CLASS II Device recall · Reported 8 Jan 2020 · FDA Enforcement Report
Medicrea Pass LP recalled by Medicrea International
Medicrea International is recalling Medicrea Pass LP, distributed in 7 states. The recall was initiated on 27 Jun 2019 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0753-2020
- FDA event ID
- 84273
- Recalling firm
- Medicrea International, Rillieux La Pape, N/A
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 27 Jun 2019
- FDA classified
- 30 Dec 2019
- Reported
- 8 Jan 2020
- Type
- Voluntary: Firm initiated
- Customers notified by
- Reason category
- Other Reasons
- Device type
- Spinal Implants
Timeline
| 27 Jun 2019 | Recall initiated by company | |
| 30 Dec 2019 | Classified by FDA | +186 days |
| 8 Jan 2020 | Published in enforcement report | +9 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
It has been determined that some Pedicle system screw identified as 45mm long have an actual length of 50mm. Pedicle screws with a diameter of 7.5 mm or more are reserved for the sacrum area. As a result, a screw that is too long could come into contact with the nerves and cause postoperative pain.
Product as listed by the FDA
Medicrea Pass LP, REF B02217545, Non Sterile, Material: Ti-6Al-4V ELI, VIS PEDICULAIRE POLYAXIALE 7.5MM x 45MM, POLYAXIAL PEDICLE SCREW 7.5MM x 45 MM, POLYAXIAL PEDIKELSCHRAUBE 7.5MM X 45MM, TORNILLO PEDICULAR POLIAXIAL 7.5MM x 45MM VITE POLIASSIALE PEDUNCOLARE 7.5MM x 45MM, PARAFUSO PEDICULAR POLIAXIAL 7.5MM x 45MM, POLYAXIALE PEDICULAIRE SCHROEF 7.5MM x 45MM, POLYAXIALE PEDICULAIRE SCHROEF 7.5MM x 45MM, WIELOOSIOWA SRUBA TRANSPEDIKULARNA 7.5MM x 45MM, (01)03613720177906(10)19A0630(17)290103, RX only
Where it was distributed
US: NY, CO, OH, AZ, FL, TX, PA OUS: France, Belgium, South Africa, China, Spain, Portugal
Questions about this recall
Why was it recalled?
It has been determined that some Pedicle system screw identified as 45mm long have an actual length of 50mm. Pedicle screws with a diameter of 7.5 mm or more are reserved for the sacrum area. As a result, a screw that is too long could come into contact with the nerves and cause postoperative pain.
Which lots are affected?
Lot #/Expiration Date: 19A0630/January 03, 2029 19B0329 /January 21th, 2029 19C0734 /March 03, 2029
Where was it sold?
US: NY, CO, OH, AZ, FL, TX, PA OUS: France, Belgium, South Africa, China, Spain, Portugal
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Medicrea International
All 22| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medicrea, IB3D Universal Implant Inserter recalled | Medicrea International | Other | Nationwide |
| Device | CLASS II | Medicrea, IB3D PL Instruments Set recalled by Medicrea International | Medicrea International | Other | Nationwide |
| Device | CLASS II | UNiD" Adaptive Spine Intelligence UNiD Spine Analyzer recalled | Medicrea International | Device Malfunction | 50 states |
| Device | CLASS II | Impix Alif S/a recalled by Medicrea International | Medicrea International | Packaging Defects | Nationwide |
| Device | CLASS II | PASS LP recalled by Medicrea International | Medicrea International | Packaging Defects | Nationwide |
Other Spinal Implants recalls
All 1,161| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline surgical convenience kits labeled as: Spinal CDS recalled | Medline Industries | Packaging Defects | Nationwide |
| Device | CLASS II | Medline surgical convenience kits labeled as recalled | Medline Industries | Packaging Defects | Nationwide |
| Device | CLASS II | Extended Polyaxial Head, Attached Gimbal recalled | Seaspine Orthopedics | Other | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Anterior Spine Pack recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Neuro; Neuro; Crani; Ortho Spine Pack; Bmhm Neuro Pack recalled | Avid Medical | Sterility | Nationwide |