CLASS II ONGOING Device recall · Reported 2 Oct 2024 · FDA Enforcement Report
MR 5300 - For use as a diagnostic device to obtain recalled
Philips North America is recalling MR 5300 - For use as a diagnostic device to obtain cross-sectional images, spectroscopic, distributed nationwide in the US. The recall was initiated on 3 Sep 2024 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-3201-2024
- FDA event ID
- 95289
- Recalling firm
- Philips North America, Cambridge, MA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 3 Sep 2024
- FDA classified
- 20 Sep 2024
- Reported
- 2 Oct 2024
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Lead & Heavy Metals
- Device type
- Test Kits & Diagnostics
Timeline
| 3 Sep 2024 | Recall initiated by company | |
| 20 Sep 2024 | Classified by FDA | +17 days |
| 2 Oct 2024 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop. May lead to delay in diagnosis and/or may include anxiety of a patient is left on the tabletop inside the bore for a longer duration than anticipated. movement.
Lead and other heavy metals accumulate in the body. Children are most at risk; even low levels can affect development.
Product as listed by the FDA
MR 5300 - For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 782110
Where it was distributed
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Iceland, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Jordan, Kuwait, Latvia, Maldives, Mauritius, Mexico, Morocco, Netherlands, Pakistan, Peru, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Switzerland, Thailand, T¿rkiye, United Kingdom, Utd.Arab.Emir., Uzbekistan, Vietnam.
Questions about this recall
Is MR 5300 - For use as a diagnostic device to obtain cross-sectional images, spectroscopic recalled?
Yes. Philips North America recalled MR 5300 - For use as a diagnostic device to obtain cross-sectional images, spectroscopic on 3 Sep 2024. The recall is Class II and its status is Ongoing. Recall number Z-3201-2024.
Why was it recalled?
Integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop. May lead to delay in diagnosis and/or may include anxiety of a patient is left on the tabletop inside the bore for a longer duration than anticipated. movement.
Which lots are affected?
UDI-DI: (01)00884838099364(21) Serial Numbers: 63027 63035 63039 63066 63068 63070 63076 63078 63083 63084 63086 63092 63094 63096 63097 63103 63104 63106 63109 63112 63113 63115 63122 63123 63125 63128 63130 63132 63135 63137 63140 63141 63142 63143 63144 63146 63148 63149 63150 63151 63152 63154 63157 63160 63161 63168 63169 63172 63174 63177 63178 63183 63186 63190 63191 63193 63197 63201 63203 63211 63229 63238 63248 63263 63272 63288 63367 63370
Where was it sold?
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Iceland, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Jordan, Kuwait, Latvia, Maldives, Mauritius, Mexico, Morocco, Netherlands, Pakistan, Peru, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Switzerland, Thailand, T¿rkiye, United Kingdom, Utd.Arab.Emir., Uzbekistan, Vietnam.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Philips North America
All 531Other Test Kits & Diagnostics recalls
All 2,656| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | STA Staclot Protein S recalled by Diagnostica Stago | Diagnostica Stago | Potency & Assay | Nationwide |
| Device | CLASS II | DxC 500i Clinical Analyzer Modules using recalled by Beckman Coulter | Beckman Coulter | Packaging Defects | Nationwide |
| Device | CLASS II | DxC 500 AU Clinical Analyzer Modules using recalled by Beckman Coulter | Beckman Coulter | Packaging Defects | Nationwide |
| Device | CLASS II | Syva EMIT 2000 Quinidine Assay recalled | Siemens Healthcare Diagnostics | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Beckman Coulter, Emit 2000 Quinidine Assay recalled | Siemens Healthcare Diagnostics | Lead & Heavy Metals | Nationwide |