CLASS I ONGOING Device recall · Reported 10 Jun 2026 · FDA Enforcement Report

LVP software of the Ivenix Infusion System (IIS); Product Code recalled

Fresenius Kabi USA is recalling LVP software of the Ivenix Infusion System (IIS); Product Code, distributed nationwide in the US. The recall was initiated on 6 May 2026 and the FDA has classified it as Class I.

Check your package

Lot, batch or date codesProduct Code LVP-SW-0005. UDI-DI: 00811505030122. Software Version 5.10.2
Quantity recalled32 systems

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2223-2026
FDA event ID
98929
Recalling firm
Fresenius Kabi USA, North Andover, MA
Classification
Class I
Status
Ongoing
Recall initiated
6 May 2026
FDA classified
3 Jun 2026
Reported
10 Jun 2026
Reason category
Device Malfunction
Device type
Medical Software

Timeline

6 May 2026Recall initiated by company
3 Jun 2026Classified by FDA+28 days
10 Jun 2026Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Potential software anomaly that can cause the pump to report a false battery-health value (typically 69%) and trigger a battery-depletion alarm that will shut down the pump unless it is reconnected to AC mains power.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

LVP software of the Ivenix Infusion System (IIS); Product Code LVP-SW-0005.

Where it was distributed

US Nationwide distribution in the states of CA, CO, FL, GA, GA, ID, IL, MD, MI, MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI.

NationwideCaliforniaColoradoFloridaGeorgiaIdahoIllinoisMarylandMichiganMinnesotaMissouriMississippiNebraskaNew JerseyNevadaOklahomaOregonSouth CarolinaTexasVirginiaWashington

Questions about this recall

Why was it recalled?

Potential software anomaly that can cause the pump to report a false battery-health value (typically 69%) and trigger a battery-depletion alarm that will shut down the pump unless it is reconnected to AC mains power.

Which lots are affected?

Product Code LVP-SW-0005. UDI-DI: 00811505030122. Software Version 5.10.2

Where was it sold?

US Nationwide distribution in the states of CA, CO, FL, GA, GA, ID, IL, MD, MI, MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class I mean?

Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Fresenius Kabi USA

All 95

Other Medical Software recalls

All 946