11)20181124(21)20659148; 20659149 (01)04562122958868(11)20181124(21)20659149; 20659150 (01)04562122958868(11)20181124(21)20659150; 20659153 (01)04562122958868(11)20181126(21)20659153; 20659154 (01)04562122958868(11)20181126(21)20659154; 20659155 (01)04562122958868(11)20181126(21)20659155; 20659156 (01)04562122958868(11)20181126(21)20659156; 205G0225 (01)04562122958868(11)20170418(21)205G0225; 205G0226 (01)04562122958868(11)20170424(21)205G0226; 205P2146 (01)04562122958868(11)20170615(21)205P2146; 205P2147 (01)04562122958868(11)20170615(21)205P2147; 205P2148 (01)04562122958868(11)20170615(21)205P2148; 205P2149 (01)04562122958868(11)20170615(21)205P2149; 205Q0809 (01)04562122958868(11)20170707(21)205Q0809; 205Q0810 (01)04562122958868(11)20170707(21)205Q0810; 205Q0812 (01)04562122958868(11)20170707(21)205Q0812; 205Q0815 (01)04562122958868(11)20171006(21)205Q0815; 205T7139 (01)04562122958868(11)20171006(21)205T7139; 205T7143 (01)04562122958868(11)20171026(21)205T7143; 205V5544 (01)04562122958868(11)20171026(21)205V5544; 205V5545 (01)04562122958868(11)20171030(21)205V5545; 205V5547 (01)04562122958868(11)20171031(21)205V5547; 205V5548 (01)04562122958868(11)20171031(21)205V5548; G3005239 (01)04562122958868(11)20190308(21)G3005239; G3005240 (01)04562122958868(11)20190308(21)G3005240; G3025041 (01)04562122958868(11)20190603(21)G3025041; G3025042 (01)04562122958868(11)20190604(21)G3025042; G3025044 (01)04562122958868(11)20190605(21)G3025044; G3025045 (01)04562122958868(11)20190605(21)G3025045; G3025048 (01)04562122958868(11)20190606(21)G3025048; G3025049 (01)04562122958868(11)20190606(21)G3025049; G3037183 (01)04562122958868(11)20190624(21)G3037183; G3037184 (01)04562122958868(11)20190624(21)G3037184; G3037192 (01)04562122958868(11)20190709(21)G3037192; G3043580 (01)04562122958868(11)20190820(21)G3043580; G3043581 (01)04562122958868(11)20190820(21)G3043581; G3043582 (01)04562122958868(11)20190820(21)G3043582; G3052217 (01)04562122958868(11)20191016(21)G3052217; G3052218 (01)04562122958868(11)20191016(21)G3052218; G3052225 (01)04562122958868(11)20191202(21)G3052225; G3052234 (01)04562122958868(11)20200128(21)G3052234; G3052236 (01)04562122958868(11)20200109(21)G3052236; G3054831 (01)04562122958868(11)20200115(21)G3054831; G3054833 (01)04562122958868(11)20191212(21)G3054833; G3054840 (01)04562122958868(11)20200115(21)G3054840; G3080616 (01)04562122958868(11)20201215(21)G3080616; G3080617 (01)04562122958868(11)20201215(21)G3080617; G3106256 (01)04562122958868(11)20210413(21)G3106256; G3106258 (01)04562122958868(11)20210413(21)G3106258; G3111572 (01)04562122958868(11)20210413(21)G3111572; G3111573 (01)04562122958868(11)20210421(21)G3111573
Quantity recalledN/A

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0191-2022
FDA event ID
88840
Recalling firm
FujiFilm Healthcare Americas, Twinsburg, OH
Classification
Class II
Status
Ongoing
Recall initiated
14 Oct 2021
FDA classified
2 Nov 2021
Reported
10 Nov 2021
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

14 Oct 2021Recall initiated by company
2 Nov 2021Classified by FDA+19 days
10 Nov 2021Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Diagnostic ultrasound system with the specified software version and used in conjunction with a Fujifilm (formally Hitachi) transducer does not display the measurement results correctly with two cardiology measurement functions: Mitral Regurgitation Flow (MR Flow) Measurement; Mitral Valve (MV) Measurement. In addition there is an error in the instruction manual.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Lisendo 880 Ultrasound Software Version: V1.0 through V4.1.3

Where it was distributed

Nationwide

Nationwide

Questions about this recall

Why was it recalled?

Diagnostic ultrasound system with the specified software version and used in conjunction with a Fujifilm (formally Hitachi) transducer does not display the measurement results correctly with two cardiology measurement functions: Mitral Regurgitation Flow (MR Flow) Measurement; Mitral Valve (MV) Measurement. In addition there is an error in the instruction manual.

Which lots are affected?

Serial Numbers/UDI: 20514657 (01)04562122958868(11)161220(21)20514657; 20659147 (01)04562122958868(11)20181124(21)20659147; 20659148 (01)04562122958868(11)20181124(21)20659148; 20659149 (01)04562122958868(11)20181124(21)20659149; 20659150 (01)04562122958868(11)20181124(21)20659150; 20659153 (01)04562122958868(11)20181126(21)20659153; 20659154 (01)04562122958868(11)20181126(21)20659154; 20659155 (01)04562122958868(11)20181126(21)20659155; 20659156 (01)04562122958868(11)20181126(21)20659156; 205G0225 (01)04562122958868(11)20170418(21)205G0225; 205G0226 (01)04562122958868(11)20170424(21)205G0226; 205P2146 (01)04562122958868(11)20170615(21)205P2146; 205P2147 (01)04562122958868(11)20170615(21)205P2147; 205P2148 (01)04562122958868(11)20170615(21)205P2148; 205P2149 (01)04562122958868(11)20170615(21)205P2149; 205Q0809 (01)04562122958868(11)20170707(21)205Q0809; 205Q0810 (01)04562122958868(11)20170707(21)205Q0810; 205Q0812 (01)04562122958868(11)20170707(21)205Q0812; 205Q0815 (01)04562122958868(11)20171006(21)205Q0815; 205T7139 (01)04562122958868(11)20171006(21)205T7139; 205T7143 (01)04562122958868(11)20171026(21)205T7143; 205V5544 (01)04562122958868(11)20171026(21)205V5544; 205V5545 (01)04562122958868(11)20171030(21)205V5545; 205V5547 (01)04562122958868(11)20171031(21)205V5547; 205V5548 (01)04562122958868(11)20171031(21)205V5548; G3005239 (01)04562122958868(11)20190308(21)G3005239; G3005240 (01)04562122958868(11)20190308(21)G3005240; G3025041 (01)04562122958868(11)20190603(21)G3025041; G3025042 (01)04562122958868(11)20190604(21)G3025042; G3025044 (01)04562122958868(11)20190605(21)G3025044; G3025045 (01)04562122958868(11)20190605(21)G3025045; G3025048 (01)04562122958868(11)20190606(21)G3025048; G3025049 (01)04562122958868(11)20190606(21)G3025049; G3037183 (01)04562122958868(11)20190624(21)G3037183; G3037184 (01)04562122958868(11)20190624(21)G3037184; G3037192 (01)04562122958868(11)20190709(21)G3037192; G3043580 (01)04562122958868(11)20190820(21)G3043580; G3043581 (01)04562122958868(11)20190820(21)G3043581; G3043582 (01)04562122958868(11)20190820(21)G3043582; G3052217 (01)04562122958868(11)20191016(21)G3052217; G3052218 (01)04562122958868(11)20191016(21)G3052218; G3052225 (01)04562122958868(11)20191202(21)G3052225; G3052234 (01)04562122958868(11)20200128(21)G3052234; G3052236 (01)04562122958868(11)20200109(21)G3052236; G3054831 (01)04562122958868(11)20200115(21)G3054831; G3054833 (01)04562122958868(11)20191212(21)G3054833; G3054840 (01)04562122958868(11)20200115(21)G3054840; G3080616 (01)04562122958868(11)20201215(21)G3080616; G3080617 (01)04562122958868(11)20201215(21)G3080617; G3106256 (01)04562122958868(11)20210413(21)G3106256; G3106258 (01)04562122958868(11)20210413(21)G3106258; G3111572 (01)04562122958868(11)20210413(21)G3111572; G3111573 (01)04562122958868(11)20210421(21)G3111573

Where was it sold?

Nationwide

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

2 other products · FDA event 88840

This product is part of a recall event; the main page for the event is Arietta 850 Ultrasound recalled by FujiFilm Healthcare Americas.

More recalls from FujiFilm Healthcare Americas

All 24

Other Imaging Equipment recalls

All 2,336