CLASS II ONGOING Device recall · Reported 23 Apr 2025 · FDA Enforcement Report
Maquet VH-3010 Power Supply recalled by Maquet Cardiovascular
Maquet Cardiovascular is recalling Maquet VH-3010 Power Supply, distributed nationwide in the US. The recall was initiated on 10 Mar 2025 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1577-2025
- FDA event ID
- 96392
- Recalling firm
- Maquet Cardiovascular, Wayne, NJ
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 10 Mar 2025
- FDA classified
- 16 Apr 2025
- Reported
- 23 Apr 2025
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Other Medical Devices
Timeline
| 10 Mar 2025 | Recall initiated by company | |
| 16 Apr 2025 | Classified by FDA | +37 days |
| 23 Apr 2025 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Incorrect resistor utilized in the VH-3010 Power Supply, which may cause the jaws of the HemoPro 2 to not adequately heat up.
Product as listed by the FDA
Maquet VH-3010 Power Supply. Used to deliver power to the harvesting tool of the Vasoview Hemopro EVH System (HemoPro 2).
Where it was distributed
Worldwide - US Nationwide distribution in the states of AL, CA, CO, DC, FL, GA, IA, ID, IL, IN, KY, MD, MI, MN, MO, NC, NH, NJ, NY, OH, PA, PR, TX, UT, VA, WI, WV and the countries of China, Japan, Netherlands, New Zealand, Singapore, United Kingdom, United States.
Nationwide AlabamaCaliforniaColoradoDistrict of ColumbiaFloridaGeorgiaIowaIdahoIllinoisIndianaKentuckyMarylandMichiganMinnesotaMissouriNorth CarolinaNew HampshireNew JerseyNew YorkOhio
Questions about this recall
Why was it recalled?
Incorrect resistor utilized in the VH-3010 Power Supply, which may cause the jaws of the HemoPro 2 to not adequately heat up.
Which lots are affected?
Model No. VH-3010; UDI-DI: 0060756770826; Serial No. H24080065G, H24080074G, H24080012G, H24080022G, H24080184G, H24080185G, H24080051G, H24080117G, H24080069G, H24080016G, H24080079G, H24080041G, H24080056G, H24080030G, H24080068G, H24080034G, H24080058G, H24080026G, H24080033G, H24080020G, H24080032G, H24080053G, H24080066G, H24080077G, H24080144G, H24080123G, H24080042G, H24080070G, H24080008G, H24080003G, H24080174G, H24080064G, H24080027G, H24080044G, H24080047G, H24080049G, H24080061G, H24080078G, H24080010G, H24080138G, H24080080G, H24080006G, H24080015G, H24080038G, H24080085G, H24080084G, H24080031G, H24080101G, H24080148G, H24080040G, H24080055G, H24080025G, H24080052G, H24080048G, H24080093G, H24080126G, H24080073G, H24080007G, H23060082G, H24080075G, H24080113G, H24080127G, H24080011G, H24080005G, H24080013G, H24080023G, H24080057G, H24080110G, H24080009G, H24080067G, H24080029G, H24080043G, H24080183G, H24080165G, H24080017G, H24080001G, H24080037G, H24080060G, H24080072G, H24080076G, H24080045G, H24080024G, H24080028G, H24080114G, H24080136G, H24080062G, H24080018G, H24080054G, H24080063G, H24080112G, H24080118G, H24080071G, H24080182G, H24080002G, H24080004G, H24080014G, H24080019G, H24080021G, H24080035G, H24080036G, H24080039G, H24080046G, H24080050G, H24080059G, H24080081G, H24080082G, H24080083G, H24080125G, H24080186G, H24080189G.
Where was it sold?
Worldwide - US Nationwide distribution in the states of AL, CA, CO, DC, FL, GA, IA, ID, IL, IN, KY, MD, MI, MN, MO, NC, NH, NJ, NY, OH, PA, PR, TX, UT, VA, WI, WV and the countries of China, Japan, Netherlands, New Zealand, Singapore, United Kingdom, United States.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Maquet Cardiovascular
All 252| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | The Vasoview Hemopro 2 Endoscopic Vessel Harvesting System recalled | Maquet Cardiovascular | Sterility | Nationwide |
| Device | CLASS II | Vasoview Hemopro 3 Endoscopic Vessel Harvesting System recalled | Maquet Cardiovascular | Device Malfunction | Nationwide |
| Device | CLASS II | Heartstring III Proximal Seal System Intravascular recalled | Maquet Cardiovascular | Packaging Defects | Nationwide |
| Device | CLASS II | Heartstring III Proximal Seal System Intravascular recalled | Maquet Cardiovascular | Packaging Defects | Nationwide |
| Device | CLASS II | Heartstring III Proximal Seal System recalled by Maquet Cardiovascular | Maquet Cardiovascular | Packaging Defects | Nationwide |
Other Other Medical Devices recalls
All 13,950| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline Drape C-section W/Pouch ST 7/CS, Medline Product Number recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Medline medical convenience kits labeled as recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Medline medical convenience kits labeled as: Fetal Surgery Cds-lf recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Medline Drape C-section Fen Clr Screen W, Medline Product Number recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Medline medical convenience kits labeled as: C-section Pack recalled | Medline Industries | Other | Nationwide |