CLASS II Device recall · Reported 31 Jan 2018 · FDA Enforcement Report

Mars Treatment Kit, Use only with Mars(R) Monitor 1TC recalled

Baxter Healthcare is recalling Mars Treatment Kit, Use only with Mars(R) Monitor 1TC, Gambro. The recall was initiated on 21 Dec 2017 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI: 37332414107728; Lot Numbers: 0000022597, 0000022648, 0000022663, 0000022671, 0000022721, 0000022767
Quantity recalled186 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0411-2018
FDA event ID
78835
Recalling firm
Baxter Healthcare, Deerfield, IL
Classification
Class II
Status
Terminated
Recall initiated
21 Dec 2017
FDA classified
24 Jan 2018
Reported
31 Jan 2018
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Packaging Defects
Device type
Patient Monitors

Timeline

21 Dec 2017Recall initiated by company
24 Jan 2018Classified by FDA+34 days
31 Jan 2018Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Baxter has received customer complaints regarding leakage in the albumin circuit. The leakage was caused by an inadequate adhesive connection of the tubing to the Hansen connector of the MARS Tube Set which is part of the MARS Treatment Kit.

Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.

Product as listed by the FDA

MARS Treatment Kit, Use only with MARS(R) Monitor 1TC, GAMBRO, Product Code 800541

Where it was distributed

US, Germany, Sweden, Italy, Belgium, Portugal, Spain, France, Russia, Poland, Turkey, China, Thailand, Taiwan, Hong Kong, Malaysia, Korea, and Mexico.

Questions about this recall

Why was it recalled?

Baxter has received customer complaints regarding leakage in the albumin circuit. The leakage was caused by an inadequate adhesive connection of the tubing to the Hansen connector of the MARS Tube Set which is part of the MARS Treatment Kit.

Which lots are affected?

UDI: 37332414107728; Lot Numbers: 0000022597, 0000022648, 0000022663, 0000022671, 0000022721, 0000022767

Where was it sold?

US, Germany, Sweden, Italy, Belgium, Portugal, Spain, France, Russia, Poland, Turkey, China, Thailand, Taiwan, Hong Kong, Malaysia, Korea, and Mexico.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.