CLASS II Device recall · Reported 18 Nov 2015 · FDA Enforcement Report

Maquet Getinge Group, Universal Frame recalled

Maquet Cardiovascular Us Sales is recalling Maquet Getinge Group, Universal Frame for Positioning Patients (1007.24XX) Manufacturer, distributed nationwide in the US. The recall was initiated on 17 Aug 2015 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesPart Nos 1007.28A0 (Adjustable Positioning Elements and 1007.24F0 (Universal Frame with Adjustable Positioning Elements)
Quantity recalledUnited States (3 sets); Rest of World (47 sets)

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0248-2016
FDA event ID
71998
Recalling firm
Maquet Cardiovascular Us Sales, Wayne, NJ
Classification
Class II
Status
Terminated
Recall initiated
17 Aug 2015
FDA classified
9 Nov 2015
Reported
18 Nov 2015
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

17 Aug 2015Recall initiated by company
9 Nov 2015Classified by FDA+84 days
18 Nov 2015Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Two serious thromboembolic events were reported involving the Adjustable Positioning Elements (Part Number 1007.28A0) in conjunction with Universal Frame (Part Number 1007.24F0) for lumbar spine surgery. The hip elements may have a negative effect on patients- venous backflow under certain conditions and may have been a contributing factor to thromboembolic events.

Product as listed by the FDA

Maquet Getinge Group, Universal Frame for Positioning Patients (1007.24XX) Manufacturer: Maquet GmbH, Germany The MAQUET Adjustable Positioning Elements are designed for the placement and positioning of patients in the prone position immediately before, during and after surgical interventions as well as for examination and treatment. With a patient in the prone position, the adjustable positioning elements stabilize the upper body while simultaneously relieving the abdominal region in the vertical position with regard to the spine and the large blood vessels there. This positioning method is thus suitable for interventions on the spine with dorsal access to prevent positioning-specific compression with resulting bleeding tenancies. They are made up of the following components: (2) chest and (2) hip elements with padding which are secured in a fixed arrangement to the frames and can be adjusted in all directions independently of each other.

Where it was distributed

Worldwide Distribution - US including AL, CA, MO and Internationally to Australia, Brazil, Canada, China, Croatia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Russia, Singapore, South Africa and United Arab Emirates.

Nationwide AlabamaCaliforniaMissouri

Questions about this recall

Is Maquet Getinge Group, Universal Frame for Positioning Patients (1007.24XX) Manufacturer recalled?

Yes. Maquet Cardiovascular Us Sales recalled Maquet Getinge Group, Universal Frame for Positioning Patients (1007.24XX) Manufacturer on 17 Aug 2015. The recall is Class II and its status is Terminated. Recall number Z-0248-2016.

Why was it recalled?

Two serious thromboembolic events were reported involving the Adjustable Positioning Elements (Part Number 1007.28A0) in conjunction with Universal Frame (Part Number 1007.24F0) for lumbar spine surgery. The hip elements may have a negative effect on patients- venous backflow under certain conditions and may have been a contributing factor to thromboembolic events.

Which lots are affected?

Part Nos 1007.28A0 (Adjustable Positioning Elements and 1007.24F0 (Universal Frame with Adjustable Positioning Elements)

Where was it sold?

Worldwide Distribution - US including AL, CA, MO and Internationally to Australia, Brazil, Canada, China, Croatia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Russia, Singapore, South Africa and United Arab Emirates.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Maquet Cardiovascular Us Sales

All 62

Other Hip & Knee Implants recalls

All 3,097