CLASS II Device recall · Reported 13 Jul 2016 · FDA Enforcement Report
Magnetic Resonance System G-scan Brio recalled by Esaote S.p.A
Esaote is recalling Magnetic Resonance System G-scan Brio (, distributed in Florida, Minnesota, Nevada, Ohio, Texas, Virginia. The recall was initiated on 6 Jun 2016 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2124-2016
- FDA event ID
- 74391
- Recalling firm
- Esaote, Firenze, N/A
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 6 Jun 2016
- FDA classified
- 5 Jul 2016
- Reported
- 13 Jul 2016
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Other Medical Devices
Timeline
| 6 Jun 2016 | Recall initiated by company | |
| 5 Jul 2016 | Classified by FDA | +29 days |
| 13 Jul 2016 | Published in enforcement report | +8 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
The device is being recalled because dynamic loading forces from larger patients can fracture the extremity portion of the patient table. Additionally, a weight limit is added to the patient seat for weight-bearing examinations.
Product as listed by the FDA
Magnetic Resonance System G-scan Brio (Part Number /Model No.101002000)
Where it was distributed
Domestic: FL, TX, OH, MN, NV; Foreign: No customers in Canada; No VA/DOD.
Questions about this recall
Why was it recalled?
The device is being recalled because dynamic loading forces from larger patients can fracture the extremity portion of the patient table. Additionally, a weight limit is added to the patient seat for weight-bearing examinations.
Which lots are affected?
Item No., Unit Description and Serial No.s: 101002000, UNIT, G-SCAN BRIO, 05202; 101002000, UNIT, G-SCAN BRIO, 05203; 101002000, UNIT, G-SCAN BRIO 2.1A; 05205; 101002000, UNIT, G-SCAN BRIO 2.1A, 05207; 101002000 UNIT, G-SCAN BRIO 2.1A, 05208; 101002000 UNIT, G-SCAN BRIO 2.1A, 05209; 101002000 UNIT, G-SCAN BRIO 2.1A, 05210; 101002000 UNIT, G-SCAN BRIO 2.1A, 05211; 101002000 UNIT, G-SCAN BRIO 2.1A, 05305; 101002000 UNIT, G-SCAN BRIO 2.1A, 05308
Where was it sold?
Domestic: FL, TX, OH, MN, NV; Foreign: No customers in Canada; No VA/DOD.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Esaote
All 9| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Endocavity ultrasonic probe recalled by Esaote S.p.a | Esaote | Packaging Defects | Nationwide |
| Device | CLASS II | AC/DC Power Supply recalled by Esaote S.p.a | Esaote | Other | Nationwide |
| Device | CLASS II | AC/DC Power Supply recalled by Esaote S.p.a | Esaote | Other | Nationwide |
| Device | CLASS II | MyLab Seven recalled by Esaote S.p.a | Esaote | Device Malfunction | Nationwide |
| Device | CLASS II | MyLab Alpha recalled by Esaote S.p.a | Esaote | Device Malfunction | Nationwide |
Other Other Medical Devices recalls
All 13,929| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Rapicide PA Part A High-Level Disinfectant recalled by Steris | Steris | Other | Nationwide |
| Device | CLASS II | NeuroStar Advanced Therapy System recalled by Neuronetics | Neuronetics | Other | Nationwide |
| Device | CLASS II | TS Misc. Instruments Tray, Triathlon recalled by Howmedica Osteonics | Howmedica Osteonics | Sterility | NJ |
| Device | CLASS II | BD Phoenix Panel Nmic recalled by Becton Dickinson | Becton Dickinson | Device Malfunction | Nationwide |
| Device | CLASS II | BD Phoenix Panel Nmic recalled by Becton Dickinson | Becton Dickinson | Device Malfunction | Nationwide |