CLASS II Device recall · Reported 13 Feb 2013 · FDA Enforcement Report

Medacta Quadra Trial Broach size 0 part # recalled

Medacta Usa is recalling Medacta Quadra Trial Broach size 0 part # 01.10.10.045, for use with the Medacta Total, distributed nationwide in the US. The recall was initiated on 17 Dec 2012 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesLot/Unit Codes: 075023, 075064, 084000, 084001, 084326, 084743, 084815, 085202, 085455
Quantity recalled26 pieces U.S.A. (226 pieces Worldwide)

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0791-2013
FDA event ID
64160
Recalling firm
Medacta Usa, Camarillo, CA
Classification
Class II
Status
Terminated
Recall initiated
17 Dec 2012
FDA classified
6 Feb 2013
Reported
13 Feb 2013
Type
Voluntary: Firm initiated
Customers notified by
E-Mail
Reason category
Other Reasons

Timeline

17 Dec 2012Recall initiated by company
6 Feb 2013Classified by FDA+51 days
13 Feb 2013Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The tip of the Quadra Trial Broach size 0, broke during a surgery.

Product as listed by the FDA

Medacta Quadra Trial Broach size 0 part # 01.10.10.045, for use with the Medacta Total Hip Prosthesis System. Product Usage - used to broach the intramedullary canal of the femur in order to create space for the femoral stem during a hip replacement surgery.

Where it was distributed

US Nationwide Distribution including the states of Arizona, California, Idaho, Illinois, North Carolina, Ohio, Utah, Washington and US Virgin Islands

Nationwide ArizonaCaliforniaIdahoIllinoisNorth CarolinaOhioUtahWashington

Questions about this recall

Why was it recalled?

The tip of the Quadra Trial Broach size 0, broke during a surgery.

Which lots are affected?

Lot/Unit Codes: 075023, 075064, 084000, 084001, 084326, 084743, 084815, 085202, 085455

Where was it sold?

US Nationwide Distribution including the states of Arizona, California, Idaho, Illinois, North Carolina, Ohio, Utah, Washington and US Virgin Islands

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Medacta Usa

All 20

Other Hip & Knee Implants recalls

All 3,111