CLASS II ONGOING Device recall · Reported 4 Nov 2026 · FDA Enforcement Report
Nucleic acid sample preparation instrument IVD, automated recalled
BioMerieux is recalling Nucleic acid sample preparation instrument IVD, automated, distributed nationwide in the US. The recall was initiated on 11 Sep 2026 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0301-2027
- FDA event ID
- 99931
- Recalling firm
- BioMerieux, Marcy L'Etoile
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 11 Sep 2026
- FDA classified
- 27 Oct 2026
- Reported
- 4 Nov 2026
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Other Medical Devices
Timeline
| 11 Sep 2026 | Recall initiated by company | |
| 27 Oct 2026 | Classified by FDA | +46 days |
| 4 Nov 2026 | Published in enforcement report | +8 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Delayed or no result when analyzing clinical samples with EMAG system (ref: 418591/418591U) due to a randomly occurring software error, Error 10003 Heater timeout after updating to recent firmware SP4 Patch 2
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
EMAG system Model/Catalog Number: 418591/418591U Product Description: Nucleic acid sample preparation instrument IVD, automated
Where it was distributed
Worldwide - US Nationwide distribution in the states of CA, IN, KS, LA, MN, MO, NY, OH, and TX. The countries of Canada, United Kingdom, Sweden, France, Norway, Netherlands, Belgium, Denmark, Germany, Austria.
Nationwide CaliforniaIndianaKansasLouisianaMinnesotaMissouriNew YorkOhioTexas
Questions about this recall
Why was it recalled?
Delayed or no result when analyzing clinical samples with EMAG system (ref: 418591/418591U) due to a randomly occurring software error, Error 10003 Heater timeout after updating to recent firmware SP4 Patch 2
Which lots are affected?
UDI-DI 357302BUDI000544TP Serial Numbers IM03343 IM03283 IM03104 IM03329 IM03376 IM03474 IM03321 IM03071 IM03175 IM03194 IM03515 IM03516 IM03240 IM03478 IM03479 IM03482 IM03034 IM03092 IM03378 IM03414 IM03177 IM03193 IM03506 IM3054 IM03439 IM03350 IM03380 IM03131 IM03148 IM03153 IM03275 IM03314 IM03351
Where was it sold?
Worldwide - US Nationwide distribution in the states of CA, IN, KS, LA, MN, MO, NY, OH, and TX. The countries of Canada, United Kingdom, Sweden, France, Norway, Netherlands, Belgium, Denmark, Germany, Austria.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 10 Nov 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from BioMerieux
All 478| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Vitek 2 Ast-N461 Test Kit, Gram Negative Susceptibility recalled | Biomerieux | Other | MO |
| Device | CLASS II | Vitek 2 Ast-N447 Test Kit Product reference recalled by Biomerieux | Biomerieux | Other | MO |
| Device | CLASS II | Vitek-negative Susceptibility Cards Containing o Polymyxin recalled | Biomerieux | Sterility | Nationwide |
| Device | CLASS II | Vitek¿-negative AST Test Kits (Cards) recalled by bioMerieux | bioMerieux | Sterility | NC |
| Device | CLASS I | Vitek 2 - AST Test kits (in vitro diagnostic test cards recalled | Biomerieux | Other | MO |
Other Other Medical Devices recalls
All 13,968| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Philips Azurion 3 M15 recalled by Philips Medical Systems Nederl | Philips Medical Systems Nederl | Other | Nationwide |
| Device | CLASS II | Orthoscan Versa Accessory Cart recalled by Ziehm-orthoscan | Ziehm-orthoscan | Device Malfunction | Nationwide |
| Device | CLASS II | Philips Azurion 7 B12 recalled by Philips Medical Systems Nederl | Philips Medical Systems Nederl | Other | Nationwide |
| Device | CLASS II | Philips Azurion 7 M20 recalled by Philips Medical Systems Nederl | Philips Medical Systems Nederl | Other | Nationwide |
| Device | CLASS II | Carevo, Height-adjustable shower trolley recalled | Arjohuntleigh Polska Sp. z.o.o | Other | Nationwide |