CLASS I ONGOING Device recall · Reported 1 May 2024 · FDA Enforcement Report

Vitek 2 - AST Test kits (in vitro diagnostic test cards recalled

Biomerieux is recalling Vitek 2 - AST Test kits (in vitro diagnostic test cards for antimicrobial. The recall was initiated on 14 Mar 2024 and the FDA has classified it as Class I.

Check your package

Lot, batch or date codesa) AST-N334, REF 418984, UDI/DI 3573026490720, Lot/Serial Numbers: 7742790503, 7742805403, 7742828503, 7742830203; b) AST-GN95, REF 421982, UDI/DI 3573026565923, Lot/Serial Numbers: 6852740103, 6852751503, 6852778503, 6852811503; c) AST-GN99, REF 423102, UDI/DI 3573026608132, Lot/Serial Numbers: 6892721103, 6892727403, 6892740503, 6892740403, 6892743503, 6892761503, 6892770403, 6892790103, 6892799403, 6892799103, 6892810103, 6892818103, 6892831503; d) AST-N390, REF 423340, UDI/DI 3573026612177, Lot/Serial Numbers: 1402832104, 1402761504; e) AST-N391, REF 423341, UDI/DI 3573026612269, Lot/Serial Numbers: 1412831104, 1412754504; f) AST-N392, REF 423342, UDI/DI 3573026612283, Lot/Serial Numbers: 1422819204; g) AST-N395, REF 423491, UDI/DI 3573026615802, Lot/Serial Numbers: 1452725204; h) AST-N401, REF 423643, UDI/DI 3573026616816, Lot/Serial Numbers: 1512776404, 1512789204, 1512811104, 1512831104, 1512720404, 1512729404, 1512729104, 1512748504, 1512754104, 1512785404; i) AST-N402, REF 423644, UDI/DI 3573026616830, Lot/Serial Numbers: 1522720504, 1522733404; j) AST-N404, REF 423664, UDI/DI 3573026617356, Lot/Serial Numbers: 1542819204; k) AST-N802, REF 423706, UDI/DI 3573026617752, Lot/Serial Numbers: 0422750104, 0422763104, 0422764504, 0422782404, 0422820504, 0422820404; l) AST-N405, REF 423864, UDI/DI 3573026621360, Lot/Serial Numbers: 1552757504, 1552792404, 1552800404, 1552820104, 1552733104, 1552741504, 1552757404; m) AST-N417, REF 423880, UDI/DI 3573026621704, Lot/Serial Numbers: 0192763504, 0192819404; n) AST-N408, REF 423924, UDI/DI 3573026622725, Lot/Serial Numbers: 1582810504, 1582820104, 1582734404, 1582742504, 1582764404, 1582782504; o) AST-N409, REF 423925, UDI/DI 3573026622749, Lot/Serial Numbers: 1592811104, 1592746404; p) AST-N420, REF 424039, UDI/DI 3573026625283, Lot/Serial Numbers: 0462810404; q) AST-N423, REF 424042, UDI/DI 3573026625337, Lot/Serial Numbers: 0502750404, 0502825504; r) AST-N422, REF 424056, UDI/DI 3573026625610, Lot/Serial Numbers: 0482819504; s) AST-N436, REF 424440, UDI/DI 3573026631666, Lot/Serial Numbers: 0672734104, 0672754104; t) AST-N809, REF 424703, UDI/DI 3573026635978, Lot/Serial Numbers: 0962799404; u) AST-N806, REF 424709, UDI/DI 3573026636067, Lot/Serial Numbers: 0932729404; v) AST-N807, REF 424710, UDI/DI 3573026636081, Lot/Serial Numbers: 0942723404; w) AST-N808, REF 424711, UDI/DI 3573026636104, Lot/Serial Numbers: 0952758104; x) AST-N810, REF 424712, UDI/DI 3573026636128, Lot/Serial Numbers: 0972729504; y) AST-N812, REF 424721, UDI/DI 3573026636296, Lot/Serial Numbers: 0992777104, 0992776504, 0992821404
Quantity recalled49215 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1464-2024
FDA event ID
94264
Recalling firm
Biomerieux, Hazelwood, MO
Classification
Class I
Status
Ongoing
Recall initiated
14 Mar 2024
FDA classified
25 Apr 2024
Reported
1 May 2024
Type
Voluntary: Firm initiated
Customers notified by
E-Mail
Reason category
Other Reasons

Timeline

14 Mar 2024Recall initiated by company
25 Apr 2024Classified by FDA+42 days
1 May 2024Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Ceftriaxone (cro02n) concentration errors associated with specific recently manufactured test kits which would have the potential for for false susceptible results. The risk is present for isolates that have a result of MIC=0.5, 1, 2.

Product as listed by the FDA

VITEK 2 - AST Test kits (in vitro diagnostic test cards for antimicrobial susceptibility): a) AST-N334, REF 418984; b) AST-GN95, REF 421982; c) AST-GN99, REF 423102; d) AST-N390, REF 423340; e) AST-N391, REF 423341; f) AST-N392, REF 423342; g) AST-N395, REF 423491; h) AST-N401, REF 423643; i) AST-N402, REF 423644; j) AST-N404, REF 423664; k) AST-N802, REF 423706; l) AST-N405, REF 423864; m) AST-N417, REF 423880; n) AST-N408, REF 423924; o) AST-N409, REF 423925; p) AST-N420, REF 424039; q) AST-N423, REF 424042; r) AST-N422, REF 424056; s) AST-N436, REF 424440; t) AST-N809, REF 424703; u) AST-N806, REF 424709; v) AST-N807, REF 424710; w) AST-N808, REF 424711; x) AST-N810, REF 424712; y) AST-N812, REF 424721

Where it was distributed

Worldwide

Questions about this recall

Why was it recalled?

Ceftriaxone (cro02n) concentration errors associated with specific recently manufactured test kits which would have the potential for for false susceptible results. The risk is present for isolates that have a result of MIC=0.5, 1, 2.

Which lots are affected?

a) AST-N334, REF 418984, UDI/DI 3573026490720, Lot/Serial Numbers: 7742790503, 7742805403, 7742828503, 7742830203; b) AST-GN95, REF 421982, UDI/DI 3573026565923, Lot/Serial Numbers: 6852740103, 6852751503, 6852778503, 6852811503; c) AST-GN99, REF 423102, UDI/DI 3573026608132, Lot/Serial Numbers: 6892721103, 6892727403, 6892740503, 6892740403, 6892743503, 6892761503, 6892770403, 6892790103, 6892799403, 6892799103, 6892810103, 6892818103, 6892831503; d) AST-N390, REF 423340, UDI/DI 3573026612177, Lot/Serial Numbers: 1402832104, 1402761504; e) AST-N391, REF 423341, UDI/DI 3573026612269, Lot/Serial Numbers: 1412831104, 1412754504; f) AST-N392, REF 423342, UDI/DI 3573026612283, Lot/Serial Numbers: 1422819204; g) AST-N395, REF 423491, UDI/DI 3573026615802, Lot/Serial Numbers: 1452725204; h) AST-N401, REF 423643, UDI/DI 3573026616816, Lot/Serial Numbers: 1512776404, 1512789204, 1512811104, 1512831104, 1512720404, 1512729404, 1512729104, 1512748504, 1512754104, 1512785404; i) AST-N402, REF 423644, UDI/DI 3573026616830, Lot/Serial Numbers: 1522720504, 1522733404; j) AST-N404, REF 423664, UDI/DI 3573026617356, Lot/Serial Numbers: 1542819204; k) AST-N802, REF 423706, UDI/DI 3573026617752, Lot/Serial Numbers: 0422750104, 0422763104, 0422764504, 0422782404, 0422820504, 0422820404; l) AST-N405, REF 423864, UDI/DI 3573026621360, Lot/Serial Numbers: 1552757504, 1552792404, 1552800404, 1552820104, 1552733104, 1552741504, 1552757404; m) AST-N417, REF 423880, UDI/DI 3573026621704, Lot/Serial Numbers: 0192763504, 0192819404; n) AST-N408, REF 423924, UDI/DI 3573026622725, Lot/Serial Numbers: 1582810504, 1582820104, 1582734404, 1582742504, 1582764404, 1582782504; o) AST-N409, REF 423925, UDI/DI 3573026622749, Lot/Serial Numbers: 1592811104, 1592746404; p) AST-N420, REF 424039, UDI/DI 3573026625283, Lot/Serial Numbers: 0462810404; q) AST-N423, REF 424042, UDI/DI 3573026625337, Lot/Serial Numbers: 0502750404, 0502825504; r) AST-N422, REF 424056, UDI/DI 3573026625610, Lot/Serial Numbers: 0482819504; s) AST-N436, REF 424440, UDI/DI 3573026631666, Lot/Serial Numbers: 0672734104, 0672754104; t) AST-N809, REF 424703, UDI/DI 3573026635978, Lot/Serial Numbers: 0962799404; u) AST-N806, REF 424709, UDI/DI 3573026636067, Lot/Serial Numbers: 0932729404; v) AST-N807, REF 424710, UDI/DI 3573026636081, Lot/Serial Numbers: 0942723404; w) AST-N808, REF 424711, UDI/DI 3573026636104, Lot/Serial Numbers: 0952758104; x) AST-N810, REF 424712, UDI/DI 3573026636128, Lot/Serial Numbers: 0972729504; y) AST-N812, REF 424721, UDI/DI 3573026636296, Lot/Serial Numbers: 0992777104, 0992776504, 0992821404

Where was it sold?

Worldwide

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class I mean?

Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.