CLASS I ONGOING Device recall · Reported 10 Jun 2026 · FDA Enforcement Report
LVP software of the Ivenix Infusion System (IIS); Product Code recalled
Fresenius Kabi USA is recalling LVP software of the Ivenix Infusion System (IIS); Product Code, distributed nationwide in the US. The recall was initiated on 6 May 2026 and the FDA has classified it as Class I.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2223-2026
- FDA event ID
- 98929
- Recalling firm
- Fresenius Kabi USA, North Andover, MA
- Classification
- Class I
- Status
- Ongoing
- Recall initiated
- 6 May 2026
- FDA classified
- 3 Jun 2026
- Reported
- 10 Jun 2026
- Reason category
- Device Malfunction
- Device type
- Medical Software
Timeline
| 6 May 2026 | Recall initiated by company | |
| 3 Jun 2026 | Classified by FDA | +28 days |
| 10 Jun 2026 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Potential software anomaly that can cause the pump to report a false battery-health value (typically 69%) and trigger a battery-depletion alarm that will shut down the pump unless it is reconnected to AC mains power.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
LVP software of the Ivenix Infusion System (IIS); Product Code LVP-SW-0005.
Where it was distributed
US Nationwide distribution in the states of CA, CO, FL, GA, GA, ID, IL, MD, MI, MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI.
NationwideCaliforniaColoradoFloridaGeorgiaIdahoIllinoisMarylandMichiganMinnesotaMissouriMississippiNebraskaNew JerseyNevadaOklahomaOregonSouth CarolinaTexasVirginiaWashington
Questions about this recall
Why was it recalled?
Potential software anomaly that can cause the pump to report a false battery-health value (typically 69%) and trigger a battery-depletion alarm that will shut down the pump unless it is reconnected to AC mains power.
Which lots are affected?
Product Code LVP-SW-0005. UDI-DI: 00811505030122. Software Version 5.10.2
Where was it sold?
US Nationwide distribution in the states of CA, CO, FL, GA, GA, ID, IL, MD, MI, MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class I mean?
Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Fresenius Kabi USA
All 95| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Drug | CLASS I | Tyenne (tocilizumab-aazg) Injection, 400 mg/ (20 mg/mL) vial recalledTyenne | Fresenius Kabi USA | Particulate Matter | Nationwide |
| Drug | CLASS I | Morphine Sulfate Injection recalled by Fresenius Kabi USAMorphine Sulfate | Fresenius Kabi USA | Packaging Defects | Nationwide |
| Drug | CLASS II | Glycopyrrolate Injection, USP, 1 mg per (0.2 mg per mL) recalledGlycopyrrolate | Fresenius Kabi USA | Particulate Matter | Nationwide |
| Drug | CLASS II | Xylocaine (lidocaine HCl Injection, USP), 1%, 500 mg recalledXylocaine | Fresenius Kabi USA | Particulate Matter | Nationwide |
| Drug | CLASS II | Lidocaine HCl Injection, USP, 2%, 100 mg recalled by Fresenius Kabi USALidocaine | Fresenius Kabi USA | Particulate Matter | Nationwide |
Other Medical Software recalls
All 946| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | OCS Lung Mission Critical Plus Perfusion Set recalled by TransMedics | TransMedics | Device Malfunction | Nationwide |
| Device | CLASS II | OCS Heart Perfusion Module, included recalled by TransMedics | TransMedics | Device Malfunction | Nationwide |
| Device | CLASS II | Nitric Oxide Calibration Gas recalled by Beyond Air | Beyond Air | Device Malfunction | Nationwide |
| Device | CLASS II | Physitemp Sst-1 Skin Sensor recalled by Physitemp Instruments | Physitemp Instruments | Temperature Abuse | Nationwide |
| Device | CLASS II | CareLink SmartSync Device Manager App recalled by Medtronic | Medtronic | Device Malfunction | Nationwide |